[MSB-42] Impact of Phosphorylcholine-Coated Circuits on Inflammatory Response and Renal Function: A Retrospective Study in Coronary Artery Bypass Grafting.
{"title":"[MSB-42] Impact of Phosphorylcholine-Coated Circuits on Inflammatory Response and Renal Function: A Retrospective Study in Coronary Artery Bypass Grafting.","authors":"Ercan Kahraman, Mehmet Niyaz, Şirin Çetin","doi":"10.5606/tgkdc.dergisi.2024.msb-42","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study aimed to evaluate the effects of phosphorylcholine-coated circuits compared to traditional standard tubing sets on postoperative inflammatory response, renal function, and intubation duration.</p><p><strong>Methods: </strong>This retrospective study was conducted with 45 patients who underwent coronary artery bypass grafting. Surgeries were performed by the same team. The patients were divided into two groups: standard tubing sets were used for Group 1 (n=12), and phosphorylcholine-coated sets were used for Group 2 (n=33). Data from the operation were analyzed.</p><p><strong>Results: </strong>No significant difference was found between standard and phosphorylcholine-coated tubing sets regarding lactate levels after bypass and in intensive care (p>0.05). C-reactive protein levels were similar on the first day but lower in the phosphorylcholine group on the second day (p<0.05). Preoperative glomerular filtration rates were similar on the first day, but higher glomerular filtration rates were noted on the second day in the phosphorylcholine group (p<0.05). Intubation duration was shorter with phosphorylcholine-coated circuits, with no differences in erythrocyte suspension and drainage volumes (p>0.05).</p><p><strong>Conclusion: </strong>Phosphorylcholine-coated circuits reduced postoperative inflammatory response, preserved renal function, and shortened intubation duration. Although coated systems may improve cardiac surgery outcomes, no system completely prevents pathological inflammatory responses. Collaboration between biomedical engineering and cardiac surgery teams is essential for further advancements.</p>","PeriodicalId":49413,"journal":{"name":"Turk Gogus Kalp Damar Cerrahisi Dergisi-Turkish Journal of Thoracic and Cardiovascular Surgery","volume":"32 4 Suppl 2","pages":"067-67"},"PeriodicalIF":0.5000,"publicationDate":"2024-12-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12045224/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Turk Gogus Kalp Damar Cerrahisi Dergisi-Turkish Journal of Thoracic and Cardiovascular Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5606/tgkdc.dergisi.2024.msb-42","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"SURGERY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: This study aimed to evaluate the effects of phosphorylcholine-coated circuits compared to traditional standard tubing sets on postoperative inflammatory response, renal function, and intubation duration.
Methods: This retrospective study was conducted with 45 patients who underwent coronary artery bypass grafting. Surgeries were performed by the same team. The patients were divided into two groups: standard tubing sets were used for Group 1 (n=12), and phosphorylcholine-coated sets were used for Group 2 (n=33). Data from the operation were analyzed.
Results: No significant difference was found between standard and phosphorylcholine-coated tubing sets regarding lactate levels after bypass and in intensive care (p>0.05). C-reactive protein levels were similar on the first day but lower in the phosphorylcholine group on the second day (p<0.05). Preoperative glomerular filtration rates were similar on the first day, but higher glomerular filtration rates were noted on the second day in the phosphorylcholine group (p<0.05). Intubation duration was shorter with phosphorylcholine-coated circuits, with no differences in erythrocyte suspension and drainage volumes (p>0.05).
Conclusion: Phosphorylcholine-coated circuits reduced postoperative inflammatory response, preserved renal function, and shortened intubation duration. Although coated systems may improve cardiac surgery outcomes, no system completely prevents pathological inflammatory responses. Collaboration between biomedical engineering and cardiac surgery teams is essential for further advancements.
期刊介绍:
The Turkish Journal of Thoracic and Cardiovascular Surgery is an international open access journal which publishes original articles on topics in generality of Cardiac, Thoracic, Arterial, Venous, Lymphatic Disorders and their managements. These encompass all relevant clinical, surgical and experimental studies, editorials, current and collective reviews, technical know-how papers, case reports, interesting images, How to Do It papers, correspondences, and commentaries.