Anna V. Phillips;Cherie M. Kuzmiak;Gabriela Torres;Caterina M. Gallippi
{"title":"VisR Ultrasound Improves Diagnosis of Breast Cancer by the Elastogram-to-B-Mode Ratio in a Blinded Reader Study","authors":"Anna V. Phillips;Cherie M. Kuzmiak;Gabriela Torres;Caterina M. Gallippi","doi":"10.1109/OJUFFC.2025.3560585","DOIUrl":null,"url":null,"abstract":"Ultrasound elastography is increasingly being used alongside mammography for breast cancer diagnosis, particularly in women with radiographically dense breasts. The elastogram-to-B-Mode ratio (E/B), which compares lesion sizes in B-Mode and stiffness images, has been shown to differentiate malignant (E/B >1) from benign (E/B <1)>1 in malignant lesions and <1 in benign lesions) to those observed in stiffness images. Additionally, E/B values calculated from RE alone, RV alone, or a combination of RE and RV achieved 4-16% higher AUCs for discriminating malignant lesions compared to E/B derived solely from Acoustic Radiation Force Impulse (ARFI) peak displacement. These results suggest that incorporating VisR-derived viscosity and elasticity metrics into E/B calculations could significantly improve diagnostic accuracy for breast cancer detection.","PeriodicalId":73301,"journal":{"name":"IEEE open journal of ultrasonics, ferroelectrics, and frequency control","volume":"5 ","pages":"67-76"},"PeriodicalIF":0.0000,"publicationDate":"2025-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10964276","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE open journal of ultrasonics, ferroelectrics, and frequency control","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10964276/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Ultrasound elastography is increasingly being used alongside mammography for breast cancer diagnosis, particularly in women with radiographically dense breasts. The elastogram-to-B-Mode ratio (E/B), which compares lesion sizes in B-Mode and stiffness images, has been shown to differentiate malignant (E/B >1) from benign (E/B <1)>1 in malignant lesions and <1 in benign lesions) to those observed in stiffness images. Additionally, E/B values calculated from RE alone, RV alone, or a combination of RE and RV achieved 4-16% higher AUCs for discriminating malignant lesions compared to E/B derived solely from Acoustic Radiation Force Impulse (ARFI) peak displacement. These results suggest that incorporating VisR-derived viscosity and elasticity metrics into E/B calculations could significantly improve diagnostic accuracy for breast cancer detection.