{"title":"Long-term Titanium mesh cage implant failure after total en bloc spondylectomy due to spinal chondrosarcoma","authors":"Ezequiel Martinez, Matias Pereira Duarte, Gonzalo Kido, Gaston Camino-Willhuber, Julio Bassani, Marcelo Gruenberg, Matias Petracchi","doi":"10.31053/1853.0605.v82.n1.44961","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To report an unusual titanium mesh cage failure after ten years of follow-up.</p><p><strong>Clinical case: </strong>A 40-year-old woman diagnosed with low-grade spinal chondrosarcoma underwent a total en-bloc spondylenctomy (TES) surgery. This technique required a wide resection, anterior reconstruction with a titanium mesh cage (TMC), and posterior stabilization. For several years a favorable clinical outcome was achieved. After 10 years of follow-up, the patient developed acute non-traumatic low back pain without neurological repercussions. The images confirmed a rupture of the titanium cage. After ruling out tumor recurrence, the patient went through spinal instrumentation revision.</p><p><strong>Conclusion: </strong>Isolated TMC rupture is a rare event scarcely reported in the literature. There is a strong recommendation to instrument two levels proximal and distal to the bone defect in 360° reconstructions after a TES. Stability had likely been lost over the years due to the slow resorption of the bone graft. We consider necessary studies with a higher casuistry and level of evidence to analyze the contributing factors for the event of TMC rupture.</p>","PeriodicalId":38814,"journal":{"name":"Revista de la Facultad de Ciencias Medicas de Cordoba","volume":"82 1","pages":"140-153"},"PeriodicalIF":0.0000,"publicationDate":"2025-03-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista de la Facultad de Ciencias Medicas de Cordoba","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31053/1853.0605.v82.n1.44961","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To report an unusual titanium mesh cage failure after ten years of follow-up.
Clinical case: A 40-year-old woman diagnosed with low-grade spinal chondrosarcoma underwent a total en-bloc spondylenctomy (TES) surgery. This technique required a wide resection, anterior reconstruction with a titanium mesh cage (TMC), and posterior stabilization. For several years a favorable clinical outcome was achieved. After 10 years of follow-up, the patient developed acute non-traumatic low back pain without neurological repercussions. The images confirmed a rupture of the titanium cage. After ruling out tumor recurrence, the patient went through spinal instrumentation revision.
Conclusion: Isolated TMC rupture is a rare event scarcely reported in the literature. There is a strong recommendation to instrument two levels proximal and distal to the bone defect in 360° reconstructions after a TES. Stability had likely been lost over the years due to the slow resorption of the bone graft. We consider necessary studies with a higher casuistry and level of evidence to analyze the contributing factors for the event of TMC rupture.
期刊介绍:
The Journal of the Faculty of Medical Sciences is a scientific publication of the Secretariat of Science and Technology of the Faculty of Medical Sciences of the National University of Cordoba. Its objective is to disseminate and promote research work related to Medical and Biological Sciences. It publishes scientific works of national and international professionals on different topics related to health sciences from the field of medicine, nursing, kinesiology, diagnostic imaging, phonoaudiology, nutrition, public health, chemical sciences, dentistry and related.