Christopher Reilly-Stitt, Ian Jennings, Steve Kitchen, Michael Cahillane, Christine Saunders, Chloe George, Tom Scorer, Isobel D. Walker
{"title":"Cold Stored Platelets: A Solution for Platelet Aggregation External Quality Assessment/Proficiency Testing","authors":"Christopher Reilly-Stitt, Ian Jennings, Steve Kitchen, Michael Cahillane, Christine Saunders, Chloe George, Tom Scorer, Isobel D. Walker","doi":"10.1111/ijlh.14445","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>To achieve accreditation for ISO 15189, laboratories are required to either participate in EQA exercises or intra-laboratory comparisons to meet the standard. Light transmission aggregometry performed by laboratory scientists for the clinical investigation of possible platelet function defects is time dependent. Current EQA available audits the interpretation of platelet aggregation traces.</p>\n </section>\n \n <section>\n \n <h3> Objectives</h3>\n \n <p>Could NEQAS BC provide external quality assessment (EQA) material to centres employing Light Transmission Aggregometry LTA for clinical investigation of platelet function? The use of fresh donor platelets could audit more of the analytical and post analytical aspects of light transmission aggregation than currently available.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A pool of donor platelets was split into aliquots that were distributed to testing centres across England, Scotland and Wales for testing within a 72 h window. Participating centres employed their locally validated testing methods for LTA assays, for agonists including ADP; Arachidonic acid; Collagen; Epinephrine; Ristocetin; TRAP and U46619.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Five different aggregation platforms were used including: Chronolog 700; Helena Aggram; PAP-8; Stago TA-8V and Sysmex CN-series. Sample packs were tested through the 72 h window with most sites performing LTA on the EQA material on day one of the three.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The % consensus of interpretations provided by participants for agonists including: ADP; Arachidonic acid; Collagen; Epinephrine and Ristocetin ranged from 94% to 100% indicating that the material is stable plus centres using different aggregometers returned the same clinical interpretations.</p>\n </section>\n </div>","PeriodicalId":14120,"journal":{"name":"International Journal of Laboratory Hematology","volume":"47 3","pages":"529-535"},"PeriodicalIF":2.2000,"publicationDate":"2025-03-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Laboratory Hematology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/ijlh.14445","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
To achieve accreditation for ISO 15189, laboratories are required to either participate in EQA exercises or intra-laboratory comparisons to meet the standard. Light transmission aggregometry performed by laboratory scientists for the clinical investigation of possible platelet function defects is time dependent. Current EQA available audits the interpretation of platelet aggregation traces.
Objectives
Could NEQAS BC provide external quality assessment (EQA) material to centres employing Light Transmission Aggregometry LTA for clinical investigation of platelet function? The use of fresh donor platelets could audit more of the analytical and post analytical aspects of light transmission aggregation than currently available.
Methods
A pool of donor platelets was split into aliquots that were distributed to testing centres across England, Scotland and Wales for testing within a 72 h window. Participating centres employed their locally validated testing methods for LTA assays, for agonists including ADP; Arachidonic acid; Collagen; Epinephrine; Ristocetin; TRAP and U46619.
Results
Five different aggregation platforms were used including: Chronolog 700; Helena Aggram; PAP-8; Stago TA-8V and Sysmex CN-series. Sample packs were tested through the 72 h window with most sites performing LTA on the EQA material on day one of the three.
Conclusion
The % consensus of interpretations provided by participants for agonists including: ADP; Arachidonic acid; Collagen; Epinephrine and Ristocetin ranged from 94% to 100% indicating that the material is stable plus centres using different aggregometers returned the same clinical interpretations.
期刊介绍:
The International Journal of Laboratory Hematology provides a forum for the communication of new developments, research topics and the practice of laboratory haematology.
The journal publishes invited reviews, full length original articles, and correspondence.
The International Journal of Laboratory Hematology is the official journal of the International Society for Laboratory Hematology, which addresses the following sub-disciplines: cellular analysis, flow cytometry, haemostasis and thrombosis, molecular diagnostics, haematology informatics, haemoglobinopathies, point of care testing, standards and guidelines.
The journal was launched in 2006 as the successor to Clinical and Laboratory Hematology, which was first published in 1979. An active and positive editorial policy ensures that work of a high scientific standard is reported, in order to bridge the gap between practical and academic aspects of laboratory haematology.