Optical confirmation by a thick-lens formula of the optimized A-constant (as calculated by a thin-lens formula) of a new intraocular lens.

IF 2.6 3区 医学 Q2 OPHTHALMOLOGY
Giacomo Savini, Daniela Nicolosi, Kenneth J Hoffer, Catarina P Coutinho, Maria Cristina Curatolo, Domenico Schiano-Lomoriello
{"title":"Optical confirmation by a thick-lens formula of the optimized A-constant (as calculated by a thin-lens formula) of a new intraocular lens.","authors":"Giacomo Savini, Daniela Nicolosi, Kenneth J Hoffer, Catarina P Coutinho, Maria Cristina Curatolo, Domenico Schiano-Lomoriello","doi":"10.1097/j.jcrs.0000000000001671","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To provide a physical explanation for the A-constant of a new enhanced monofocal intraocular lens (IOL) by evaluating the relationship between the postoperative anterior chamber depth constant (ACD-constant) predicted by the Olsen formula, the ACD predicted by the Holladay formula and the postoperative ACD measured by anterior segment optical coherence tomography (AS-OCT).</p><p><strong>Setting: </strong>IRCCS Bietti Foundation, Rome, Italy.</p><p><strong>Design: </strong>Retrospective case series.</p><p><strong>Methods: </strong>In a series of consecutive patients that underwent cataract surgery with the same enhanced monofocal IOL, the A-constant of the SRK/T formula was optimized. The postoperative distance between the corneal epithelium and the anterior surface of the IOL was measured by AS-OCT (MS-39, CSO) and compared to the ACD predicted by the Olsen formula (PhacoOptics) and the method described by Holladay to correlate the thin-lens IOL position to the thick-lens IOL position.</p><p><strong>Results: </strong>Twenty-one eyes were investigated. The back-calculated optimized A-constant was 118.33. The mean measured IOL distance from the corneal epithelium was 4.24 ± 0.28 mm. Based on the A-constant value of 118.33, PhacoOptics calculated that the mean predicted ACD constant specific for this IOL was 4.28 mm, while the mean predicted ACD based on the individual preoperative biometric parameters was 4.17 ± 0.23 mm. According to the Holladay method, the IOL distance from the corneal vertex was 4.21 ± 0.36 mm. Repeated measures ANOVA disclosed that the difference among the 3 values was not significant (p = 0.3535).</p><p><strong>Conclusions: </strong>The optimized A-constant correlates well with the thick-lens values calculated for the postoperative ACD.</p>","PeriodicalId":15214,"journal":{"name":"Journal of cataract and refractive surgery","volume":" ","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2025-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cataract and refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/j.jcrs.0000000000001671","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To provide a physical explanation for the A-constant of a new enhanced monofocal intraocular lens (IOL) by evaluating the relationship between the postoperative anterior chamber depth constant (ACD-constant) predicted by the Olsen formula, the ACD predicted by the Holladay formula and the postoperative ACD measured by anterior segment optical coherence tomography (AS-OCT).

Setting: IRCCS Bietti Foundation, Rome, Italy.

Design: Retrospective case series.

Methods: In a series of consecutive patients that underwent cataract surgery with the same enhanced monofocal IOL, the A-constant of the SRK/T formula was optimized. The postoperative distance between the corneal epithelium and the anterior surface of the IOL was measured by AS-OCT (MS-39, CSO) and compared to the ACD predicted by the Olsen formula (PhacoOptics) and the method described by Holladay to correlate the thin-lens IOL position to the thick-lens IOL position.

Results: Twenty-one eyes were investigated. The back-calculated optimized A-constant was 118.33. The mean measured IOL distance from the corneal epithelium was 4.24 ± 0.28 mm. Based on the A-constant value of 118.33, PhacoOptics calculated that the mean predicted ACD constant specific for this IOL was 4.28 mm, while the mean predicted ACD based on the individual preoperative biometric parameters was 4.17 ± 0.23 mm. According to the Holladay method, the IOL distance from the corneal vertex was 4.21 ± 0.36 mm. Repeated measures ANOVA disclosed that the difference among the 3 values was not significant (p = 0.3535).

Conclusions: The optimized A-constant correlates well with the thick-lens values calculated for the postoperative ACD.

用厚晶状体公式确定新型人工晶状体的优化a常数(用薄晶状体公式计算)。
目的:通过评价Olsen公式预测的术后前房深度常数(ACD-constant)、Holladay公式预测的ACD与术后前段光学相干断层扫描(AS-OCT)测量的ACD之间的关系,为新型增强单焦点人工晶状体(IOL)的a常数提供物理解释。地点:意大利罗马IRCCS Bietti基金会。设计:回顾性病例系列。方法:对连续行同一增强单焦点人工晶状体白内障手术的患者,优化SRK/T配方的a常数。采用AS-OCT (MS-39, CSO)测量术后角膜上皮与人工晶状体前表面之间的距离,并与Olsen公式(PhacoOptics)预测的ACD和Holladay描述的薄晶状体IOL位置与厚晶状体IOL位置相关的方法进行比较。结果:检查21只眼。反算优化a常数为118.33。人工晶状体距角膜上皮的平均距离为4.24±0.28 mm。基于a常数值118.33,PhacoOptics计算出该IOL特异性的预测ACD常数平均值为4.28 mm,而基于个体术前生物特征参数的预测ACD平均值为4.17±0.23 mm。Holladay法人工晶状体距角膜顶点的距离为4.21±0.36 mm。重复测量方差分析显示,3个值之间的差异不显著(p = 0.3535)。结论:优化后的a常数与术后ACD计算的厚晶状体值具有良好的相关性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.60
自引率
14.30%
发文量
259
审稿时长
8.5 weeks
期刊介绍: The Journal of Cataract & Refractive Surgery (JCRS), a preeminent peer-reviewed monthly ophthalmology publication, is the official journal of the American Society of Cataract and Refractive Surgery (ASCRS) and the European Society of Cataract and Refractive Surgeons (ESCRS). JCRS publishes high quality articles on all aspects of anterior segment surgery. In addition to original clinical studies, the journal features a consultation section, practical techniques, important cases, and reviews as well as basic science articles.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信