Quantitative multislice and jointly optimized rapid CEST for in vivo whole-brain imaging.

IF 3 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Ouri Cohen, Robert J Young, Ricardo Otazo
{"title":"Quantitative multislice and jointly optimized rapid CEST for in vivo whole-brain imaging.","authors":"Ouri Cohen, Robert J Young, Ricardo Otazo","doi":"10.1002/mrm.30488","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To develop a quantitative multislice chemical exchange saturation transfer (CEST) schedule optimization and pulse sequence that reduces the loss of sensitivity inherent to multislice sequences.</p><p><strong>Methods: </strong>A deep learning framework was developed for simultaneous optimization of scan parameters and slice order. The optimized sequence was tested in numerical simulations against a random schedule and an optimized single-slice schedule. The scan efficiency of each schedule was quantified. Three healthy subjects were scanned with the proposed sequence. Regions of interest in white matter (WM) and gray matter (GM) were defined. The sequence was compared with the single-slice sequence in vivo and differences quantified using Bland-Altman plots. Test-retest reproducibility was assessed, and the Lin's concordance correlation coefficient (CCC) was calculated for WM and GM. Intersubject variability was also measured with the CCC. Feasibility of whole-brain clinical imaging was tested using a multislab acquisition in 1 subject.</p><p><strong>Results: </strong>The optimized multislice sequence yielded a lower mean error than the random schedule for all tissue parameters and a lower error than the optimized single-slice schedule for four of six parameters. The optimized multislice sequence provided the highest scan efficiency. In vivo tissue-parameter values obtained with the proposed sequence agreed well with those of the optimized single-slice sequence and prior studies. The average WM/GM CCC was 0.8151/0.7779 for the test-retest scans and 0.7792/0.7191 for the intersubject variability experiment.</p><p><strong>Conclusion: </strong>A multislice schedule optimization framework and pulse sequence were demonstrated for quantitative CEST. The proposed approach enables accurate and reproducible whole-brain quantitative CEST imaging in clinically relevant scan times.</p>","PeriodicalId":18065,"journal":{"name":"Magnetic Resonance in Medicine","volume":" ","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2025-03-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance in Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/mrm.30488","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To develop a quantitative multislice chemical exchange saturation transfer (CEST) schedule optimization and pulse sequence that reduces the loss of sensitivity inherent to multislice sequences.

Methods: A deep learning framework was developed for simultaneous optimization of scan parameters and slice order. The optimized sequence was tested in numerical simulations against a random schedule and an optimized single-slice schedule. The scan efficiency of each schedule was quantified. Three healthy subjects were scanned with the proposed sequence. Regions of interest in white matter (WM) and gray matter (GM) were defined. The sequence was compared with the single-slice sequence in vivo and differences quantified using Bland-Altman plots. Test-retest reproducibility was assessed, and the Lin's concordance correlation coefficient (CCC) was calculated for WM and GM. Intersubject variability was also measured with the CCC. Feasibility of whole-brain clinical imaging was tested using a multislab acquisition in 1 subject.

Results: The optimized multislice sequence yielded a lower mean error than the random schedule for all tissue parameters and a lower error than the optimized single-slice schedule for four of six parameters. The optimized multislice sequence provided the highest scan efficiency. In vivo tissue-parameter values obtained with the proposed sequence agreed well with those of the optimized single-slice sequence and prior studies. The average WM/GM CCC was 0.8151/0.7779 for the test-retest scans and 0.7792/0.7191 for the intersubject variability experiment.

Conclusion: A multislice schedule optimization framework and pulse sequence were demonstrated for quantitative CEST. The proposed approach enables accurate and reproducible whole-brain quantitative CEST imaging in clinically relevant scan times.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.70
自引率
24.20%
发文量
376
审稿时长
2-4 weeks
期刊介绍: Magnetic Resonance in Medicine (Magn Reson Med) is an international journal devoted to the publication of original investigations concerned with all aspects of the development and use of nuclear magnetic resonance and electron paramagnetic resonance techniques for medical applications. Reports of original investigations in the areas of mathematics, computing, engineering, physics, biophysics, chemistry, biochemistry, and physiology directly relevant to magnetic resonance will be accepted, as well as methodology-oriented clinical studies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信