Evaluation of the Sysmex CN-6000 Coagulation Analyzer for Routine and Specialized Coagulation Testing in a Central Laboratory.

IF 0.7 4区 医学 Q4 MEDICAL LABORATORY TECHNOLOGY
Marija Kocijancic, Janina Roche, Andreas Peter, Sebastian Hörber
{"title":"Evaluation of the Sysmex CN-6000 Coagulation Analyzer for Routine and Specialized Coagulation Testing in a Central Laboratory.","authors":"Marija Kocijancic, Janina Roche, Andreas Peter, Sebastian Hörber","doi":"10.7754/Clin.Lab.2024.240922","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The Sysmex CN-6000 is a fully automated high-throughput coagulation analyzer. The objective of this study was to evaluate the analytical performance of the analyzer for routine and special coagulation testing in a high-throughput central laboratory of a university hospital.</p><p><strong>Methods: </strong>The within- and between-day precision and accuracy of 29 coagulation parameters were evaluated on the Sysmex CN-6000 using commercially available quality control materials. Patient plasma samples were used to compare results of coagulation measurements between the Sysmex CN-6000 and the Atellica COAG 360, including plasma samples with visual interference. The sample throughputs of both analyzers were compared using plasma samples from healthy volunteers.</p><p><strong>Results: </strong>Within- and between-day coefficients of variation were acceptable for all assays tested on the Sysmex CN-6000. High correlation and good agreement were observed when comparing coagulation results from the Sysmex CN-6000 and the Atellica COAG 360. Samples with visual interference showed comparable coagulation results between the two analyzers, with slightly better detection by the Sysmex CN-6000. The sample throughput per hour for analysis of a panel of five coagulation parameters was higher with the Sysmex CN-6000 compared to the Atellica COAG 360 (247 vs. 193 tests).</p><p><strong>Conclusions: </strong>The Sysmex CN-6000 demonstrated excellent analytical performance for a large number of coagulation parameters and has a high throughput capacity, ideal for the needs of a central laboratory with a high volume of routine and specialized coagulation testing.</p>","PeriodicalId":10384,"journal":{"name":"Clinical laboratory","volume":"71 3","pages":""},"PeriodicalIF":0.7000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical laboratory","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.7754/Clin.Lab.2024.240922","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The Sysmex CN-6000 is a fully automated high-throughput coagulation analyzer. The objective of this study was to evaluate the analytical performance of the analyzer for routine and special coagulation testing in a high-throughput central laboratory of a university hospital.

Methods: The within- and between-day precision and accuracy of 29 coagulation parameters were evaluated on the Sysmex CN-6000 using commercially available quality control materials. Patient plasma samples were used to compare results of coagulation measurements between the Sysmex CN-6000 and the Atellica COAG 360, including plasma samples with visual interference. The sample throughputs of both analyzers were compared using plasma samples from healthy volunteers.

Results: Within- and between-day coefficients of variation were acceptable for all assays tested on the Sysmex CN-6000. High correlation and good agreement were observed when comparing coagulation results from the Sysmex CN-6000 and the Atellica COAG 360. Samples with visual interference showed comparable coagulation results between the two analyzers, with slightly better detection by the Sysmex CN-6000. The sample throughput per hour for analysis of a panel of five coagulation parameters was higher with the Sysmex CN-6000 compared to the Atellica COAG 360 (247 vs. 193 tests).

Conclusions: The Sysmex CN-6000 demonstrated excellent analytical performance for a large number of coagulation parameters and has a high throughput capacity, ideal for the needs of a central laboratory with a high volume of routine and specialized coagulation testing.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Clinical laboratory
Clinical laboratory 医学-医学实验技术
CiteScore
1.50
自引率
0.00%
发文量
494
审稿时长
3 months
期刊介绍: Clinical Laboratory is an international fully peer-reviewed journal covering all aspects of laboratory medicine and transfusion medicine. In addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies. The journal publishes original articles, review articles, posters, short reports, case studies and letters to the editor dealing with 1) the scientific background, implementation and diagnostic significance of laboratory methods employed in hospitals, blood banks and physicians'' offices and with 2) scientific, administrative and clinical aspects of transfusion medicine and 3) in addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信