Accuracy of bi-coordinate and multi-coordinate handpiece calibration methods for robot-assisted implant placement.

IF 4.3 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Yi Li, Yin Zhou, Jizhe Lyu, Xunning Cao, Jianguo Tan, Xiaoqiang Liu
{"title":"Accuracy of bi-coordinate and multi-coordinate handpiece calibration methods for robot-assisted implant placement.","authors":"Yi Li, Yin Zhou, Jizhe Lyu, Xunning Cao, Jianguo Tan, Xiaoqiang Liu","doi":"10.1016/j.prosdent.2025.02.057","DOIUrl":null,"url":null,"abstract":"<p><strong>Statement of problem: </strong>To ensure accurate robot-assisted surgery, it is essential to identify the handpiece position at the end effector of the robotic arm. Clinically, the relationship between the optical tracking device and the handpiece has been typically confirmed by using a calibration plate at the end effector of the robotic arm. However, the accuracy of the handpiece calibration methods for robot-assisted implant placement remains unclear.</p><p><strong>Purpose: </strong>The purpose of this in vitro study was to evaluate the accuracy of bi-coordinate and multi-coordinate handpiece calibration methods, as well as the multi-coordinate handpiece plate under partial obstruction, in the context of robot-assisted implant placement.</p><p><strong>Material and methods: </strong>In total, 120 implants were divided into 6 groups based on the calibration plate used in the study: bi-coordinate handpiece calibration plate for the maxilla (Bmx), bi-coordinate handpiece calibration plate for the mandible (Bmn), multi-coordinate handpiece calibration plate for the maxilla (Mmx), multi-coordinate handpiece calibration plate for the mandible (Mmn), partially obscured multi-coordinate handpiece calibration plate for the mandible with the primary coordinate unblocked and the auxiliary coordinate covered (MmnPrim), and partially obscured multi-coordinate handpiece calibration plate for the mandible with the auxiliary coordinate unblocked and the primary coordinate covered (MmnAux). Calibration of the robotic arm was conducted separately for each group. Then the robot autonomously performed osteotomies and implant placements at the first and second premolars according to the preoperative plan. Following surgery, the robotic software program calculated the deviation values between the planned and actual implant positions. Differences between the test groups were analyzed using 1-way analysis of variance (ANOVA) and the Bonferroni post hoc test (α=.05).</p><p><strong>Results: </strong>The ranges of angular deviation and 3-dimensional deviations at the implant platform and apex across the 6 groups were 0.30 degrees to 0.48 degrees, 0.31 to 0.36 mm, and 0.31 to 0.38 mm, respectively. No statistically significant differences were found among the groups (P>.05).</p><p><strong>Conclusions: </strong>Both the bi-coordinate and multi-coordinate handpiece calibration methods demonstrated acceptable accuracy for robot-assisted implant placement. The multi-coordinate calibration plate provides a feasible method for robot calibration in scenarios where the mandible is partially obstructed.</p>","PeriodicalId":16866,"journal":{"name":"Journal of Prosthetic Dentistry","volume":" ","pages":""},"PeriodicalIF":4.3000,"publicationDate":"2025-03-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthetic Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.prosdent.2025.02.057","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Statement of problem: To ensure accurate robot-assisted surgery, it is essential to identify the handpiece position at the end effector of the robotic arm. Clinically, the relationship between the optical tracking device and the handpiece has been typically confirmed by using a calibration plate at the end effector of the robotic arm. However, the accuracy of the handpiece calibration methods for robot-assisted implant placement remains unclear.

Purpose: The purpose of this in vitro study was to evaluate the accuracy of bi-coordinate and multi-coordinate handpiece calibration methods, as well as the multi-coordinate handpiece plate under partial obstruction, in the context of robot-assisted implant placement.

Material and methods: In total, 120 implants were divided into 6 groups based on the calibration plate used in the study: bi-coordinate handpiece calibration plate for the maxilla (Bmx), bi-coordinate handpiece calibration plate for the mandible (Bmn), multi-coordinate handpiece calibration plate for the maxilla (Mmx), multi-coordinate handpiece calibration plate for the mandible (Mmn), partially obscured multi-coordinate handpiece calibration plate for the mandible with the primary coordinate unblocked and the auxiliary coordinate covered (MmnPrim), and partially obscured multi-coordinate handpiece calibration plate for the mandible with the auxiliary coordinate unblocked and the primary coordinate covered (MmnAux). Calibration of the robotic arm was conducted separately for each group. Then the robot autonomously performed osteotomies and implant placements at the first and second premolars according to the preoperative plan. Following surgery, the robotic software program calculated the deviation values between the planned and actual implant positions. Differences between the test groups were analyzed using 1-way analysis of variance (ANOVA) and the Bonferroni post hoc test (α=.05).

Results: The ranges of angular deviation and 3-dimensional deviations at the implant platform and apex across the 6 groups were 0.30 degrees to 0.48 degrees, 0.31 to 0.36 mm, and 0.31 to 0.38 mm, respectively. No statistically significant differences were found among the groups (P>.05).

Conclusions: Both the bi-coordinate and multi-coordinate handpiece calibration methods demonstrated acceptable accuracy for robot-assisted implant placement. The multi-coordinate calibration plate provides a feasible method for robot calibration in scenarios where the mandible is partially obstructed.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Prosthetic Dentistry
Journal of Prosthetic Dentistry 医学-牙科与口腔外科
CiteScore
7.00
自引率
13.00%
发文量
599
审稿时长
69 days
期刊介绍: The Journal of Prosthetic Dentistry is the leading professional journal devoted exclusively to prosthetic and restorative dentistry. The Journal is the official publication for 24 leading U.S. international prosthodontic organizations. The monthly publication features timely, original peer-reviewed articles on the newest techniques, dental materials, and research findings. The Journal serves prosthodontists and dentists in advanced practice, and features color photos that illustrate many step-by-step procedures. The Journal of Prosthetic Dentistry is included in Index Medicus and CINAHL.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信