A cost-saving, picture-assisted method for toric intraocular lens alignment versus manual self-leveling marking via RoboMarker.

IF 1.4 4区 医学 Q3 OPHTHALMOLOGY
David Kiessling, Gernot F Roessler, Randolf A Widder
{"title":"A cost-saving, picture-assisted method for toric intraocular lens alignment versus manual self-leveling marking via RoboMarker.","authors":"David Kiessling, Gernot F Roessler, Randolf A Widder","doi":"10.1007/s10792-025-03479-5","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To determine the accuracy of two axis-marking methods for toric intraocular lens (IOL) implantation, one picture-assisted approach based on scleral vessel vectors, and the other based on a self-leveling device for manual marking.</p><p><strong>Methods: </strong>This retrospective single-center study involved 60 eyes of 51 participants, who underwent phacoemulsification with toric IOL implantation. In all eyes, preoperative markings were made in a seated position both manually via a self-leveling corneal marker (RoboMarker), and digitally on slit-lamp photographs, defining scleral vessels as landmarks, aiding to find the correct intraoperative orientation for an angular graduation instrument. The axis of IOL alignment at the end of surgery was determined from high resolution, intraoperative footage from a microscope-integrated camera and the axis-marking error served as an outcome measurement for both marking techniques. The endpoint was the alignment of the lens at the end of surgery.</p><p><strong>Results: </strong>The average axis-marking error was 2.5 ± 1.9 degrees for picture-assisted marking, which was significantly less than that of the self-leveling corneal marker, being 5.4 ± 4.4 degrees.</p><p><strong>Conclusion: </strong>Our results indicate that scleral vessel vector marking leads to highly accurate toric IOL alignments, while being an inexpensive technique, as solely a slit-lamp camera is required for preoperative preparation.</p>","PeriodicalId":14473,"journal":{"name":"International Ophthalmology","volume":"45 1","pages":"98"},"PeriodicalIF":1.4000,"publicationDate":"2025-03-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10792-025-03479-5","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To determine the accuracy of two axis-marking methods for toric intraocular lens (IOL) implantation, one picture-assisted approach based on scleral vessel vectors, and the other based on a self-leveling device for manual marking.

Methods: This retrospective single-center study involved 60 eyes of 51 participants, who underwent phacoemulsification with toric IOL implantation. In all eyes, preoperative markings were made in a seated position both manually via a self-leveling corneal marker (RoboMarker), and digitally on slit-lamp photographs, defining scleral vessels as landmarks, aiding to find the correct intraoperative orientation for an angular graduation instrument. The axis of IOL alignment at the end of surgery was determined from high resolution, intraoperative footage from a microscope-integrated camera and the axis-marking error served as an outcome measurement for both marking techniques. The endpoint was the alignment of the lens at the end of surgery.

Results: The average axis-marking error was 2.5 ± 1.9 degrees for picture-assisted marking, which was significantly less than that of the self-leveling corneal marker, being 5.4 ± 4.4 degrees.

Conclusion: Our results indicate that scleral vessel vector marking leads to highly accurate toric IOL alignments, while being an inexpensive technique, as solely a slit-lamp camera is required for preoperative preparation.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.20
自引率
0.00%
发文量
451
期刊介绍: International Ophthalmology provides the clinician with articles on all the relevant subspecialties of ophthalmology, with a broad international scope. The emphasis is on presentation of the latest clinical research in the field. In addition, the journal includes regular sections devoted to new developments in technologies, products, and techniques.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信