Yuanli Zhang, Hong Huang, Chongzhi Yin, Guizhi Zhang, Yang Wang, Rui Gao, Jinlin Song
{"title":"Clinical Evaluation of ODIS-1 Orthodontic Operation and Image Quality of Digital Imaging System.","authors":"Yuanli Zhang, Hong Huang, Chongzhi Yin, Guizhi Zhang, Yang Wang, Rui Gao, Jinlin Song","doi":"10.2174/0115734056345020250223150845","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>With the rapid development of computer technology, the application of digital technology to the display and processing of medical images has become a common concern. In recent years, oral digital imaging technology has received more and more attention.</p><p><strong>Objective: </strong>This paper mainly aims at the ODIS-1 oral digital imaging system to analyze and study the image quality and image aims at the ODIS-1 oral digital imaging system to analyze and study the image quality and processing technology, of which X-ray imaging is indispensable.</p><p><strong>Methods: </strong>In this paper, the ODIS-1 digital scanning technology is used to detect different types of dental tissues, and its application in diagnosing oral diseases is evaluated. This paper takes 320 inpatients as the research object and uses Kodak dental film to compare the image quality of different positions.</p><p><strong>Results: </strong>It is found that there is no significant difference in image quality between the maxillary anterior teeth and mandibular anterior teeth and the maxillary posterior teeth and mandibular posterior teeth (P>0.05); the image quality of maxillary anterior teeth, mandibular anterior teeth, and maxillary posterior teeth and mandibular teeth are significantly different (P<0.05); among the various positions of the ODIS-1 oral digital imaging system, the image quality of the anterior teeth area is the best, while the image quality of the maxillary posterior teeth area is the worst.</p><p><strong>Conclusion: </strong>However, the system has a variety of image post-processing functions, which can adjust the brightness and contrast of the image arbitrarily, select the area of interest in the image according to the detection requirements, and perform local amplification, edge enhancement, and other technologies to make the image achieve the best effect. In the case of poor image quality, the clarity of the image can be further improved through image post-processing and analysis.</p>","PeriodicalId":54215,"journal":{"name":"Current Medical Imaging Reviews","volume":" ","pages":""},"PeriodicalIF":1.1000,"publicationDate":"2025-03-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Current Medical Imaging Reviews","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2174/0115734056345020250223150845","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background: With the rapid development of computer technology, the application of digital technology to the display and processing of medical images has become a common concern. In recent years, oral digital imaging technology has received more and more attention.
Objective: This paper mainly aims at the ODIS-1 oral digital imaging system to analyze and study the image quality and image aims at the ODIS-1 oral digital imaging system to analyze and study the image quality and processing technology, of which X-ray imaging is indispensable.
Methods: In this paper, the ODIS-1 digital scanning technology is used to detect different types of dental tissues, and its application in diagnosing oral diseases is evaluated. This paper takes 320 inpatients as the research object and uses Kodak dental film to compare the image quality of different positions.
Results: It is found that there is no significant difference in image quality between the maxillary anterior teeth and mandibular anterior teeth and the maxillary posterior teeth and mandibular posterior teeth (P>0.05); the image quality of maxillary anterior teeth, mandibular anterior teeth, and maxillary posterior teeth and mandibular teeth are significantly different (P<0.05); among the various positions of the ODIS-1 oral digital imaging system, the image quality of the anterior teeth area is the best, while the image quality of the maxillary posterior teeth area is the worst.
Conclusion: However, the system has a variety of image post-processing functions, which can adjust the brightness and contrast of the image arbitrarily, select the area of interest in the image according to the detection requirements, and perform local amplification, edge enhancement, and other technologies to make the image achieve the best effect. In the case of poor image quality, the clarity of the image can be further improved through image post-processing and analysis.
期刊介绍:
Current Medical Imaging Reviews publishes frontier review articles, original research articles, drug clinical trial studies and guest edited thematic issues on all the latest advances on medical imaging dedicated to clinical research. All relevant areas are covered by the journal, including advances in the diagnosis, instrumentation and therapeutic applications related to all modern medical imaging techniques.