Implementing 100% quality control in a cervical cytology workflow using whole slide images and artificial intelligence provided by the Techcyte SureView™ System
Maria del Mar Rivera Rolon MD, FCAP, Erik Gustafson PhD, Riley Cole BA, MA, Jaylene Matos CT (ASCP), CM, Kellie Hicken CT (ASCP), BS, Jacob Hicks BS, MBA, Brian Cahoon BS, MBA, Mariano de Socarraz, Juan Carlos Santa-Rosario MD, FCAP, FASCP
{"title":"Implementing 100% quality control in a cervical cytology workflow using whole slide images and artificial intelligence provided by the Techcyte SureView™ System","authors":"Maria del Mar Rivera Rolon MD, FCAP, Erik Gustafson PhD, Riley Cole BA, MA, Jaylene Matos CT (ASCP), CM, Kellie Hicken CT (ASCP), BS, Jacob Hicks BS, MBA, Brian Cahoon BS, MBA, Mariano de Socarraz, Juan Carlos Santa-Rosario MD, FCAP, FASCP","doi":"10.1002/cncy.70019","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Recent advancements in digital pathology have extended into cytopathology. Laboratories screening cervical cytology specimens now choose between limited imaging options and traditional manual microscopy. The Techcyte SureView™ Cervical Cytology System, designed for digital cytopathology, was validated at CorePlus, a pathology laboratory in Puerto Rico, and adopted as a 100% quality control (QC) tool.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The validation study included 1442 whole slide images (WSIs) from 1273 ThinPrep® and 169 SurePath™ cervical cytology slides, digitized with the 3DHISTECH Panoramic 1000 DX scanner using dry and water immersion scanning profiles. These WSIs were processed by the Techcyte SureView™ system, with a board-certified cytopathologist reviewing artificial intelligence (AI)-identified objects of interest and comparing them to traditional light microscopy results.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Techcyte SureView™ with the water immersion scanning profile outperformed both the dry scanning profile and light microscopy in detecting squamous and glandular abnormalities. It achieved 97% accuracy, 82% sensitivity, 99% specificity, 98% negative predictive value, and 86% positive predictive value. Additionally, the review time was rapid. The system has been operational for several months, enhancing accuracy and workflow efficiency.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>This study demonstrates that digital cytopathology, particularly through the Techcyte SureView™ system, can improve laboratory workflow and performance. Successful validation led CorePlus to integrate the AI algorithm into their workflow as a 100% QC review tool, resulting in improved accuracy, benefiting both laboratory professionals and patients.</p>\n </section>\n </div>","PeriodicalId":9410,"journal":{"name":"Cancer Cytopathology","volume":"133 6","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2025-05-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/cncy.70019","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cancer Cytopathology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cncy.70019","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Recent advancements in digital pathology have extended into cytopathology. Laboratories screening cervical cytology specimens now choose between limited imaging options and traditional manual microscopy. The Techcyte SureView™ Cervical Cytology System, designed for digital cytopathology, was validated at CorePlus, a pathology laboratory in Puerto Rico, and adopted as a 100% quality control (QC) tool.
Methods
The validation study included 1442 whole slide images (WSIs) from 1273 ThinPrep® and 169 SurePath™ cervical cytology slides, digitized with the 3DHISTECH Panoramic 1000 DX scanner using dry and water immersion scanning profiles. These WSIs were processed by the Techcyte SureView™ system, with a board-certified cytopathologist reviewing artificial intelligence (AI)-identified objects of interest and comparing them to traditional light microscopy results.
Results
Techcyte SureView™ with the water immersion scanning profile outperformed both the dry scanning profile and light microscopy in detecting squamous and glandular abnormalities. It achieved 97% accuracy, 82% sensitivity, 99% specificity, 98% negative predictive value, and 86% positive predictive value. Additionally, the review time was rapid. The system has been operational for several months, enhancing accuracy and workflow efficiency.
Conclusions
This study demonstrates that digital cytopathology, particularly through the Techcyte SureView™ system, can improve laboratory workflow and performance. Successful validation led CorePlus to integrate the AI algorithm into their workflow as a 100% QC review tool, resulting in improved accuracy, benefiting both laboratory professionals and patients.
期刊介绍:
Cancer Cytopathology provides a unique forum for interaction and dissemination of original research and educational information relevant to the practice of cytopathology and its related oncologic disciplines. The journal strives to have a positive effect on cancer prevention, early detection, diagnosis, and cure by the publication of high-quality content. The mission of Cancer Cytopathology is to present and inform readers of new applications, technological advances, cutting-edge research, novel applications of molecular techniques, and relevant review articles related to cytopathology.