Use of Three-Dimensional Surface Imaging to Measure Breast Volume in the Upright Position With Acceptable Accuracy.

IF 1.3 Q4 ONCOLOGY
Hironori Goto, Satsuki Fukumoto, Yukiko Hara, Keiichiro Tada
{"title":"Use of Three-Dimensional Surface Imaging to Measure Breast Volume in the Upright Position With Acceptable Accuracy.","authors":"Hironori Goto, Satsuki Fukumoto, Yukiko Hara, Keiichiro Tada","doi":"10.4274/ejbh.galenos.2025.2025-2-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The utility of three-dimensional surface imaging (3DSI) for measuring breast volume in the upright position has not been established.</p><p><strong>Materials and methods: </strong>First, the accuracy of 3DSI measurements was evaluated using plastic models with known breast volumes. Then, the breast volumes of 30 patients were measured using 3DSI in the upright position, computed tomography (CT) in the supine position, magnetic resonance imaging (MRI) in the prone position, and mammography (MMG) of the compressed breast. To determine the impact of 3DSI measurements, the correlation coefficients between 3DSI and CT, between MMG and CT, and between MRI and CT were calculated.</p><p><strong>Results: </strong>The accuracy of 3DSI measurements was confirmed using plastic models. The correlation coefficients between 3DSI and CT, between MRI and CT, and between MMG and CT were 0.83, 0.997, and 0.84, respectively. Although the breast volume measured by 3DSI was closely associated with that measured by CT, this correlation was weaker than that between the MRI- and CT-measured volumes and comparable with that between the MMG- and CT-measured volumes.</p><p><strong>Conclusion: </strong>3DSI can be used to measure breast volume in the upright position with clinically acceptable accuracy for the evaluation of cosmetic surgical outcomes.</p>","PeriodicalId":93996,"journal":{"name":"European journal of breast health","volume":" ","pages":"230-236"},"PeriodicalIF":1.3000,"publicationDate":"2025-06-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12180109/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European journal of breast health","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4274/ejbh.galenos.2025.2025-2-3","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/5/13 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The utility of three-dimensional surface imaging (3DSI) for measuring breast volume in the upright position has not been established.

Materials and methods: First, the accuracy of 3DSI measurements was evaluated using plastic models with known breast volumes. Then, the breast volumes of 30 patients were measured using 3DSI in the upright position, computed tomography (CT) in the supine position, magnetic resonance imaging (MRI) in the prone position, and mammography (MMG) of the compressed breast. To determine the impact of 3DSI measurements, the correlation coefficients between 3DSI and CT, between MMG and CT, and between MRI and CT were calculated.

Results: The accuracy of 3DSI measurements was confirmed using plastic models. The correlation coefficients between 3DSI and CT, between MRI and CT, and between MMG and CT were 0.83, 0.997, and 0.84, respectively. Although the breast volume measured by 3DSI was closely associated with that measured by CT, this correlation was weaker than that between the MRI- and CT-measured volumes and comparable with that between the MMG- and CT-measured volumes.

Conclusion: 3DSI can be used to measure breast volume in the upright position with clinically acceptable accuracy for the evaluation of cosmetic surgical outcomes.

使用三维表面成像测量直立位置的乳房体积,精度可接受。
目的:三维体表成像(3DSI)在竖直位乳房体积测量中的应用尚不明确。材料和方法:首先,使用已知乳房体积的塑料模型评估3DSI测量的准确性。然后,对30例患者采用直立位3DSI、仰卧位CT、俯卧位MRI和压缩乳房x线摄影(MMG)测量乳房体积。为了确定3DSI测量的影响,计算3DSI与CT、MMG与CT、MRI与CT之间的相关系数。结果:利用塑料模型证实了3DSI测量的准确性。3DSI与CT、MRI与CT、MMG与CT的相关系数分别为0.83、0.997、0.84。虽然3DSI测量的乳房体积与CT测量的体积密切相关,但这种相关性弱于MRI和CT测量的体积之间的相关性,与MMG和CT测量的体积之间的相关性相当。结论:3DSI可用于直立位乳房体积测量,准确度可用于评价美容手术效果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.60
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信