Javier Orbegozo, Angelica Pérez, Iñaki Basterra, Helena Noguera, Sergio Díaz, Laura González, Gorka Lauzirika, David P Piñero
{"title":"Clinical Outcomes of a Refractive Extended Depth of Focus Intraocular Lens.","authors":"Javier Orbegozo, Angelica Pérez, Iñaki Basterra, Helena Noguera, Sergio Díaz, Laura González, Gorka Lauzirika, David P Piñero","doi":"10.3928/1081597X-20250204-02","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the clinical and patient-reported outcomes of the toric version of a refractive extended depth of focus intraocular lens (IOL).</p><p><strong>Methods: </strong>This was a prospective observational study including 26 eyes of 26 patients (age range: 58 to 88 years) who underwent cataract surgery with implantation of the LuxSmart toric IOL (Bausch & Lomb GmbH). Changes in distance, intermediate (80 cm), and near (40 cm) visual acuities, manifest refraction, defocus curve, distance photopic contrast sensitivity (CSV-1000 test), and patient-reported outcomes (Catquest-9SF questionnaire) were evaluated during a 3-month follow-up.</p><p><strong>Results: </strong>Mean 3-month postoperative uncorrected distance, corrected distance, distance-corrected intermediate, and distance-corrected near visual acuities of 0.10 ± 0.11, 0.03 ± 0.08, 0.06 ± 0.13, and 0.38 ± 0.22 logarithm of the minimum angle of resolution (logMAR) were found, respectively. Uncorrected and corrected distance visual acuity improved significantly with surgery (<i>P</i> < .001). In the defocus curve, mean visual acuity was 0.30 logMAR or better for defocus between +0.50 and -2.00 D, without significant changes during the follow-up (<i>P</i> ⩾ .071). The postoperative spherical equivalent was within ±0.50 and ±1.00 diopters (D) in 92% and 96% of eyes, respectively. Surgically induced astigmatism prediction error ranged between -0.37 and 0.80 D, with a mean value of 0.11 ± 0.28 D. Mean 3-month postoperative contrast sensitivity was within or close to the upper limit of the normality range for most spatial frequencies. Mean 3-month postoperative overall Rasch calibrated Catquest score was -1.68 ± 1.04, with 96.2% of patients satisfied with the vision achieved. IOL rotations of 5º or less were observed in 92.3% of eyes.</p><p><strong>Conclusions: </strong>The toric extended depth of focus IOL evaluated provides a good distance and intermediate functional vision in eyes with preexisting corneal astigmatism, with functional near visual performance for most patients. <b>[<i>J Refract Surg</i>. 2025;41(3):e238-e247.]</b>.</p>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"41 3","pages":"e238-e247"},"PeriodicalIF":2.9000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597X-20250204-02","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate the clinical and patient-reported outcomes of the toric version of a refractive extended depth of focus intraocular lens (IOL).
Methods: This was a prospective observational study including 26 eyes of 26 patients (age range: 58 to 88 years) who underwent cataract surgery with implantation of the LuxSmart toric IOL (Bausch & Lomb GmbH). Changes in distance, intermediate (80 cm), and near (40 cm) visual acuities, manifest refraction, defocus curve, distance photopic contrast sensitivity (CSV-1000 test), and patient-reported outcomes (Catquest-9SF questionnaire) were evaluated during a 3-month follow-up.
Results: Mean 3-month postoperative uncorrected distance, corrected distance, distance-corrected intermediate, and distance-corrected near visual acuities of 0.10 ± 0.11, 0.03 ± 0.08, 0.06 ± 0.13, and 0.38 ± 0.22 logarithm of the minimum angle of resolution (logMAR) were found, respectively. Uncorrected and corrected distance visual acuity improved significantly with surgery (P < .001). In the defocus curve, mean visual acuity was 0.30 logMAR or better for defocus between +0.50 and -2.00 D, without significant changes during the follow-up (P ⩾ .071). The postoperative spherical equivalent was within ±0.50 and ±1.00 diopters (D) in 92% and 96% of eyes, respectively. Surgically induced astigmatism prediction error ranged between -0.37 and 0.80 D, with a mean value of 0.11 ± 0.28 D. Mean 3-month postoperative contrast sensitivity was within or close to the upper limit of the normality range for most spatial frequencies. Mean 3-month postoperative overall Rasch calibrated Catquest score was -1.68 ± 1.04, with 96.2% of patients satisfied with the vision achieved. IOL rotations of 5º or less were observed in 92.3% of eyes.
Conclusions: The toric extended depth of focus IOL evaluated provides a good distance and intermediate functional vision in eyes with preexisting corneal astigmatism, with functional near visual performance for most patients. [J Refract Surg. 2025;41(3):e238-e247.].
期刊介绍:
The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as:
• Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics”
• Supplemental videos and materials available for many articles
• Access to current articles, as well as several years of archived content
• Articles posted online just 2 months after acceptance.