{"title":"The quantified method for blood clot detection in the extraction socket.","authors":"Suwat Tanya, Piyachat Patcharanuchat, Sajee Sattayut","doi":"10.12688/f1000research.155330.1","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Currently, there is no objective and quantified measurement for detecting blood clots during extraction socket hemostasis. It has relied solely on clinical observation, even when conducting clinical research by using extraction sockets as samples. This study aimed to assess the in vitro reliability and clinical-relevant validity of a new objective measurement providing quantified data called blood clot detection (BCD) using a standard capillary tube.</p><p><strong>Methods: </strong>The in vitro part of the study was conducted using surplus blood samples from ten healthy participants. Two identical sets of blood samples in simulation reservoirs mimicking bleeding sockets were prepared for reliability tests. Then, the capillary tubes were concurrently placed in the reservoirs. The blood-filled distances were measured. The part of clinical-relevant validity study was conducted in sixteen extraction sockets from each healthy participant. Clinical observation and BCD measurement were evaluated by two calibrated assessors. The total duration of the assessment was a 30-minute.</p><p><strong>Results: </strong>The distances of the blood-filled capillary tube were decreased by time. Test and retest reliability analysis of the BCD measurement showed an excellent intraclass correlation coefficient of 0.980 (0.968 to 0.988). The medians of blood-filled distance categorized by clinical observation into active bleeding, sluggish oozing, and clot formation were 13.0 mm (Q1 = 11.7, Q3 = 13.8), 5.6 mm (Q1 = 4.3, Q3 = 7.0), and 0.9 mm (Q1 = 0.5, Q3 = 1.3), respectively. The blood-filled distance of the clot formation group was significantly less than the active bleeding and sluggish oozing (p<0.001). Therefore, the BCD measurement also significantly indicated the completion of extraction socket hemostasis.</p><p><strong>Conclusions: </strong>A distance of blood-filled in capillary tube of 0.9 mm from the BCD measurement significantly ensured complete clot formation. The BCD measurement proved to be a quantified tool for objectively measuring hemostasis of bleeding socket.</p>","PeriodicalId":12260,"journal":{"name":"F1000Research","volume":"13 ","pages":"1043"},"PeriodicalIF":0.0000,"publicationDate":"2025-04-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11914871/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"F1000Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.12688/f1000research.155330.1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"Pharmacology, Toxicology and Pharmaceutics","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Currently, there is no objective and quantified measurement for detecting blood clots during extraction socket hemostasis. It has relied solely on clinical observation, even when conducting clinical research by using extraction sockets as samples. This study aimed to assess the in vitro reliability and clinical-relevant validity of a new objective measurement providing quantified data called blood clot detection (BCD) using a standard capillary tube.
Methods: The in vitro part of the study was conducted using surplus blood samples from ten healthy participants. Two identical sets of blood samples in simulation reservoirs mimicking bleeding sockets were prepared for reliability tests. Then, the capillary tubes were concurrently placed in the reservoirs. The blood-filled distances were measured. The part of clinical-relevant validity study was conducted in sixteen extraction sockets from each healthy participant. Clinical observation and BCD measurement were evaluated by two calibrated assessors. The total duration of the assessment was a 30-minute.
Results: The distances of the blood-filled capillary tube were decreased by time. Test and retest reliability analysis of the BCD measurement showed an excellent intraclass correlation coefficient of 0.980 (0.968 to 0.988). The medians of blood-filled distance categorized by clinical observation into active bleeding, sluggish oozing, and clot formation were 13.0 mm (Q1 = 11.7, Q3 = 13.8), 5.6 mm (Q1 = 4.3, Q3 = 7.0), and 0.9 mm (Q1 = 0.5, Q3 = 1.3), respectively. The blood-filled distance of the clot formation group was significantly less than the active bleeding and sluggish oozing (p<0.001). Therefore, the BCD measurement also significantly indicated the completion of extraction socket hemostasis.
Conclusions: A distance of blood-filled in capillary tube of 0.9 mm from the BCD measurement significantly ensured complete clot formation. The BCD measurement proved to be a quantified tool for objectively measuring hemostasis of bleeding socket.
F1000ResearchPharmacology, Toxicology and Pharmaceutics-Pharmacology, Toxicology and Pharmaceutics (all)
CiteScore
5.00
自引率
0.00%
发文量
1646
审稿时长
1 weeks
期刊介绍:
F1000Research publishes articles and other research outputs reporting basic scientific, scholarly, translational and clinical research across the physical and life sciences, engineering, medicine, social sciences and humanities. F1000Research is a scholarly publication platform set up for the scientific, scholarly and medical research community; each article has at least one author who is a qualified researcher, scholar or clinician actively working in their speciality and who has made a key contribution to the article. Articles must be original (not duplications). All research is suitable irrespective of the perceived level of interest or novelty; we welcome confirmatory and negative results, as well as null studies. F1000Research publishes different type of research, including clinical trials, systematic reviews, software tools, method articles, and many others. Reviews and Opinion articles providing a balanced and comprehensive overview of the latest discoveries in a particular field, or presenting a personal perspective on recent developments, are also welcome. See the full list of article types we accept for more information.