EVALUATION OF HAND TREMOR SUPPRESSION USING A PASSIVE SURGICAL SUPPORT ROBOT IN SIMULATED RETINAL VASCULAR CANNULATION.

IF 2.3 2区 医学 Q2 OPHTHALMOLOGY
Shin Tanaka, Yuta Taniguchi, Hiroshi Enaida, Kentaro Nakamura, Hiroaki Sakai, Soichiro Yamamoto, Shohei Kitahata, Tatsuya Inoue, Toshiro Yamanaka, Fumihito Arai, Kazuaki Kadonosono
{"title":"EVALUATION OF HAND TREMOR SUPPRESSION USING A PASSIVE SURGICAL SUPPORT ROBOT IN SIMULATED RETINAL VASCULAR CANNULATION.","authors":"Shin Tanaka, Yuta Taniguchi, Hiroshi Enaida, Kentaro Nakamura, Hiroaki Sakai, Soichiro Yamamoto, Shohei Kitahata, Tatsuya Inoue, Toshiro Yamanaka, Fumihito Arai, Kazuaki Kadonosono","doi":"10.1097/IAE.0000000000004447","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To quantitatively evaluate hand tremors during retinal vascular cannulation surgery with a microneedle and investigate if a customized passive surgical robot suppresses hand tremors.</p><p><strong>Methods: </strong>A surgical simulation eye was modified, and an analysis system was used that synchronized the surgical video and accelerometer attached to the tail of the microneedle syringe. A cannulation target mimicking the optic disc and retinal vessels was designed in the bottom of the artificial eye. Two vitreous surgeons used this system to evaluate hand tremor 15 seconds before cannulation and 60 seconds during cannulation with and without the robot.</p><p><strong>Results: </strong>Fifteen seconds before cannulation, the mean horizontal, vertical, and combined components of hand tremor were 3.8 ± 1.7, 0.93 ± 0.48, and 4.0 ± 1.7 mG without the robot, decreasing to 3.3 ± 1.6, 0.87 ± 0.52, and 3.5 ± 1.8 mG with the robot. Significant reductions were observed in the horizontal (p = 0.012) and combined components (p = 0.020). Sixty seconds during cannulation, the mean horizontal, vertical, and combined components of hand tremor were 3.3 ± 1.0, 0.53 ± 0.24, and 3.4 ± 1.0 mG, respectively, without the robot. These components significantly decreased with the robot to 2.2 ± 0.7, 0.40 ± 0.15, and 2.2 ± 0.7 mG, respectively, for all components (p < 0.001). Hand tremor decreased significantly by 35% in the combined components during cannulation with the robot.</p><p><strong>Conclusion: </strong>The customized passive surgical support robot reduces hand tremor during simulated retinal vascular cannulation surgery.</p>","PeriodicalId":54486,"journal":{"name":"Retina-The Journal of Retinal and Vitreous Diseases","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2025-03-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Retina-The Journal of Retinal and Vitreous Diseases","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/IAE.0000000000004447","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To quantitatively evaluate hand tremors during retinal vascular cannulation surgery with a microneedle and investigate if a customized passive surgical robot suppresses hand tremors.

Methods: A surgical simulation eye was modified, and an analysis system was used that synchronized the surgical video and accelerometer attached to the tail of the microneedle syringe. A cannulation target mimicking the optic disc and retinal vessels was designed in the bottom of the artificial eye. Two vitreous surgeons used this system to evaluate hand tremor 15 seconds before cannulation and 60 seconds during cannulation with and without the robot.

Results: Fifteen seconds before cannulation, the mean horizontal, vertical, and combined components of hand tremor were 3.8 ± 1.7, 0.93 ± 0.48, and 4.0 ± 1.7 mG without the robot, decreasing to 3.3 ± 1.6, 0.87 ± 0.52, and 3.5 ± 1.8 mG with the robot. Significant reductions were observed in the horizontal (p = 0.012) and combined components (p = 0.020). Sixty seconds during cannulation, the mean horizontal, vertical, and combined components of hand tremor were 3.3 ± 1.0, 0.53 ± 0.24, and 3.4 ± 1.0 mG, respectively, without the robot. These components significantly decreased with the robot to 2.2 ± 0.7, 0.40 ± 0.15, and 2.2 ± 0.7 mG, respectively, for all components (p < 0.001). Hand tremor decreased significantly by 35% in the combined components during cannulation with the robot.

Conclusion: The customized passive surgical support robot reduces hand tremor during simulated retinal vascular cannulation surgery.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.70
自引率
9.10%
发文量
554
审稿时长
3-6 weeks
期刊介绍: ​RETINA® focuses exclusively on the growing specialty of vitreoretinal disorders. The Journal provides current information on diagnostic and therapeutic techniques. Its highly specialized and informative, peer-reviewed articles are easily applicable to clinical practice. In addition to regular reports from clinical and basic science investigators, RETINA® publishes special features including periodic review articles on pertinent topics, special articles dealing with surgical and other therapeutic techniques, and abstract cards. Issues are abundantly illustrated in vivid full color. Published 12 times per year, RETINA® is truly a “must have” publication for anyone connected to this field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信