Effect of damage or contamination to the tips of 200 light-curing units.

IF 2.6 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Abdulrahman A Balhaddad, Afnan O Al-Zain, Hassan A Alyami, Husain A Almakrami, Osama A Alsulaiman, Eman H Ismail, Richard B Price, Ahmed A Alsulaiman
{"title":"Effect of damage or contamination to the tips of 200 light-curing units.","authors":"Abdulrahman A Balhaddad, Afnan O Al-Zain, Hassan A Alyami, Husain A Almakrami, Osama A Alsulaiman, Eman H Ismail, Richard B Price, Ahmed A Alsulaiman","doi":"10.1186/s12903-025-06126-8","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The light-curing unit (LCU) has become a vital piece of dental equipment that must be correctly maintained. This study investigated the impact of contamination and physical damage to the light tip on the power and radiant emittance values from old and new LCUs.</p><p><strong>Materials and methods: </strong>Two investigators assessed 200 LCUs in dental clinics. The extent of contamination and physical damage to the light-curing unit (LCU) tips was recorded using a scale ranging from 0 to 8, where 0 indicates the absence of damage or contamination, and 8 represents severe damage or contamination. Then, the radiant emittance and power values of the LCU tip were measured using a digital radiometer (Bluephase meter II; Ivoclar, Schaan, Liechtenstein). LCUs that were more than five years old were classified as old. Spearman correlation coefficient was used to determine the relationship between the condition of the LCU and radiant emittance/power (p = 0.05).</p><p><strong>Results: </strong>There were no significant differences in the percent reduction of the power and radiant emittance from the values reported by the manufacturers, as well as the presence of contamination or physical damage scores between old and relatively new light-curing tips (p > 0.05). The mean ± standard deviation percentage reductions in power and radiant emittance from the manufacturer's stated values were 19.2 ± 17.63% and 3.9 ± 16.49%, respectively. Contamination and physical damage had significant positive correlations with the reduction in the power (r = 0.22070, p = 0.0017 and r = 0.27422, p < 0.0001, respectively) and the reduction in the radiant emittance (r = 0.28626, p < 0.0001 and r = 0.36650, p < 0.0001). Increased contamination and physical damage scores corresponded to greater percent reductions in the power and radiant emittance (p < 0.05).</p><p><strong>Conclusions: </strong>Contamination and physical damage to the LCU can negatively impact the light output from LCUs.</p><p><strong>Clinical relevance: </strong>To ensure optimal performance, dentists should regularly monitor the output of their LCUs and examine the devices for any signs of physical damage or contamination.</p>","PeriodicalId":9072,"journal":{"name":"BMC Oral Health","volume":"25 1","pages":"718"},"PeriodicalIF":2.6000,"publicationDate":"2025-05-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12079966/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"BMC Oral Health","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s12903-025-06126-8","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The light-curing unit (LCU) has become a vital piece of dental equipment that must be correctly maintained. This study investigated the impact of contamination and physical damage to the light tip on the power and radiant emittance values from old and new LCUs.

Materials and methods: Two investigators assessed 200 LCUs in dental clinics. The extent of contamination and physical damage to the light-curing unit (LCU) tips was recorded using a scale ranging from 0 to 8, where 0 indicates the absence of damage or contamination, and 8 represents severe damage or contamination. Then, the radiant emittance and power values of the LCU tip were measured using a digital radiometer (Bluephase meter II; Ivoclar, Schaan, Liechtenstein). LCUs that were more than five years old were classified as old. Spearman correlation coefficient was used to determine the relationship between the condition of the LCU and radiant emittance/power (p = 0.05).

Results: There were no significant differences in the percent reduction of the power and radiant emittance from the values reported by the manufacturers, as well as the presence of contamination or physical damage scores between old and relatively new light-curing tips (p > 0.05). The mean ± standard deviation percentage reductions in power and radiant emittance from the manufacturer's stated values were 19.2 ± 17.63% and 3.9 ± 16.49%, respectively. Contamination and physical damage had significant positive correlations with the reduction in the power (r = 0.22070, p = 0.0017 and r = 0.27422, p < 0.0001, respectively) and the reduction in the radiant emittance (r = 0.28626, p < 0.0001 and r = 0.36650, p < 0.0001). Increased contamination and physical damage scores corresponded to greater percent reductions in the power and radiant emittance (p < 0.05).

Conclusions: Contamination and physical damage to the LCU can negatively impact the light output from LCUs.

Clinical relevance: To ensure optimal performance, dentists should regularly monitor the output of their LCUs and examine the devices for any signs of physical damage or contamination.

损坏或污染对200个光固化单元尖端的影响。
目的:光固化装置(LCU)已成为牙科设备的重要组成部分,必须正确维护。研究了光尖污染和物理损伤对新旧lcu功率和辐射发射度的影响。材料与方法:两位研究者对200个牙科诊所的lcu进行了评估。光固化单元(LCU)尖端的污染程度和物理损坏程度使用从0到8的刻度进行记录,其中0表示没有损坏或污染,8表示严重损坏或污染。然后,利用数字辐射计(Bluephase meter II;Ivoclar, Schaan,列支敦士登)超过5年的lcu被归类为老的。采用Spearman相关系数确定LCU状态与辐射发射度/功率的关系(p = 0.05)。结果:新旧光固化针尖在功率和辐射发射率的降低百分比以及污染或物理损伤评分方面没有显著差异(p > 0.05)。功率和辐射度相对于制造商声明值的平均±标准偏差百分比分别为19.2±17.63%和3.9±16.49%。污染和物理损伤与功率降低呈显著正相关(r = 0.22070, p = 0.0017和r = 0.27422, p)。结论:污染和物理损伤会对LCU的光输出产生负面影响。临床相关性:为确保最佳性能,牙医应定期监测lcu的输出,并检查设备是否有物理损坏或污染的迹象。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
BMC Oral Health
BMC Oral Health DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
3.90
自引率
6.90%
发文量
481
审稿时长
6-12 weeks
期刊介绍: BMC Oral Health is an open access, peer-reviewed journal that considers articles on all aspects of the prevention, diagnosis and management of disorders of the mouth, teeth and gums, as well as related molecular genetics, pathophysiology, and epidemiology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信