First clinical application of Comprehensive Motion Management with 1.5T MR-linac on prostate cancer patients treated with radiotherapy: technical aspects and suggested workflow recommendations.
IF 9.7 1区 医学Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Luca Nicosia, Michele Rigo, Roberto Giuseppe Pellegrini, Andrea Gaetano Allegra, Chiara De-Colle, Niccolò Giaj-Levra, Edoardo Pastorello, Francesco Ricchetti, Ruggero Ruggieri, Filippo Alongi
{"title":"First clinical application of Comprehensive Motion Management with 1.5T MR-linac on prostate cancer patients treated with radiotherapy: technical aspects and suggested workflow recommendations.","authors":"Luca Nicosia, Michele Rigo, Roberto Giuseppe Pellegrini, Andrea Gaetano Allegra, Chiara De-Colle, Niccolò Giaj-Levra, Edoardo Pastorello, Francesco Ricchetti, Ruggero Ruggieri, Filippo Alongi","doi":"10.1007/s11547-025-02014-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>MR-guided radiotherapy (MRgRT) on the MR-linac (MRL) with daily online plan adaptation enables better control of inter-fraction variability. Recently, Comprehensive Motion Management (CMM) was introduced for the 1.5T MRL. CMM halts dose delivery when the target moves outside its defined position and allows for target drift corrections, compensating for intra-fraction variability. This study aims to report the first clinical experience with prostate MRgRT using CMM, focusing on intra-fraction motion management and treatment delivery time.</p><p><strong>Material and methods: </strong>Sixty patients with low- to intermediate-risk prostate cancer were treated with the 1.5T MRL using CMM. PTV margins were 5 mm in all directions, with a 3-mm margin posteriorly. Fifty patients received ultra-hypofractionated radiotherapy (SBRT), and 10 received hypofractionated radiotherapy (hypoRT). The CMM threshold was set to 100% of the GTV volume.</p><p><strong>Results: </strong>A total of 450 treatment fractions were administered. The median beam-on time was 10.3 min, with a median duty cycle of 98.9% (95% CI 98.6-99.2%). Beam hold occurred in 158 fractions (35%), with an average of 24.5 instances per fraction (95% CI 16-32). Thirty-two baseline shift replans were performed. Both acute and late toxicities were low, with no Grade 3 or higher toxicities.</p><p><strong>Conclusions: </strong>CMM has been successfully implemented in MRgRT. In-treatment corrections and baseline shift replanning did not significantly impact treatment time, enhancing treatment quality without compromising patient compliance or treatment feasibility. The low incidence of beam-hold events suggests that CMM could potentially allow for a safe reduction in PTV margins.</p>","PeriodicalId":20817,"journal":{"name":"Radiologia Medica","volume":" ","pages":""},"PeriodicalIF":9.7000,"publicationDate":"2025-04-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiologia Medica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s11547-025-02014-x","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: MR-guided radiotherapy (MRgRT) on the MR-linac (MRL) with daily online plan adaptation enables better control of inter-fraction variability. Recently, Comprehensive Motion Management (CMM) was introduced for the 1.5T MRL. CMM halts dose delivery when the target moves outside its defined position and allows for target drift corrections, compensating for intra-fraction variability. This study aims to report the first clinical experience with prostate MRgRT using CMM, focusing on intra-fraction motion management and treatment delivery time.
Material and methods: Sixty patients with low- to intermediate-risk prostate cancer were treated with the 1.5T MRL using CMM. PTV margins were 5 mm in all directions, with a 3-mm margin posteriorly. Fifty patients received ultra-hypofractionated radiotherapy (SBRT), and 10 received hypofractionated radiotherapy (hypoRT). The CMM threshold was set to 100% of the GTV volume.
Results: A total of 450 treatment fractions were administered. The median beam-on time was 10.3 min, with a median duty cycle of 98.9% (95% CI 98.6-99.2%). Beam hold occurred in 158 fractions (35%), with an average of 24.5 instances per fraction (95% CI 16-32). Thirty-two baseline shift replans were performed. Both acute and late toxicities were low, with no Grade 3 or higher toxicities.
Conclusions: CMM has been successfully implemented in MRgRT. In-treatment corrections and baseline shift replanning did not significantly impact treatment time, enhancing treatment quality without compromising patient compliance or treatment feasibility. The low incidence of beam-hold events suggests that CMM could potentially allow for a safe reduction in PTV margins.
期刊介绍:
Felice Perussia founded La radiologia medica in 1914. It is a peer-reviewed journal and serves as the official journal of the Italian Society of Medical and Interventional Radiology (SIRM). The primary purpose of the journal is to disseminate information related to Radiology, especially advancements in diagnostic imaging and related disciplines. La radiologia medica welcomes original research on both fundamental and clinical aspects of modern radiology, with a particular focus on diagnostic and interventional imaging techniques. It also covers topics such as radiotherapy, nuclear medicine, radiobiology, health physics, and artificial intelligence in the context of clinical implications. The journal includes various types of contributions such as original articles, review articles, editorials, short reports, and letters to the editor. With an esteemed Editorial Board and a selection of insightful reports, the journal is an indispensable resource for radiologists and professionals in related fields. Ultimately, La radiologia medica aims to serve as a platform for international collaboration and knowledge sharing within the radiological community.