Effect of Fluid Contamination on Reverse Torque of Healing Abutments in Two Implant Connection Systems: In Vitro Study

IF 3.7 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Shifra Levartovsky, Guy Ronen, Perry Raz, Ilan Beitlitum
{"title":"Effect of Fluid Contamination on Reverse Torque of Healing Abutments in Two Implant Connection Systems: In Vitro Study","authors":"Shifra Levartovsky,&nbsp;Guy Ronen,&nbsp;Perry Raz,&nbsp;Ilan Beitlitum","doi":"10.1111/cid.70056","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>This in vitro study investigates the impact of blood, saliva, and chlorhexidine contamination on the reverse torque values (RTVs) of healing abutments in two types of implant-abutment connections: internal hex and conical.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A total of 88 Ti6Al4V titanium alloy dental implants were tested, comprising two types of connections: internal hex (<i>n</i> = 44, MIS Seven) and conical (<i>n</i> = 44, MIS C1). Each group was further divided into four subgroups (<i>n</i> = 11) based on the type of contamination medium used. Healing abutments were tightened onto the implant fixtures using a torque of 25 Ncm with the designated medium, followed by retightening after a 10-min interval. After a two-week incubation period, reverse torque values (RTVs) were recorded using a digital torque gauge. To assess differences between the implant systems within the same medium, as well as differences between the medium types within the same implant system, the Welch two-sample <i>t</i>-test was employed. Additionally, linear regression analysis was conducted to evaluate the interaction between medium types and the implant systems.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>For MIS Seven implants, RTVs were consistently lower than the initial 25 Ncm torque across all contamination groups. In contrast, for MIS C1 implants, RTVs exceeded 25 Ncm in the saliva and blood contamination groups; however, the presence of chlorhexidine reduced the RTVs. A comparative analysis revealed that the C1 conical connection required significantly higher RTVs than the Seven internal hex system (coefficient = 3.318). No significant differences were observed between implant systems in the control and chlorhexidine groups. However, the C1 system required higher RTVs than the Seven system in the presence of saliva (<i>p</i> = 0.0006) and blood (<i>p</i> = 0.00009). Furthermore, the interaction analysis indicated that in the presence of saliva and blood contamination, C1 implants required significantly higher RTVs, with mean differences of 7.18 and 4.9 (Ncm), respectively, compared to the Seven implants.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Conical implant-abutment connections generally require higher RTVs for abutment disconnection compared to internal hex connections. Both implant connection and the type of contaminant significantly affect RTVs. Chlorhexidine has been shown to reduce reverse torque for both types of connections. In contrast, contamination with saliva and blood tends to increase reverse torque, particularly for conical connections.</p>\n </section>\n </div>","PeriodicalId":50679,"journal":{"name":"Clinical Implant Dentistry and Related Research","volume":"27 3","pages":""},"PeriodicalIF":3.7000,"publicationDate":"2025-05-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/cid.70056","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Implant Dentistry and Related Research","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/cid.70056","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction

This in vitro study investigates the impact of blood, saliva, and chlorhexidine contamination on the reverse torque values (RTVs) of healing abutments in two types of implant-abutment connections: internal hex and conical.

Methods

A total of 88 Ti6Al4V titanium alloy dental implants were tested, comprising two types of connections: internal hex (n = 44, MIS Seven) and conical (n = 44, MIS C1). Each group was further divided into four subgroups (n = 11) based on the type of contamination medium used. Healing abutments were tightened onto the implant fixtures using a torque of 25 Ncm with the designated medium, followed by retightening after a 10-min interval. After a two-week incubation period, reverse torque values (RTVs) were recorded using a digital torque gauge. To assess differences between the implant systems within the same medium, as well as differences between the medium types within the same implant system, the Welch two-sample t-test was employed. Additionally, linear regression analysis was conducted to evaluate the interaction between medium types and the implant systems.

