Unmasking loss of capture and prolonged left ventricular latency by the Biotronik desynchronization-detection algorithm

IF 1.3 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
S. Serge Barold MD , Andreas Kucher MSc, PhD
{"title":"Unmasking loss of capture and prolonged left ventricular latency by the Biotronik desynchronization-detection algorithm","authors":"S. Serge Barold MD ,&nbsp;Andreas Kucher MSc, PhD","doi":"10.1016/j.jelectrocard.2025.154023","DOIUrl":null,"url":null,"abstract":"<div><div>Loss of pacing capture of the right or left ventricle as well as prolongation of the device-defined LV latency can be automatically unmasked in Biotronik CRT-D devices by Home Monitory of the CRT pacing interrupt algorithm. These situations are associated with short intervals from a pacing stimulus to conducted ventricular activation. These intervals are relatively constant with device-defined latency while those from loss of capture are longer and may be more variable. The true LV latency interval should be re-evaluated whenever the device-defined latency is apparent and prolonged.</div></div>","PeriodicalId":15606,"journal":{"name":"Journal of electrocardiology","volume":"91 ","pages":"Article 154023"},"PeriodicalIF":1.3000,"publicationDate":"2025-05-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of electrocardiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0022073625001517","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Loss of pacing capture of the right or left ventricle as well as prolongation of the device-defined LV latency can be automatically unmasked in Biotronik CRT-D devices by Home Monitory of the CRT pacing interrupt algorithm. These situations are associated with short intervals from a pacing stimulus to conducted ventricular activation. These intervals are relatively constant with device-defined latency while those from loss of capture are longer and may be more variable. The true LV latency interval should be re-evaluated whenever the device-defined latency is apparent and prolonged.
通过Biotronik去同步检测算法揭开捕获丢失和左心室潜伏期延长的掩蔽
在Biotronik CRT- d设备中,通过CRT起搏中断算法的Home monitor,可以自动发现右心室或左心室起搏捕获的丢失以及设备定义的LV延迟的延长。这些情况与从起搏刺激到传导心室激活的短间隔有关。这些时间间隔相对于设备定义的延迟是恒定的,而丢失捕获的时间间隔则更长,并且可能变化更大。当设备定义的延迟明显且延长时,应该重新评估真实的LV延迟间隔。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of electrocardiology
Journal of electrocardiology 医学-心血管系统
CiteScore
2.70
自引率
7.70%
发文量
152
审稿时长
38 days
期刊介绍: The Journal of Electrocardiology is devoted exclusively to clinical and experimental studies of the electrical activities of the heart. It seeks to contribute significantly to the accuracy of diagnosis and prognosis and the effective treatment, prevention, or delay of heart disease. Editorial contents include electrocardiography, vectorcardiography, arrhythmias, membrane action potential, cardiac pacing, monitoring defibrillation, instrumentation, drug effects, and computer applications.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信