Visual acuities as measured with an automatic intelligent visual acuity chart projector and standard logarithmic visual acuity chart: a prospective comparative study.

IF 3.1 3区 医学 Q1 MEDICINE, GENERAL & INTERNAL
Frontiers in Medicine Pub Date : 2025-02-24 eCollection Date: 2025-01-01 DOI:10.3389/fmed.2025.1447679
Mingyue Luo, Xinyu Liu, Dingding Zhang, Youxin Chen, Weijuan Kang
{"title":"Visual acuities as measured with an automatic intelligent visual acuity chart projector and standard logarithmic visual acuity chart: a prospective comparative study.","authors":"Mingyue Luo, Xinyu Liu, Dingding Zhang, Youxin Chen, Weijuan Kang","doi":"10.3389/fmed.2025.1447679","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to test the repeatability of visual acuity (VA) measurement using an novel intelligent projector visual acuity (IP) chart (LSJ-IVAC-6000A, Hunan Liangshi Jia Biotechnology Co., Ltd.) and the consistency of VA measurement between the IP chart and the 5 m standard logarithmic visual acuity (SL) chart (GB11533-2011).</p><p><strong>Methods: </strong>In this prospective comparative study, 53 subjects were included to test the internal repeatability of the IP chart. Both eyes of the subjects were tested three times, with a minimum interval of 10 min. The intraclass correlation coefficient (ICC) was used to evaluate the repeatability. One hundred subjects were included to test the consistency between two charts. Both eyes of the subjects were tested with two charts in random order with an interval of 10 min for each test. ICC and the Bland-Altman statistical analyses were used to analyze the consistency of the two charts. VA values were expressed by the logMAR recording method. The time consumption for each test was recorded in seconds.</p><p><strong>Results: </strong>In the repeatability analysis, the ICC values of the right and left eye were 0.968 (95% CI, 0.950-0.981, <i>p</i> < 0.001) and 0.964 (95% CI, 0.944-0.978, <i>p</i> < 0.001), respectively. In the consistency analysis, the ICC values of the right and left eye were 0.946 (95% CI, 0.919-0.963, <i>p</i> < 0.001) and 0.817 (95% CI, 0.727-0.877, <i>p</i> < 0.001), respectively. However, the 95% limits of agreement (LoA) of the differences in the VA measurements between the IP and SL chart for the right and left eye were -0.26 to 0.31 and -0.34 to 0.39, respectively, suggesting a certain degree of instability in the measurements of the SL chart. There was no significant difference in the time consumption of VA measurements between the two charts (<i>p</i> = 0.668). In the consistency analysis of subgroups by age, the ICC values were >0.8 (<i>p</i> < 0.001) in the most groups except for the left eyes of 51-70 years old subgroup which had an ICC value of 0.448 (95% CI, -0.919 to 0.841, <i>p</i> = 0.17), with 95% LoA of -1.02 to 0.77.</p><p><strong>Conclusion: </strong>The IP chart demonstrated good repeatability and overall consistency with the SL chart.</p>","PeriodicalId":12488,"journal":{"name":"Frontiers in Medicine","volume":"12 ","pages":"1447679"},"PeriodicalIF":3.1000,"publicationDate":"2025-02-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11892698/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3389/fmed.2025.1447679","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q1","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study aimed to test the repeatability of visual acuity (VA) measurement using an novel intelligent projector visual acuity (IP) chart (LSJ-IVAC-6000A, Hunan Liangshi Jia Biotechnology Co., Ltd.) and the consistency of VA measurement between the IP chart and the 5 m standard logarithmic visual acuity (SL) chart (GB11533-2011).

Methods: In this prospective comparative study, 53 subjects were included to test the internal repeatability of the IP chart. Both eyes of the subjects were tested three times, with a minimum interval of 10 min. The intraclass correlation coefficient (ICC) was used to evaluate the repeatability. One hundred subjects were included to test the consistency between two charts. Both eyes of the subjects were tested with two charts in random order with an interval of 10 min for each test. ICC and the Bland-Altman statistical analyses were used to analyze the consistency of the two charts. VA values were expressed by the logMAR recording method. The time consumption for each test was recorded in seconds.

Results: In the repeatability analysis, the ICC values of the right and left eye were 0.968 (95% CI, 0.950-0.981, p < 0.001) and 0.964 (95% CI, 0.944-0.978, p < 0.001), respectively. In the consistency analysis, the ICC values of the right and left eye were 0.946 (95% CI, 0.919-0.963, p < 0.001) and 0.817 (95% CI, 0.727-0.877, p < 0.001), respectively. However, the 95% limits of agreement (LoA) of the differences in the VA measurements between the IP and SL chart for the right and left eye were -0.26 to 0.31 and -0.34 to 0.39, respectively, suggesting a certain degree of instability in the measurements of the SL chart. There was no significant difference in the time consumption of VA measurements between the two charts (p = 0.668). In the consistency analysis of subgroups by age, the ICC values were >0.8 (p < 0.001) in the most groups except for the left eyes of 51-70 years old subgroup which had an ICC value of 0.448 (95% CI, -0.919 to 0.841, p = 0.17), with 95% LoA of -1.02 to 0.77.

Conclusion: The IP chart demonstrated good repeatability and overall consistency with the SL chart.

目的:本研究旨在测试使用新型智能投影视力表(IP)(LSJ-IVAC-6000A,湖南良视佳生物科技有限公司)测量视力的重复性,以及IP视力表与5 m标准对数视力表(SL)(GB11533-2011)测量视力的一致性:在这项前瞻性比较研究中,共纳入了 53 名受试者来测试 IP 视力表的内部重复性。受试者的双眼均接受了三次测试,测试间隔至少为 10 分钟。使用类内相关系数(ICC)评估重复性。100 名受试者被纳入测试两个图表之间的一致性。受试者的双眼随机使用两张图表进行测试,每次测试间隔 10 分钟。采用 ICC 和 Bland-Altman 统计分析来分析两个图表的一致性。VA 值用对数马尔记录法表示。每次测试消耗的时间以秒为单位记录:在重复性分析中,左右眼的 ICC 值为 0.968(95% CI,0.950-0.981,p p p p = 0.668)。在按年龄分组的一致性分析中,ICC 值均大于 0.8(P P = 0.17),95% LoA 为-1.02 至 0.77:IP 图表与 SL 图表具有良好的重复性和整体一致性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Frontiers in Medicine
Frontiers in Medicine Medicine-General Medicine
CiteScore
5.10
自引率
5.10%
发文量
3710
审稿时长
12 weeks
期刊介绍: Frontiers in Medicine publishes rigorously peer-reviewed research linking basic research to clinical practice and patient care, as well as translating scientific advances into new therapies and diagnostic tools. Led by an outstanding Editorial Board of international experts, this multidisciplinary open-access journal is at the forefront of disseminating and communicating scientific knowledge and impactful discoveries to researchers, academics, clinicians and the public worldwide. In addition to papers that provide a link between basic research and clinical practice, a particular emphasis is given to studies that are directly relevant to patient care. In this spirit, the journal publishes the latest research results and medical knowledge that facilitate the translation of scientific advances into new therapies or diagnostic tools. The full listing of the Specialty Sections represented by Frontiers in Medicine is as listed below. As well as the established medical disciplines, Frontiers in Medicine is launching new sections that together will facilitate - the use of patient-reported outcomes under real world conditions - the exploitation of big data and the use of novel information and communication tools in the assessment of new medicines - the scientific bases for guidelines and decisions from regulatory authorities - access to medicinal products and medical devices worldwide - addressing the grand health challenges around the world
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信