Optical and Visual Outcomes of a New Refractive Extended Depth of Focus Intraocular Lens.

IF 2.9 3区 医学 Q1 OPHTHALMOLOGY
Belén Alfonso-Bartolozzi, Irene Martinez-Alberquilla, Luis Fernández-Vega-Cueto, Fátima Cuellar, Fidel Vega, María S Millán, David Madrid-Costa, José F Alfonso
{"title":"Optical and Visual Outcomes of a New Refractive Extended Depth of Focus Intraocular Lens.","authors":"Belén Alfonso-Bartolozzi, Irene Martinez-Alberquilla, Luis Fernández-Vega-Cueto, Fátima Cuellar, Fidel Vega, María S Millán, David Madrid-Costa, José F Alfonso","doi":"10.3928/1081597X-20250221-02","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To characterize the optical design and analyze the optical quality and halo induced by the new extended depth of focus (EDOF) TECNIS PureSee intraocular lens (IOL) (Johnson & Johnson Surgical Vision) and to evaluate its clinical performance.</p><p><strong>Methods: </strong>The power profile of the TECNIS PureSee was measured with the NIMO TR1504 device (LAMBDA-X). The through-focus modulation transfer function area (TF-MTFa) metric from +2.00 to -4.00 diopters (D) and the halo induced by the IOL in the best focus plane for distance vision (0.00 D), were assessed for 2-, 3-, and 4.5-mm aperture diameters at the IOL plane. The clinical evaluation in a cohort of patients included postoperative refraction, visual acuities, defocus curve, and subjective symptomatology.</p><p><strong>Results: </strong>The power profile shows distinct zones that meet relatively smoothly. The largest power change occurs in the central zone of the lens. The TF-MTFa curves showed an increase in depth of focus for smaller pupils. Larger pupils resulted in smaller, lower-intensity halos. For 2-, 3-, and 4.5-mm pupils, halo sizes were 18.64, 14.73, and 13.23 arcmin, respectively, and normalized energies were 0.55, 0.46, and 0.42, respectively. Clinical assessment yielded excellent visual and refractive outcomes, a depth of field from +1.00 to -2.00 D with visual acuities better than 0.2 logarithm of the minimum angle of resolution, and with almost no dysphotopic symptoms.</p><p><strong>Conclusions: </strong>The TECNIS PureSee IOL provided good distance optical and visual quality and an extended depth of field of approximately 2.00 D. The depth of focus increased for smaller pupils. The halo induced by the lens was of low size and energy. <b>[<i>J Refract Surg</i>. 2025;41(4):e333-e341.]</b>.</p>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"41 4","pages":"e333-e341"},"PeriodicalIF":2.9000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597X-20250221-02","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To characterize the optical design and analyze the optical quality and halo induced by the new extended depth of focus (EDOF) TECNIS PureSee intraocular lens (IOL) (Johnson & Johnson Surgical Vision) and to evaluate its clinical performance.

Methods: The power profile of the TECNIS PureSee was measured with the NIMO TR1504 device (LAMBDA-X). The through-focus modulation transfer function area (TF-MTFa) metric from +2.00 to -4.00 diopters (D) and the halo induced by the IOL in the best focus plane for distance vision (0.00 D), were assessed for 2-, 3-, and 4.5-mm aperture diameters at the IOL plane. The clinical evaluation in a cohort of patients included postoperative refraction, visual acuities, defocus curve, and subjective symptomatology.

Results: The power profile shows distinct zones that meet relatively smoothly. The largest power change occurs in the central zone of the lens. The TF-MTFa curves showed an increase in depth of focus for smaller pupils. Larger pupils resulted in smaller, lower-intensity halos. For 2-, 3-, and 4.5-mm pupils, halo sizes were 18.64, 14.73, and 13.23 arcmin, respectively, and normalized energies were 0.55, 0.46, and 0.42, respectively. Clinical assessment yielded excellent visual and refractive outcomes, a depth of field from +1.00 to -2.00 D with visual acuities better than 0.2 logarithm of the minimum angle of resolution, and with almost no dysphotopic symptoms.

Conclusions: The TECNIS PureSee IOL provided good distance optical and visual quality and an extended depth of field of approximately 2.00 D. The depth of focus increased for smaller pupils. The halo induced by the lens was of low size and energy. [J Refract Surg. 2025;41(4):e333-e341.].

一种新型屈光扩展焦深度人工晶状体的光学和视觉效果。
目的:对新型扩展焦深(EDOF) TECNIS PureSee人工晶状体(IOL) (Johnson & Johnson Surgical Vision)的光学设计进行表征,分析其光学质量和光晕,评价其临床性能。方法:采用NIMO TR1504装置(LAMBDA-X)测量TECNIS PureSee的功率分布。通过对焦调制传递函数面积(TF-MTFa)从+2.00到-4.00屈光度(D)和最佳焦距平面(0.00 D)的IOL在2-,3-和4.5 mm孔径的IOL引起的晕。一组患者的临床评价包括术后屈光、视力、离焦曲线和主观症状。结果:功率剖面有明显的交汇区域,相对平滑。最大的功率变化发生在透镜的中心区域。TF-MTFa曲线显示小瞳孔的聚焦深度增加。瞳孔越大,光晕越小,强度越低。对于2毫米、3毫米和4.5毫米瞳孔,光晕大小分别为18.64、14.73和13.23弧分,归一化能量分别为0.55、0.46和0.42。临床评估获得了良好的视力和屈光效果,景深在+1.00到-2.00 D之间,视力优于最小分辨率的0.2对数,几乎没有光线不良症状。结论:TECNIS PureSee人工晶状体提供了良好的距离光学和视觉质量,扩展的景深约为2.00 d,小瞳孔的焦点深度增加。透镜产生的光晕尺寸小,能量小。[J].中国光学精密工程,2015;41(4):393 - 391。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.10
自引率
12.50%
发文量
160
审稿时长
4-8 weeks
期刊介绍: The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as: • Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics” • Supplemental videos and materials available for many articles • Access to current articles, as well as several years of archived content • Articles posted online just 2 months after acceptance.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信