Results

For MIS Seven implants, RTVs were consistently lower than the initial 25 Ncm torque across all contamination groups. In contrast, for MIS C1 implants, RTVs exceeded 25 Ncm in the saliva and blood contamination groups; however, the presence of chlorhexidine reduced the RTVs. A comparative analysis revealed that the C1 conical connection required significantly higher RTVs than the Seven internal hex system (coefficient = 3.318). No significant differences were observed between implant systems in the control and chlorhexidine groups. However, the C1 system required higher RTVs than the Seven system in the presence of saliva (p = 0.0006) and blood (p = 0.00009). Furthermore, the interaction analysis indicated that in the presence of saliva and blood contamination, C1 implants required significantly higher RTVs, with mean differences of 7.18 and 4.9 (Ncm), respectively, compared to the Seven implants.

Conclusions

Conical implant-abutment connections generally require higher RTVs for abutment disconnection compared to internal hex connections. Both implant connection and the type of contaminant significantly affect RTVs. Chlorhexidine has been shown to reduce reverse torque for both types of connections. In contrast, contamination with saliva and blood tends to increase reverse torque, particularly for conical connections.

液体污染对两种种植体连接系统修复基台反向扭矩的影响:体外研究
本体外研究探讨了血液、唾液和氯己定污染对两种类型种植体-基台连接中愈合基台的反向扭矩值(RTVs)的影响:内六角和锥形。方法对88个Ti6Al4V钛合金种植体进行试验,包括内六角(n = 44, MIS 7)和锥形(n = 44, MIS C1)两种连接方式。根据使用的污染介质类型将每组进一步分为4个亚组(n = 11)。使用指定介质将愈合基牙以25 Ncm的扭矩紧固到种植体固定物上,间隔10分钟后重新紧固。在两周的潜伏期后,使用数字扭矩计记录反向扭矩值(RTVs)。为了评估同一培养基中不同种植体系统之间的差异,以及同一种植体系统中不同培养基类型之间的差异,采用Welch双样本t检验。此外,还进行了线性回归分析,以评估培养基类型与种植体系统之间的相互作用。结果对于MIS - 7种植体,所有污染组的rtv均低于初始25 Ncm扭矩。相比之下,对于MIS C1植入物,唾液和血液污染组的rtv超过25 Ncm;然而,氯己定的存在降低了rtv。对比分析表明,C1型锥形连接的rtv值明显高于7型内六角连接(系数= 3.318)。在对照组和氯己定组的种植体系统之间没有观察到显著差异。然而,在唾液(p = 0.0006)和血液(p = 0.00009)存在的情况下,C1系统比Seven系统需要更高的rtv。此外,交互作用分析表明,在存在唾液和血液污染的情况下,C1种植体所需的rtv显著高于7种种植体,平均差异分别为7.18和4.9 (Ncm)。结论锥形种植体-基台连接比内六角连接需要更高的rtv。种植体连接和污染物类型对rtv都有显著影响。氯己定已被证明可以减少这两种类型连接的反向扭矩。相反,唾液和血液的污染往往会增加反向扭矩,特别是对于锥形连接。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.00
自引率
13.90%
发文量
103
审稿时长
4-8 weeks
期刊介绍: The goal of Clinical Implant Dentistry and Related Research is to advance the scientific and technical aspects relating to dental implants and related scientific subjects. Dissemination of new and evolving information related to dental implants and the related science is the primary goal of our journal. The range of topics covered by the journals will include but be not limited to: New scientific developments relating to bone Implant surfaces and their relationship to the surrounding tissues Computer aided implant designs Computer aided prosthetic designs Immediate implant loading Immediate implant placement Materials relating to bone induction and conduction New surgical methods relating to implant placement New materials and methods relating to implant restorations Methods for determining implant stability A primary focus of the journal is publication of evidenced based articles evaluating to new dental implants, techniques and multicenter studies evaluating these treatments. In addition basic science research relating to wound healing and osseointegration will be an important focus for the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信