{"title":"A Flexible PVDF Sensor for Forcecardiography.","authors":"Salvatore Parlato, Jessica Centracchio, Eliana Cinotti, Gaetano D Gargiulo, Daniele Esposito, Paolo Bifulco, Emilio Andreozzi","doi":"10.3390/s25051608","DOIUrl":null,"url":null,"abstract":"<p><p>Forcecardiography (FCG) uses force sensors to record the mechanical vibrations induced on the chest wall by cardiac and respiratory activities. FCG is usually performed via piezoelectric lead-zirconate titanate (PZT) sensors, which simultaneously record the very slow respiratory movements of the chest, the slow infrasonic vibrations due to emptying and filling of heart chambers, the faster infrasonic vibrations due to movements of heart valves, which are usually recorded via Seismocardiography (SCG), and the audible vibrations corresponding to heart sounds, commonly recorded via Phonocardiography (PCG). However, PZT sensors are not flexible and do not adapt very well to the deformations of soft tissues on the chest. This study presents a flexible FCG sensor based on a piezoelectric polyvinylidene fluoride (PVDF) transducer. The PVDF FCG sensor was compared with a well-assessed PZT FCG sensor, as well as with an electro-resistive respiratory band (ERB), an accelerometric SCG sensor, and an electronic stethoscope for PCG. Simultaneous recordings were acquired with these sensors and an electrocardiography (ECG) monitor from a cohort of 35 healthy subjects (16 males and 19 females). The PVDF sensor signals were compared in terms of morphology with those acquired simultaneously via the PZT sensor, the SCG sensor and the electronic stethoscope. Moreover, the estimation accuracies of PVDF and PZT sensors for inter-beat intervals (IBIs) and inter-breath intervals (IBrIs) were assessed against reference ECG and ERB measurements. The results of statistical analyses confirmed that the PVDF sensor provides FCG signals with very high similarity to those acquired via PZT sensors (median cross-correlation index of 0.96 across all subjects) as well as with SCG and PCG signals (median cross-correlation indices of 0.85 and 0.80, respectively). Moreover, the PVDF sensor provides very accurate estimates of IBIs, with R<sup>2</sup> > 0.99 and Bland-Altman limits of agreement (LoA) of [-5.30; 5.00] ms, and of IBrIs, with R<sup>2</sup> > 0.96 and LoA of [-0.510; 0.513] s. The flexibility of the PVDF sensor makes it more comfortable and ideal for wearable applications. Unlike PZT, PVDF is lead-free, which increases safety and biocompatibility for prolonged skin contact.</p>","PeriodicalId":21698,"journal":{"name":"Sensors","volume":"25 5","pages":""},"PeriodicalIF":3.4000,"publicationDate":"2025-03-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11902622/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sensors","FirstCategoryId":"103","ListUrlMain":"https://doi.org/10.3390/s25051608","RegionNum":3,"RegionCategory":"综合性期刊","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Forcecardiography (FCG) uses force sensors to record the mechanical vibrations induced on the chest wall by cardiac and respiratory activities. FCG is usually performed via piezoelectric lead-zirconate titanate (PZT) sensors, which simultaneously record the very slow respiratory movements of the chest, the slow infrasonic vibrations due to emptying and filling of heart chambers, the faster infrasonic vibrations due to movements of heart valves, which are usually recorded via Seismocardiography (SCG), and the audible vibrations corresponding to heart sounds, commonly recorded via Phonocardiography (PCG). However, PZT sensors are not flexible and do not adapt very well to the deformations of soft tissues on the chest. This study presents a flexible FCG sensor based on a piezoelectric polyvinylidene fluoride (PVDF) transducer. The PVDF FCG sensor was compared with a well-assessed PZT FCG sensor, as well as with an electro-resistive respiratory band (ERB), an accelerometric SCG sensor, and an electronic stethoscope for PCG. Simultaneous recordings were acquired with these sensors and an electrocardiography (ECG) monitor from a cohort of 35 healthy subjects (16 males and 19 females). The PVDF sensor signals were compared in terms of morphology with those acquired simultaneously via the PZT sensor, the SCG sensor and the electronic stethoscope. Moreover, the estimation accuracies of PVDF and PZT sensors for inter-beat intervals (IBIs) and inter-breath intervals (IBrIs) were assessed against reference ECG and ERB measurements. The results of statistical analyses confirmed that the PVDF sensor provides FCG signals with very high similarity to those acquired via PZT sensors (median cross-correlation index of 0.96 across all subjects) as well as with SCG and PCG signals (median cross-correlation indices of 0.85 and 0.80, respectively). Moreover, the PVDF sensor provides very accurate estimates of IBIs, with R2 > 0.99 and Bland-Altman limits of agreement (LoA) of [-5.30; 5.00] ms, and of IBrIs, with R2 > 0.96 and LoA of [-0.510; 0.513] s. The flexibility of the PVDF sensor makes it more comfortable and ideal for wearable applications. Unlike PZT, PVDF is lead-free, which increases safety and biocompatibility for prolonged skin contact.

用于心动描记术的柔性PVDF传感器。
力心动图(FCG)使用力传感器记录由心脏和呼吸活动引起的胸壁机械振动。FCG通常通过压电式锆钛酸铅(PZT)传感器进行,该传感器同时记录非常缓慢的胸部呼吸运动,由于心室排空和填充引起的缓慢次声振动,由于心脏瓣膜运动引起的较快次声振动,这通常通过地震心动图(SCG)记录,以及与心音对应的可听振动,通常通过心音描记术(PCG)记录。然而,PZT传感器不灵活,不能很好地适应胸部软组织的变形。本研究提出一种基于压电型聚偏氟乙烯(PVDF)换能器的柔性FCG传感器。PVDF FCG传感器与经过充分评估的PZT FCG传感器、电阻呼吸带(ERB)、加速度SCG传感器和用于PCG的电子听诊器进行了比较。这些传感器和心电图(ECG)监测器同时记录了35名健康受试者(16名男性和19名女性)的数据。将PVDF传感器信号与PZT传感器、SCG传感器和电子听诊器同时采集的信号进行形态学比较。此外,PVDF和PZT传感器对心跳间隔(IBIs)和呼吸间隔(IBIs)的估计精度与参考ECG和ERB测量结果进行了评估。统计分析结果证实,PVDF传感器提供的FCG信号与PZT传感器获得的FCG信号(所有受试者的中位数相关指数为0.96)以及SCG和PCG信号(中位数相关指数分别为0.85和0.80)具有非常高的相似性。此外,PVDF传感器提供了非常准确的ibi估计,R2 >.99和Bland-Altman一致极限(LoA)为[-5.30;5.50] ms, IBrIs的R2为0.96,LoA为-0.510;[0.513] PVDF传感器的灵活性使其更舒适,是可穿戴应用的理想选择。与PZT不同,PVDF是无铅的,这增加了长时间皮肤接触的安全性和生物相容性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Sensors
Sensors 工程技术-电化学
CiteScore
7.30
自引率
12.80%
发文量
8430
审稿时长
1.7 months
期刊介绍: Sensors (ISSN 1424-8220) provides an advanced forum for the science and technology of sensors and biosensors. It publishes reviews (including comprehensive reviews on the complete sensors products), regular research papers and short notes. Our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible. There is no restriction on the length of the papers. The full experimental details must be provided so that the results can be reproduced.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信