Anne-Maria Boldt, Walter J Schulz-Schaeffer, Hicham Benkhai, Axel Kramer
{"title":"Material compatibility of guanidine thiocyanate for decontamination of nickel-titanium root canal instruments after potential exposure to prions.","authors":"Anne-Maria Boldt, Walter J Schulz-Schaeffer, Hicham Benkhai, Axel Kramer","doi":"10.3205/dgkh000548","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>Due to the effectiveness of guanidine thiocyanate (GdnSCN) for the decontamination of prion protein aggregates, which are the causative agent of transmissible spongiform encephalopathy, the influence on the bending stiffness and torsional strength of endodontic nickel-titanium files should be tested to provide a potential alternative to single-use if necessary.</p><p><strong>Method: </strong>For the investigation, nitrite-titanium-coated EasyShape<sup>®</sup> files of sizes 25.06 and 35.04 were placed in 6 M GdnSCN solution 8 times for 15 min each for decontamination in line with the manufacturer's recommendation, with intermediate drying in each case. To simulate the worst case, the soaking time was extended to 12 h once. Both the bending stiffness and the torsional and fracture behavior were determined in accordance with DIN EN ISO 3630-1:2008-04.</p><p><strong>Results: </strong>Compared to the untreated control (n=12), decontamination with GdnSCN has no effect on the torsional strength or flexural rigidity of the tested endodontic instruments of sizes 25.06 and 35.04 (n=18 each) when used properly. On the other hand, the exposure time of 12 h reduced the bending moment and torsion angle of instrument size 25.06, while the material properties of size 35.04 files are not affected.</p><p><strong>Discussion: </strong>Based on the results, the maximum 8-fold application of 6 M GdnSCN solution for 15 min for decontamination of endodontic nickel-titanium files can be considered. Although the sterilization process has no influence on the fracture behavior, it is important to clarify for clinical practice what influence the overall reprocessing process has on the performance of the instruments.</p>","PeriodicalId":12738,"journal":{"name":"GMS Hygiene and Infection Control","volume":"20 ","pages":"Doc19"},"PeriodicalIF":1.6000,"publicationDate":"2025-05-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12171978/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"GMS Hygiene and Infection Control","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3205/dgkh000548","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH","Score":null,"Total":0}
引用次数: 0
Abstract
Aim: Due to the effectiveness of guanidine thiocyanate (GdnSCN) for the decontamination of prion protein aggregates, which are the causative agent of transmissible spongiform encephalopathy, the influence on the bending stiffness and torsional strength of endodontic nickel-titanium files should be tested to provide a potential alternative to single-use if necessary.
Method: For the investigation, nitrite-titanium-coated EasyShape® files of sizes 25.06 and 35.04 were placed in 6 M GdnSCN solution 8 times for 15 min each for decontamination in line with the manufacturer's recommendation, with intermediate drying in each case. To simulate the worst case, the soaking time was extended to 12 h once. Both the bending stiffness and the torsional and fracture behavior were determined in accordance with DIN EN ISO 3630-1:2008-04.
Results: Compared to the untreated control (n=12), decontamination with GdnSCN has no effect on the torsional strength or flexural rigidity of the tested endodontic instruments of sizes 25.06 and 35.04 (n=18 each) when used properly. On the other hand, the exposure time of 12 h reduced the bending moment and torsion angle of instrument size 25.06, while the material properties of size 35.04 files are not affected.
Discussion: Based on the results, the maximum 8-fold application of 6 M GdnSCN solution for 15 min for decontamination of endodontic nickel-titanium files can be considered. Although the sterilization process has no influence on the fracture behavior, it is important to clarify for clinical practice what influence the overall reprocessing process has on the performance of the instruments.
目的:鉴于胍硫氰酸酯(GdnSCN)对传染性海绵状脑病病原体朊病毒蛋白聚集体的净化效果,应测试其对牙髓镍钛锉的弯曲刚度和扭转强度的影响,以便在必要时提供潜在的替代一次性使用的方法。方法:将尺寸为25.06和35.04的亚硝酸盐钛包覆的easysshape®文件放入6m GdnSCN溶液中8次,每次15分钟,按照制造商的建议进行去污,每次进行中间干燥。为模拟最坏情况,将浸泡时间延长至12 h一次。弯曲刚度、扭转和断裂性能均按照DIN EN ISO 3630-1:2008-04进行测定。结果:与未处理的对照组(n=12)相比,在正确使用的情况下,GdnSCN对25.06和35.04尺寸的根管器械(n=18)的扭转强度和弯曲刚度没有影响。另一方面,12 h的曝光时间降低了25.06尺寸的仪器的弯矩和扭转角,而35.04尺寸的文件的材料性能没有受到影响。讨论:根据研究结果,可以考虑最大8倍应用6 M GdnSCN溶液,持续15 min,用于根管镍钛锉去污。虽然灭菌过程对骨折行为没有影响,但对于临床实践来说,明确整个再处理过程对器械性能的影响是很重要的。