Measurement uncertainty of the absorbed dose from fabricated Ge-doped optical fibres and nanoDot OSLD in radiotherapy electron beam dosimetry audit

IF 2.8 3区 物理与天体物理 Q3 CHEMISTRY, PHYSICAL
N. Abdullah , N. Mohd Noor , Z. Kamarul Zaman , M. Mohammad Zahid
{"title":"Measurement uncertainty of the absorbed dose from fabricated Ge-doped optical fibres and nanoDot OSLD in radiotherapy electron beam dosimetry audit","authors":"N. Abdullah ,&nbsp;N. Mohd Noor ,&nbsp;Z. Kamarul Zaman ,&nbsp;M. Mohammad Zahid","doi":"10.1016/j.radphyschem.2025.112684","DOIUrl":null,"url":null,"abstract":"<div><div>The use of transfer dosimeters to validate the accuracy of absorbed dose delivery during radiotherapy dosimetry audits requires estimation of the measurement uncertainty to ensure that the tolerance limits are met. Previous studies have explored the measurement uncertainties in absorbed dose for both Germanium-doped optical fibre (GedOF) and nanoDot optically stimulated luminescence (OSL) dosimetry systems in radiotherapy dosimetry audits involving photon and electron beams. However, there is a paucity of research comparing the detailed measurement uncertainty between GedOF and OSLD, particularly for radiotherapy electron beam dosimetry audits. In this study, the measurement uncertainties in the analysis of the absorbed dose measured from two types of GedOF in cylindrical (CGedOF) and flat (FGedOF) shapes were investigated in comparison with nanoDot OSL dosimeter (OSLD). The uncertainty was derived from two components: (i) the calibration of the dosimeters against the ionization chamber in electron beams based on the IAEA's TRS No. 398, and (ii) the calculation of the absorbed dose to water for each dosimeter, which includes the correction factors of response sensitivities, dose-response linearity, signal fading, and beam energies. Each source of uncertainty was expressed as a relative standard uncertainty, which was classified as type A (random) or type B (systematic). The total combined standard uncertainty was calculated by summing the Type A and Type B uncertainties using a quadratic method. The largest relative standard uncertainty in the absorbed dose measurement from each dosimeter arose from the calibration coefficient of these dosimetry systems (1.32 %–1.36 %), followed by the stability of the TLD/OSLD reader long term stability (0.47 %–0.52 %) and statistical TL/OSL readings (0.38 %–0.59 %). The quadratic summation of all relative standard uncertainties resulted in total combined uncertainties of 1.31 % for CGedOF and 1.36 % for both FGedOF and OSLD, resulting in tolerance limits of 3.93 % for CGedOF and 4.08 % for both FGedOF and OSLD. In conclusion, this study provides valuable insights into the measurement uncertainties of both CGedOF and FGedOF in radiotherapy applications, highlighting their potential as reliable tools for dose verification in remote dosimetry audits.</div></div>","PeriodicalId":20861,"journal":{"name":"Radiation Physics and Chemistry","volume":"234 ","pages":"Article 112684"},"PeriodicalIF":2.8000,"publicationDate":"2025-03-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation Physics and Chemistry","FirstCategoryId":"92","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0969806X25001768","RegionNum":3,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, PHYSICAL","Score":null,"Total":0}
引用次数: 0

Abstract

The use of transfer dosimeters to validate the accuracy of absorbed dose delivery during radiotherapy dosimetry audits requires estimation of the measurement uncertainty to ensure that the tolerance limits are met. Previous studies have explored the measurement uncertainties in absorbed dose for both Germanium-doped optical fibre (GedOF) and nanoDot optically stimulated luminescence (OSL) dosimetry systems in radiotherapy dosimetry audits involving photon and electron beams. However, there is a paucity of research comparing the detailed measurement uncertainty between GedOF and OSLD, particularly for radiotherapy electron beam dosimetry audits. In this study, the measurement uncertainties in the analysis of the absorbed dose measured from two types of GedOF in cylindrical (CGedOF) and flat (FGedOF) shapes were investigated in comparison with nanoDot OSL dosimeter (OSLD). The uncertainty was derived from two components: (i) the calibration of the dosimeters against the ionization chamber in electron beams based on the IAEA's TRS No. 398, and (ii) the calculation of the absorbed dose to water for each dosimeter, which includes the correction factors of response sensitivities, dose-response linearity, signal fading, and beam energies. Each source of uncertainty was expressed as a relative standard uncertainty, which was classified as type A (random) or type B (systematic). The total combined standard uncertainty was calculated by summing the Type A and Type B uncertainties using a quadratic method. The largest relative standard uncertainty in the absorbed dose measurement from each dosimeter arose from the calibration coefficient of these dosimetry systems (1.32 %–1.36 %), followed by the stability of the TLD/OSLD reader long term stability (0.47 %–0.52 %) and statistical TL/OSL readings (0.38 %–0.59 %). The quadratic summation of all relative standard uncertainties resulted in total combined uncertainties of 1.31 % for CGedOF and 1.36 % for both FGedOF and OSLD, resulting in tolerance limits of 3.93 % for CGedOF and 4.08 % for both FGedOF and OSLD. In conclusion, this study provides valuable insights into the measurement uncertainties of both CGedOF and FGedOF in radiotherapy applications, highlighting their potential as reliable tools for dose verification in remote dosimetry audits.
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiation Physics and Chemistry
Radiation Physics and Chemistry 化学-核科学技术
CiteScore
5.60
自引率
17.20%
发文量
574
审稿时长
12 weeks
期刊介绍: Radiation Physics and Chemistry is a multidisciplinary journal that provides a medium for publication of substantial and original papers, reviews, and short communications which focus on research and developments involving ionizing radiation in radiation physics, radiation chemistry and radiation processing. The journal aims to publish papers with significance to an international audience, containing substantial novelty and scientific impact. The Editors reserve the rights to reject, with or without external review, papers that do not meet these criteria. This could include papers that are very similar to previous publications, only with changed target substrates, employed materials, analyzed sites and experimental methods, report results without presenting new insights and/or hypothesis testing, or do not focus on the radiation effects.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信