{"title":"Reproducibility of the systolic and diastolic energy loss of the left ventricle in vector flow mapping.","authors":"Yuya Fukudome, Michinari Hieda, Nobuyuki Ohte, Shutaro Futami, Emi Kaneko, Akihiro Koike, Saki Kurokawa, Yuya Kunisaki, Akira Shiose, Koichi Akashi","doi":"10.1007/s12574-025-00688-w","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Vector Flow Mapping (VFM) is a novel technique for visualizing intracardiac blood flow. A few reports have been made regarding the reproducibility of energy loss (EL) measurements using VFM. This study aims to elucidate the intra-class and inter-class correlation coefficient (intra-ICC and inter-ICC) in the EL measurements.</p><p><strong>Methods: </strong>Seven healthy participants were enrolled, and echocardiographic data were obtained by two cardiac sonographers (14 images). Three independent analysts analyzed all images in three different cardiac cycles three times: 378 data points (14 images × analysts × 3 cardiac cycles × 3 times). The intra-ICC (1, 1) and inter-ICC (2, 1) were calculated. Furthermore, the intra-ICC (1, k) and inter-ICC (2, k) were applied using the averaged EL value in three cardiac cycles. An ICC value greater than 0.75 was defined as acceptable reproducibility.</p><p><strong>Results: </strong>In diastole, the intra-ICC (1, 1) by the three analysts was 0.890, 0.830, and 0.802; the intra-ICC (1, k), using the average EL value was 0.986, 0.978, and 0.973. In systole, the intra-ICC (1, 1) was 0.729, 0.698, and 0.733; the intra-ICC (1, k) was 0.960, 0.954, and 0.961. In diastole, the inter-ICC (2, 1) was 0.950, and the inter-ICC (2, k) was 0.958. In contrast, in systole, the inter-ICC (2, 1) was 0.772, and the inter-ICC (2, k) was 0.774; these values were lower than those in diastole.</p><p><strong>Conclusions: </strong>These findings indicate that the intra- and inter-measurer reproducibility of diastolic and systolic EL is favorable for clinical use.</p>","PeriodicalId":44837,"journal":{"name":"Journal of Echocardiography","volume":" ","pages":""},"PeriodicalIF":1.4000,"publicationDate":"2025-04-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Echocardiography","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12574-025-00688-w","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Vector Flow Mapping (VFM) is a novel technique for visualizing intracardiac blood flow. A few reports have been made regarding the reproducibility of energy loss (EL) measurements using VFM. This study aims to elucidate the intra-class and inter-class correlation coefficient (intra-ICC and inter-ICC) in the EL measurements.
Methods: Seven healthy participants were enrolled, and echocardiographic data were obtained by two cardiac sonographers (14 images). Three independent analysts analyzed all images in three different cardiac cycles three times: 378 data points (14 images × analysts × 3 cardiac cycles × 3 times). The intra-ICC (1, 1) and inter-ICC (2, 1) were calculated. Furthermore, the intra-ICC (1, k) and inter-ICC (2, k) were applied using the averaged EL value in three cardiac cycles. An ICC value greater than 0.75 was defined as acceptable reproducibility.
Results: In diastole, the intra-ICC (1, 1) by the three analysts was 0.890, 0.830, and 0.802; the intra-ICC (1, k), using the average EL value was 0.986, 0.978, and 0.973. In systole, the intra-ICC (1, 1) was 0.729, 0.698, and 0.733; the intra-ICC (1, k) was 0.960, 0.954, and 0.961. In diastole, the inter-ICC (2, 1) was 0.950, and the inter-ICC (2, k) was 0.958. In contrast, in systole, the inter-ICC (2, 1) was 0.772, and the inter-ICC (2, k) was 0.774; these values were lower than those in diastole.
Conclusions: These findings indicate that the intra- and inter-measurer reproducibility of diastolic and systolic EL is favorable for clinical use.
期刊介绍:
The Journal of Echocardiography, the official journal of the Japanese Society of Echocardiography, publishes work that contributes to progress in the field and articles in clinical research as well, seeking to develop a new focus and new perspectives for all who are concerned with this discipline. The journal welcomes original investigations, review articles, letters to the editor, editorials, and case image in cardiovascular ultrasound, which will be reviewed by the editorial board. The Journal of Echocardiography provides the best of up-to-date information from around the world, presenting readers with high-impact, original work focusing on pivotal issues.