{"title":"Comparison of Anterior Segment Measurements With a New Swept-source Optical Coherence Tomography Biometer and a Scheimpflug-Placido Topographer.","authors":"Chak Seng Lei, Xin Li, Huilin Xu, Rui Ning, Yiran Wang, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Kexin Li, Jinhai Huang","doi":"10.3928/1081597X-20250317-03","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To assess the agreement of anterior segment measurements between a new swept-source optical coherence tomography biometer (ZW-30; TowardPi Medical Technology) and a Scheimpflug-Placido topographer (Sirius; Costruzione Strumenti Oftalmici) in normal eyes.</p><p><strong>Methods: </strong>In this prospective study, measurement agreement was assessed in 56 right eyes of 56 healthy patients. One experienced operator obtained three consecutive measurements with both devices randomly. The comparative parameters were central corneal thickness (CCT), anterior aqueous depth (AQD), anterior chamber depth (ACD), mean keratometry, astigmatism magnitude (AST), J<sub>0</sub> and J<sub>45</sub> power vectors, and corneal diameter (CD). The paired <i>t</i>-test was used to compare the difference between the two devices. Pearson's correlation analysis and Bland-Altman plots with 95% limits of agreement (LoA) were calculated to evaluate agreement.</p><p><strong>Results: </strong>The paired <i>t</i>-test showed statistically significant differences in all parameters (<i>P</i> < .05). Correlation analysis showed high correlation, Bland-Altman plots indicated moderate to high agreement between the two devices. The 95% LoA of CCT, AQD, ACD, mean keratometry, AST, J<sub>0</sub> and J<sub>45</sub> power vectors, and CD ranged from -4.36 to 19.83 µm, -0.11 to 0.04 mm, -0.12 to 0.03 mm, -0.50 to 0.12 D, -0.29 to 0.39 D, -0.20 to 0.14 D, -0.17 to 0.13 D, and 0.18 to 0.80 mm, respectively.</p><p><strong>Conclusions: </strong>Anterior segment measurements from the new SS-OCT biometer and the Scheimpflug-Placido topographer demonstrated high agreement in AQD, ACD, AST, and J<sub>0</sub> and J<sub>45</sub> power vectors. However, for CCT, mean keratometry, and CD, the moderate inter-device agreement suggests that results should be interpreted with caution when making comparisons. <b>[<i>J Refract Surg</i>. 2025;41(5):e413-e420.]</b>.</p>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"41 5","pages":"e413-e420"},"PeriodicalIF":2.9000,"publicationDate":"2025-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597X-20250317-03","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To assess the agreement of anterior segment measurements between a new swept-source optical coherence tomography biometer (ZW-30; TowardPi Medical Technology) and a Scheimpflug-Placido topographer (Sirius; Costruzione Strumenti Oftalmici) in normal eyes.
Methods: In this prospective study, measurement agreement was assessed in 56 right eyes of 56 healthy patients. One experienced operator obtained three consecutive measurements with both devices randomly. The comparative parameters were central corneal thickness (CCT), anterior aqueous depth (AQD), anterior chamber depth (ACD), mean keratometry, astigmatism magnitude (AST), J0 and J45 power vectors, and corneal diameter (CD). The paired t-test was used to compare the difference between the two devices. Pearson's correlation analysis and Bland-Altman plots with 95% limits of agreement (LoA) were calculated to evaluate agreement.
Results: The paired t-test showed statistically significant differences in all parameters (P < .05). Correlation analysis showed high correlation, Bland-Altman plots indicated moderate to high agreement between the two devices. The 95% LoA of CCT, AQD, ACD, mean keratometry, AST, J0 and J45 power vectors, and CD ranged from -4.36 to 19.83 µm, -0.11 to 0.04 mm, -0.12 to 0.03 mm, -0.50 to 0.12 D, -0.29 to 0.39 D, -0.20 to 0.14 D, -0.17 to 0.13 D, and 0.18 to 0.80 mm, respectively.
Conclusions: Anterior segment measurements from the new SS-OCT biometer and the Scheimpflug-Placido topographer demonstrated high agreement in AQD, ACD, AST, and J0 and J45 power vectors. However, for CCT, mean keratometry, and CD, the moderate inter-device agreement suggests that results should be interpreted with caution when making comparisons. [J Refract Surg. 2025;41(5):e413-e420.].
期刊介绍:
The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as:
• Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics”
• Supplemental videos and materials available for many articles
• Access to current articles, as well as several years of archived content
• Articles posted online just 2 months after acceptance.