{"title":"Evaluation of HemoCue 201 Performance for Hemoglobin Measurement and Anemia Detection in EDTA, Heparinized, and Capillary Blood Samples","authors":"Orakan Limpornpugdee, Sithee Sodmanee, Panatda Sohab, Chitnuphong Saengnil, Narisorn Kongruttanachok","doi":"10.1111/ijlh.14495","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>This study aimed to evaluate the analytical and diagnostic performance of the HemoCue 201 device by comparing venous EDTA, venous heparinized, and capillary blood samples.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The HemoCue 201 device was compared with an automated analyzer. Hemoglobin levels were measured from EDTA, heparinized, and capillary blood samples. Analytical performance was assessed using mean ± SD, paired <i>t</i>-test, intraclass correlation coefficient (ICC), and Bland–Altman plots. The CLIA 2025 criteria for acceptable error were applied to determine agreement. Diagnostic performance was determined via sensitivity, specificity, diagnostic effectiveness, positive predictive value (PPV), negative predictive value (NPV), and likelihood ratios.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The results showed excellent reliability with ICC values of 1.00 for both EDTA and heparinized, and 0.97 for capillary blood. Bland–Altman plots indicated small biases for EDTA (0.08 [−0.27 to 0.42]) and heparin (0.09 [−0.40 to 0.57]), but larger negative biases for capillary blood (−0.12 [−1.21 to 0.97]). According to CLIA 2025 criteria, 98% of EDTA and 95% of heparin samples were within the ±4% cut-off, whereas only 68% of capillary samples met this criterion. Diagnostic performance demonstrated strong potential as a rapid and reliable anemia screening tool due to its high sensitivity (100%) and NPV (100%).</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The HemoCue 201 device reliably measured hemoglobin and detected anemia in all sample types. Heparinized blood, commonly used in point-of-care blood gas tests, can effectively support hemoglobin testing, reducing the need for additional blood draws. However, capillary samples exhibited greater bias, with variability influenced by pre-analytical factors. Standardized protocol and manufacturer guidelines helped minimize analytical variation.</p>\n </section>\n </div>","PeriodicalId":14120,"journal":{"name":"International Journal of Laboratory Hematology","volume":"47 5","pages":"817-825"},"PeriodicalIF":2.3000,"publicationDate":"2025-05-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Laboratory Hematology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/ijlh.14495","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
This study aimed to evaluate the analytical and diagnostic performance of the HemoCue 201 device by comparing venous EDTA, venous heparinized, and capillary blood samples.
Methods
The HemoCue 201 device was compared with an automated analyzer. Hemoglobin levels were measured from EDTA, heparinized, and capillary blood samples. Analytical performance was assessed using mean ± SD, paired t-test, intraclass correlation coefficient (ICC), and Bland–Altman plots. The CLIA 2025 criteria for acceptable error were applied to determine agreement. Diagnostic performance was determined via sensitivity, specificity, diagnostic effectiveness, positive predictive value (PPV), negative predictive value (NPV), and likelihood ratios.
Results
The results showed excellent reliability with ICC values of 1.00 for both EDTA and heparinized, and 0.97 for capillary blood. Bland–Altman plots indicated small biases for EDTA (0.08 [−0.27 to 0.42]) and heparin (0.09 [−0.40 to 0.57]), but larger negative biases for capillary blood (−0.12 [−1.21 to 0.97]). According to CLIA 2025 criteria, 98% of EDTA and 95% of heparin samples were within the ±4% cut-off, whereas only 68% of capillary samples met this criterion. Diagnostic performance demonstrated strong potential as a rapid and reliable anemia screening tool due to its high sensitivity (100%) and NPV (100%).
Conclusion
The HemoCue 201 device reliably measured hemoglobin and detected anemia in all sample types. Heparinized blood, commonly used in point-of-care blood gas tests, can effectively support hemoglobin testing, reducing the need for additional blood draws. However, capillary samples exhibited greater bias, with variability influenced by pre-analytical factors. Standardized protocol and manufacturer guidelines helped minimize analytical variation.
期刊介绍:
The International Journal of Laboratory Hematology provides a forum for the communication of new developments, research topics and the practice of laboratory haematology.
The journal publishes invited reviews, full length original articles, and correspondence.
The International Journal of Laboratory Hematology is the official journal of the International Society for Laboratory Hematology, which addresses the following sub-disciplines: cellular analysis, flow cytometry, haemostasis and thrombosis, molecular diagnostics, haematology informatics, haemoglobinopathies, point of care testing, standards and guidelines.
The journal was launched in 2006 as the successor to Clinical and Laboratory Hematology, which was first published in 1979. An active and positive editorial policy ensures that work of a high scientific standard is reported, in order to bridge the gap between practical and academic aspects of laboratory haematology.