Comparison of stress-induced hemolysis in neonatal intravenous catheters: Theoretical and experimental analysis of shear stress, exposure time, and index of hemolysis.

Q2 Medicine
Gordon Y Chang, Cynthia A Cox, Ellen A Kane, Giovanna Palermo Tye, Paul Fawcett, Thomas H Shaffer
{"title":"Comparison of stress-induced hemolysis in neonatal intravenous catheters: Theoretical and experimental analysis of shear stress, exposure time, and index of hemolysis.","authors":"Gordon Y Chang, Cynthia A Cox, Ellen A Kane, Giovanna Palermo Tye, Paul Fawcett, Thomas H Shaffer","doi":"10.1177/19345798251326071","DOIUrl":null,"url":null,"abstract":"<p><p>BackgroundSeveral studies report hemolysis when packed red blood cells (PRBCs) are transfused through small-inner-diameter (ID) catheters using presence of biomarkers but do not address cause and amount of hemolysis. This study aims to determine the cause and index of hemolysis percent (IH%) when PRBCs are infused through small-ID catheters.MethodsThe IH% was calculated using Giersiepen's empirical power law, which describes a relationship between hemolysis and magnitude of shear stress and exposure time. Six- and 27-day-old PRBCs were infused through five catheters with IDs of 0.20 mm to 0.70 mm at infusion rates of 3 mL and 10 mL/hour. Shear stress and exposure time were calculated. Data were analyzed as a function of catheter ID, infusion rates, and blood age.ResultsThe study demonstrated that shear stress was supraphysiologic during laminar flow in catheters with IDs of 0.20 mm and 0.28 mm. These catheters' IH% were ∼117 times higher at 3 mL/hour and ∼75 times higher at 10 mL/hour than catheters with larger IDs of 0.48 mm, 0.51 mm, and 0.70 mm. For blood age, in catheters with IDs of 0.20 mm and 0.28 mm, IH% was ∼155 times higher at 6 days and ∼76 times higher at 27 days than in catheters with IDs of 0.48 mm, 0.51 mm, and 0.70 mm.ConclusionsThis study demonstrated that when RBCs are subjected to supraphysiologic shear stress in catheters with IDs of ≤0.28 mm, index of hemolysis is greater than in catheters with IDs ≥0.48 mm.</p>","PeriodicalId":16537,"journal":{"name":"Journal of neonatal-perinatal medicine","volume":" ","pages":"19345798251326071"},"PeriodicalIF":0.0000,"publicationDate":"2025-03-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of neonatal-perinatal medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/19345798251326071","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

BackgroundSeveral studies report hemolysis when packed red blood cells (PRBCs) are transfused through small-inner-diameter (ID) catheters using presence of biomarkers but do not address cause and amount of hemolysis. This study aims to determine the cause and index of hemolysis percent (IH%) when PRBCs are infused through small-ID catheters.MethodsThe IH% was calculated using Giersiepen's empirical power law, which describes a relationship between hemolysis and magnitude of shear stress and exposure time. Six- and 27-day-old PRBCs were infused through five catheters with IDs of 0.20 mm to 0.70 mm at infusion rates of 3 mL and 10 mL/hour. Shear stress and exposure time were calculated. Data were analyzed as a function of catheter ID, infusion rates, and blood age.ResultsThe study demonstrated that shear stress was supraphysiologic during laminar flow in catheters with IDs of 0.20 mm and 0.28 mm. These catheters' IH% were ∼117 times higher at 3 mL/hour and ∼75 times higher at 10 mL/hour than catheters with larger IDs of 0.48 mm, 0.51 mm, and 0.70 mm. For blood age, in catheters with IDs of 0.20 mm and 0.28 mm, IH% was ∼155 times higher at 6 days and ∼76 times higher at 27 days than in catheters with IDs of 0.48 mm, 0.51 mm, and 0.70 mm.ConclusionsThis study demonstrated that when RBCs are subjected to supraphysiologic shear stress in catheters with IDs of ≤0.28 mm, index of hemolysis is greater than in catheters with IDs ≥0.48 mm.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of neonatal-perinatal medicine
Journal of neonatal-perinatal medicine Medicine-Pediatrics, Perinatology and Child Health
CiteScore
2.00
自引率
0.00%
发文量
124
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信