Repeatability and reproducibility of Keratograph 5M corneal topography.

IF 2.3 3区 生物学 Q2 MULTIDISCIPLINARY SCIENCES
PeerJ Pub Date : 2025-05-22 eCollection Date: 2025-01-01 DOI:10.7717/peerj.19478
Sara Ortiz-Toquero, Irene Sanchez, Raul Martin
{"title":"Repeatability and reproducibility of Keratograph 5M corneal topography.","authors":"Sara Ortiz-Toquero, Irene Sanchez, Raul Martin","doi":"10.7717/peerj.19478","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Corneal topography is an important diagnostic tool and highly repeatable and reproducible topographic devices are essential in eye care practice. Placido disc-based topography is one of the most widely used methods because of its high resolution and accuracy. The aim of this study was to analyse the intrasession repeatability and intersession reproducibility of measurements obtained with a Keratograph 5M in a sample of healthy subjects.</p><p><strong>Methods: </strong>Three consecutive measurements were performed with a Keratograph 5M during two sessions in 24 healthy subjects to calculate the within-subject standard deviation (Sw), repeatability and reproducibility limits, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) of corneal curvature (K1, K2 and Max-K), eccentricity, corneal diameter, and corneal aberration (6-mm pupil; coma, trefoil, spherical aberration, secondary astigmatism and quadrafoil).</p><p><strong>Results: </strong>No statistically significant differences were found between the three measurements in either session (<i>P</i> ≥ 0.06). The corneal parameters that demonstrated the best repeatability were corneal curvature and corneal diameter with a CoV, and the ICCs ranged from 0.41% and 0.990 (corneal diameter) to 0.28% and 0.998 (K2). Eccentricity and corneal aberrations had lower repeatability results, with CoVs and ICCs ranging from 3.88% and 0.992 (spherical aberration) to 40.21% and 0.643 (quadrafoil), respectively. In the case of reproducibility, excellent results were obtained for corneal curvature and diameter measurements (CoV ≤ 0.36% and ICC ≥ 0.987), with moderate reproducibility for corneal eccentricity (CoV ≥ 2.03% and ICC ≤ 0.986), secondary astigmatism (CoV = 20.05% and ICC = 0.787), and quadrafoil (CoV = 23.55% and ICC = 0.696).</p><p><strong>Conclusions: </strong>The Keratograph 5M demonstrated excellent repeatability and reproducibility in measuring corneal curvature and corneal diameter in healthy subjects. Corneal eccentricity shows moderate accuracy, whereas corneal aberrations (except coma, trefoil, and spherical aberrations) exhibit moderate measurement reliability and should be interpreted with caution in clinical practice.</p>","PeriodicalId":19799,"journal":{"name":"PeerJ","volume":"13 ","pages":"e19478"},"PeriodicalIF":2.3000,"publicationDate":"2025-05-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12103842/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"PeerJ","FirstCategoryId":"99","ListUrlMain":"https://doi.org/10.7717/peerj.19478","RegionNum":3,"RegionCategory":"生物学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"MULTIDISCIPLINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Corneal topography is an important diagnostic tool and highly repeatable and reproducible topographic devices are essential in eye care practice. Placido disc-based topography is one of the most widely used methods because of its high resolution and accuracy. The aim of this study was to analyse the intrasession repeatability and intersession reproducibility of measurements obtained with a Keratograph 5M in a sample of healthy subjects.

Methods: Three consecutive measurements were performed with a Keratograph 5M during two sessions in 24 healthy subjects to calculate the within-subject standard deviation (Sw), repeatability and reproducibility limits, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) of corneal curvature (K1, K2 and Max-K), eccentricity, corneal diameter, and corneal aberration (6-mm pupil; coma, trefoil, spherical aberration, secondary astigmatism and quadrafoil).

Results: No statistically significant differences were found between the three measurements in either session (P ≥ 0.06). The corneal parameters that demonstrated the best repeatability were corneal curvature and corneal diameter with a CoV, and the ICCs ranged from 0.41% and 0.990 (corneal diameter) to 0.28% and 0.998 (K2). Eccentricity and corneal aberrations had lower repeatability results, with CoVs and ICCs ranging from 3.88% and 0.992 (spherical aberration) to 40.21% and 0.643 (quadrafoil), respectively. In the case of reproducibility, excellent results were obtained for corneal curvature and diameter measurements (CoV ≤ 0.36% and ICC ≥ 0.987), with moderate reproducibility for corneal eccentricity (CoV ≥ 2.03% and ICC ≤ 0.986), secondary astigmatism (CoV = 20.05% and ICC = 0.787), and quadrafoil (CoV = 23.55% and ICC = 0.696).

Conclusions: The Keratograph 5M demonstrated excellent repeatability and reproducibility in measuring corneal curvature and corneal diameter in healthy subjects. Corneal eccentricity shows moderate accuracy, whereas corneal aberrations (except coma, trefoil, and spherical aberrations) exhibit moderate measurement reliability and should be interpreted with caution in clinical practice.

5M角膜地形图的重复性和再现性。
背景:角膜地形图是一种重要的诊断工具,高度可重复和可复制的地形图设备在眼科保健实践中是必不可少的。基于普拉西多圆盘的地形图以其较高的分辨率和精度成为应用最广泛的地形图方法之一。本研究的目的是分析5M角膜摄影仪在健康受试者样本中获得的测量结果的遗传内重复性和遗传间重复性。方法:对24名健康受试者进行两次连续3次的角膜摄影仪5M测量,计算角膜曲率(K1、K2和Max-K)、偏心率、角膜直径和角膜像差(6mm瞳孔;彗差、三叶草差、球差、次生散光和四像差)。结果:两组间三次测量均无统计学差异(P≥0.06)。重复性最好的角膜参数是角膜曲率和角膜直径,ICCs范围为0.41%和0.990(角膜直径)至0.28%和0.998 (K2)。偏心率和角膜像差的重复性较低,cov和icc分别为3.88%和0.992(球差)至40.21%和0.643(四像差)。在可重复性方面,角膜曲率和直径测量(CoV≤0.36%,ICC≥0.987)获得了良好的结果,角膜偏心率(CoV≥2.03%,ICC≤0.986)、次生散光(CoV = 20.05%, ICC = 0.787)和四联体(CoV = 23.55%, ICC = 0.696)的可重复性中等。结论:5M角膜摄影仪测量健康人角膜曲率和角膜直径具有良好的重复性和再现性。角膜偏心率具有中等的准确性,而角膜像差(除了彗差、三叶草差和球差)具有中等的测量可靠性,在临床实践中应谨慎解释。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
PeerJ
PeerJ MULTIDISCIPLINARY SCIENCES-
CiteScore
4.70
自引率
3.70%
发文量
1665
审稿时长
10 weeks
期刊介绍: PeerJ is an open access peer-reviewed scientific journal covering research in the biological and medical sciences. At PeerJ, authors take out a lifetime publication plan (for as little as $99) which allows them to publish articles in the journal for free, forever. PeerJ has 5 Nobel Prize Winners on the Board; they have won several industry and media awards; and they are widely recognized as being one of the most interesting recent developments in academic publishing.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信