Validation of the Omron J7136 and HEM-7361T blood pressure monitors according to the International Organization for Standardization protocol (ISO 81060-2:2018 + Amd.1:2020).

IF 1.2 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE
Yi Zhou, Xin-Yu Wang, Yi Zhou, Yan Li, Ji-Guang Wang
{"title":"Validation of the Omron J7136 and HEM-7361T blood pressure monitors according to the International Organization for Standardization protocol (ISO 81060-2:2018 + Amd.1:2020).","authors":"Yi Zhou, Xin-Yu Wang, Yi Zhou, Yan Li, Ji-Guang Wang","doi":"10.1097/MBP.0000000000000756","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study aimed to evaluate the accuracy of OMRON J7136 and HEM-7361T oscillometric blood pressure (BP) monitors in the general population according to the International Organization for Standardization (ISO) 81060-2:2018 and amendment (Amd) 1:2020 standard.</p><p><strong>Method: </strong>Subjects were recruited to fulfill the age, gender, BP, and cuff distributions of the ISO Universal Standard in the general population using the same arm sequential BP measurement method. Three cuffs of the OMRON J7136 were used for arm circumference 17-22 cm (small), 22-32 cm (medium), and 32-42 cm (large). A single cuff 17-36 cm was used on the OMRON HEM-7361T.</p><p><strong>Results: </strong>The mean ± SD of the differences between the OMRON J7136 and reference BP readings was -2.14 ± 6.45/-1.43 ± 5.88 mmHg (systolic/diastolic) for criterion 1. The SD of the average BP differences between the OMRON J7136 and reference BP per subject was 5.49/5.36 mmHg (systolic/diastolic) for criterion 2. For the study on the OMRON HEM-7361T, the corresponding values were -2.56 ± 6.35/-3.18 ± 5.67 mmHg (systolic/diastolic) for criterion 1 and 5.33/5.32 mmHg (systolic/diastolic) for criterion 2.</p><p><strong>Conclusion: </strong>The OMRON J7136 and HEM-7361T BP monitors both fulfilled the requirements of the ISO Universal Standard (ISO 81060-2:2018/AMD 1:2020) in the general population and can be recommended for self-measurement in adults.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":""},"PeriodicalIF":1.2000,"publicationDate":"2025-05-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000756","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: This study aimed to evaluate the accuracy of OMRON J7136 and HEM-7361T oscillometric blood pressure (BP) monitors in the general population according to the International Organization for Standardization (ISO) 81060-2:2018 and amendment (Amd) 1:2020 standard.

Method: Subjects were recruited to fulfill the age, gender, BP, and cuff distributions of the ISO Universal Standard in the general population using the same arm sequential BP measurement method. Three cuffs of the OMRON J7136 were used for arm circumference 17-22 cm (small), 22-32 cm (medium), and 32-42 cm (large). A single cuff 17-36 cm was used on the OMRON HEM-7361T.

Results: The mean ± SD of the differences between the OMRON J7136 and reference BP readings was -2.14 ± 6.45/-1.43 ± 5.88 mmHg (systolic/diastolic) for criterion 1. The SD of the average BP differences between the OMRON J7136 and reference BP per subject was 5.49/5.36 mmHg (systolic/diastolic) for criterion 2. For the study on the OMRON HEM-7361T, the corresponding values were -2.56 ± 6.35/-3.18 ± 5.67 mmHg (systolic/diastolic) for criterion 1 and 5.33/5.32 mmHg (systolic/diastolic) for criterion 2.

Conclusion: The OMRON J7136 and HEM-7361T BP monitors both fulfilled the requirements of the ISO Universal Standard (ISO 81060-2:2018/AMD 1:2020) in the general population and can be recommended for self-measurement in adults.

根据国际标准化组织协议(ISO 81060-2:2018 + Amd.1:2020)对欧姆龙J7136和HEM-7361T血压计进行验证。
目的:本研究旨在根据国际标准化组织(ISO) 81060-2:2018和修订(Amd) 1:2020 0标准,评估欧姆龙J7136和hm - 7361t振荡血压(BP)监测仪在普通人群中的准确性。方法:招募受试者,采用相同的臂序贯血压测量方法,满足一般人群中ISO通用标准的年龄、性别、血压和袖带分布。采用欧姆龙J7136三个袖口测量臂围17-22 cm(小)、22-32 cm(中)、32-42 cm(大)。欧姆龙HEM-7361T采用单袖带17-36 cm。结果:OMRON J7136与参考血压读数差异的平均值±SD为-2.14±6.45/-1.43±5.88 mmHg(收缩压/舒张压)。对于标准2,每位受试者的OMRON J7136与参考血压的平均血压差的标准差为5.49/5.36 mmHg(收缩压/舒张压)。在欧姆龙HEM-7361T的研究中,标准1的相应值为-2.56±6.35/-3.18±5.67 mmHg(收缩压/舒张压),标准2的相应值为5.33/5.32 mmHg(收缩压/舒张压)。结论:欧姆龙J7136和hm - 7361t血压监测仪在普通人群中均符合ISO通用标准(ISO 81060-2:2018/AMD 1:2020)的要求,可推荐用于成人自我测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信