{"title":"Validation assessment of a cuff-based central blood pressure device according to Artery Society recommendations.","authors":"Tan V Bui, Dean S Picone, Wojciech P Kosmala, Michał J Kosowski, Jowita Zachwyc, Giacomo Pucci, Alessio Arrivi, Rosa Curcio, Martin G Schultz, J Andrew Black, Nathan Dwyer, Philip Roberts-Thomson, Sarang Paleri, Heath Adams, James E Sharman","doi":"10.1097/MBP.0000000000000754","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Uscom BP+ is a cuff-based blood pressure (BP) device designed to noninvasively estimate central BP as distinct from conventional brachial BP. This study aimed to assess the accuracy of the Uscom BP+ device compared with invasively measured BP.</p><p><strong>Methods: </strong>Automated noninvasive cuff central BP (using the Uscom BP+ device) and invasive central aortic BP were recorded simultaneously in 191 participants (65% male, aged 66 ± 11 years) receiving coronary angiography at three independent research sites in Australia, Poland, and Italy. Validation procedures were undertaken according to the Artery Society recommendations and with a minimally acceptable error (mean ± SD) of ≤5 ± ≤8 mmHg as pass criteria.</p><p><strong>Results: </strong>Using the device default calibration technique [brachial cuff systolic blood pressure (SBP) and diastolic blood pressure (DBP)], cuff central SBP underestimated invasive central SBP [mean (SD) difference: -10.2 (11.2) mmHg] and cuff central DBP overestimated invasive central DBP [mean (SD) difference: 9.8 (8.5) mmHg]. When calibrating by brachial cuff mean arterial pressure and DBP, SBP accuracy was improved, but variability remained high [mean (SD) difference: -6.3 (14.4) mmHg, P = 0.004 vs. default calibration, whereas DBP accuracy and variability remained similar [mean (SD) difference: 10.9 (8.5) mmHg, P = 0.19 vs. default calibration].</p><p><strong>Conclusion: </strong>The Uscom BP+ cuff device does not pass the Artery Society accuracy criteria compared with invasively measured central BP.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"181-186"},"PeriodicalIF":1.2000,"publicationDate":"2025-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000754","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/5/6 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Uscom BP+ is a cuff-based blood pressure (BP) device designed to noninvasively estimate central BP as distinct from conventional brachial BP. This study aimed to assess the accuracy of the Uscom BP+ device compared with invasively measured BP.
Methods: Automated noninvasive cuff central BP (using the Uscom BP+ device) and invasive central aortic BP were recorded simultaneously in 191 participants (65% male, aged 66 ± 11 years) receiving coronary angiography at three independent research sites in Australia, Poland, and Italy. Validation procedures were undertaken according to the Artery Society recommendations and with a minimally acceptable error (mean ± SD) of ≤5 ± ≤8 mmHg as pass criteria.
Results: Using the device default calibration technique [brachial cuff systolic blood pressure (SBP) and diastolic blood pressure (DBP)], cuff central SBP underestimated invasive central SBP [mean (SD) difference: -10.2 (11.2) mmHg] and cuff central DBP overestimated invasive central DBP [mean (SD) difference: 9.8 (8.5) mmHg]. When calibrating by brachial cuff mean arterial pressure and DBP, SBP accuracy was improved, but variability remained high [mean (SD) difference: -6.3 (14.4) mmHg, P = 0.004 vs. default calibration, whereas DBP accuracy and variability remained similar [mean (SD) difference: 10.9 (8.5) mmHg, P = 0.19 vs. default calibration].
Conclusion: The Uscom BP+ cuff device does not pass the Artery Society accuracy criteria compared with invasively measured central BP.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.