Onsite Regeneration of Crystal Data in a 3/8" Foresight Detector Using a Digital Workflow.

IF 0.4 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Indian Journal of Nuclear Medicine Pub Date : 2024-09-01 Epub Date: 2025-01-25 DOI:10.4103/ijnm.ijnm_81_24
Sarika Sharma Prashar, Ajay Kumar, Ashwani Sood, Bhagwant Rai Mittal
{"title":"Onsite Regeneration of Crystal Data in a 3/8\" Foresight Detector Using a Digital Workflow.","authors":"Sarika Sharma Prashar, Ajay Kumar, Ashwani Sood, Bhagwant Rai Mittal","doi":"10.4103/ijnm.ijnm_81_24","DOIUrl":null,"url":null,"abstract":"<p><p>Generally after the accreditation of the gamma camera, its unfailing optimal operation is guaranteed with established departmental preventive maintenance schedules. However, it is not easy to always religiously follow the same maintenance routines. Therefore, instead of preventive maintenance schedule, a reactive maintenance schedule was adopted in our department for a short span of time. In our center, the humidity level in the gamma camera room was recorded to be 90%, which was much higher than the recommended value of 50%-60%. This was due to the inundated water in the gamma camera room due to an accidental breakage of fire-extinguishing water pipeline, adjacent to the room. Postaccident, higher values of nonuniformity (both intrinsic and extrinsic uniformity) were observed. The reactive maintenance was undertaken by medical physicist and field engineer to record and confirm the crystal hydration in the detector and to restore the system for the clinical studies. The digital workflow program successfully rectified the problem of crystal hydration. It is highly recommended to perform timely reactive action which may prevent the permanent damage to the systems.</p>","PeriodicalId":45830,"journal":{"name":"Indian Journal of Nuclear Medicine","volume":"39 5","pages":"411-414"},"PeriodicalIF":0.4000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11884352/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Journal of Nuclear Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/ijnm.ijnm_81_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/25 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Generally after the accreditation of the gamma camera, its unfailing optimal operation is guaranteed with established departmental preventive maintenance schedules. However, it is not easy to always religiously follow the same maintenance routines. Therefore, instead of preventive maintenance schedule, a reactive maintenance schedule was adopted in our department for a short span of time. In our center, the humidity level in the gamma camera room was recorded to be 90%, which was much higher than the recommended value of 50%-60%. This was due to the inundated water in the gamma camera room due to an accidental breakage of fire-extinguishing water pipeline, adjacent to the room. Postaccident, higher values of nonuniformity (both intrinsic and extrinsic uniformity) were observed. The reactive maintenance was undertaken by medical physicist and field engineer to record and confirm the crystal hydration in the detector and to restore the system for the clinical studies. The digital workflow program successfully rectified the problem of crystal hydration. It is highly recommended to perform timely reactive action which may prevent the permanent damage to the systems.

现场再生晶体数据在3/8”预见检测器使用数字工作流程。
一般伽玛相机的认可后,其始终如一的最佳操作是由建立部门预防性维护计划保证。然而,总是虔诚地遵循相同的维护程序并不容易。因此,我部门在短时间内采用了被动维修计划,而不是预防性维修计划。在我们中心,伽马相机室的湿度水平被记录为90%,远远高于推荐值50%-60%。这是由于相邻的伽玛摄像室灭火水管意外断裂,导致伽玛摄像室积水。事故发生后,观察到较高的非均匀性值(内在和外在均匀性)。由医学物理学家和现场工程师进行反应性维护,记录和确认探测器中的晶体水合作用,恢复系统用于临床研究。数字化工作流程程序成功地解决了晶体水化问题。强烈建议及时采取反应性措施,以防止对系统造成永久性损害。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Indian Journal of Nuclear Medicine
Indian Journal of Nuclear Medicine RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
0.70
自引率
0.00%
发文量
46
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信