A prospective comparison of Scheimpflug-based anterior segment imaging systems in normal and abnormal corneas: Oculus Pentacam vs. Remidio Visionix VX650.

IF 0.9
Frontiers in ophthalmology Pub Date : 2025-02-19 eCollection Date: 2025-01-01 DOI:10.3389/fopht.2025.1516423
Sonam Kumar, Kalpa Negiloni, Parul Chawla Gupta, Ranjitha K, Neha N, Divya Parthasarathy Rao, Jyothi Damani, Ritika Thakur, Shonraj Ballae Ganesh Rao, Mona Duggal
{"title":"A prospective comparison of Scheimpflug-based anterior segment imaging systems in normal and abnormal corneas: Oculus Pentacam vs. Remidio Visionix VX650.","authors":"Sonam Kumar, Kalpa Negiloni, Parul Chawla Gupta, Ranjitha K, Neha N, Divya Parthasarathy Rao, Jyothi Damani, Ritika Thakur, Shonraj Ballae Ganesh Rao, Mona Duggal","doi":"10.3389/fopht.2025.1516423","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Advancements in anterior segment imaging have significantly improved the precision of corneal diagnostics, aiding ocular conditions management. The study evaluated the agreement between two Scheimpflug devices-Pentacam and Remidio Visionix 650 (VX650)-in assessing corneal and anterior chamber parameters in patients with and without corneal abnormalities.</p><p><strong>Materials and methods: </strong>This prospective study at a tertiary eye care hospital measured pachymetry values (central corneal thickness, thinnest point), front corneal surface parameters (keratometry, curvature, astigmatism), back corneal surface parameters (elevation, curvature) and anterior chamber parameters (depth, volume, angle). Bland-Altman analysis assessed agreement between the devices, and the Intraclass Correlation Coefficient (ICC) evaluated VX650's repeatability.</p><p><strong>Results: </strong>Of the 202 patients (mean age was 25 ± 4.33 years), 141 eyes were classified as normal and 61 as abnormal (41 keratoconus, 17 post-LASIK, 3 post-ICL). Bland-Altman analysis showed good agreement between Pentacam and VX650 for anterior segment parameters. The VX650 measured marginally thicker corneal thickness (mean difference of central corneal thickness (CCT) of -4.0 ± 6.3 µm). Statistically significant differences in apical corneal thickness were noted for post-refractive surgery patients. K1 and K2 values showed significant variations in keratoconus but were not clinically significant. Anterior chamber depth was similar between devices, though volume differences were observed post-surgery. ICC indicated good to excellent repeatability (above 0.75) for most parameters.</p><p><strong>Conclusion: </strong>Remidio Visionix VX650 showed good agreement with Pentacam for key anterior segment measurements, confirming its reliability for corneal diagnostics. Additionally, its multifunctional capabilities offer potential for streamlining workflows, especially in clinics seeking efficient, comprehensive eye assessments.</p>","PeriodicalId":73096,"journal":{"name":"Frontiers in ophthalmology","volume":"5 ","pages":"1516423"},"PeriodicalIF":0.9000,"publicationDate":"2025-02-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11879984/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fopht.2025.1516423","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Advancements in anterior segment imaging have significantly improved the precision of corneal diagnostics, aiding ocular conditions management. The study evaluated the agreement between two Scheimpflug devices-Pentacam and Remidio Visionix 650 (VX650)-in assessing corneal and anterior chamber parameters in patients with and without corneal abnormalities.

Materials and methods: This prospective study at a tertiary eye care hospital measured pachymetry values (central corneal thickness, thinnest point), front corneal surface parameters (keratometry, curvature, astigmatism), back corneal surface parameters (elevation, curvature) and anterior chamber parameters (depth, volume, angle). Bland-Altman analysis assessed agreement between the devices, and the Intraclass Correlation Coefficient (ICC) evaluated VX650's repeatability.

Results: Of the 202 patients (mean age was 25 ± 4.33 years), 141 eyes were classified as normal and 61 as abnormal (41 keratoconus, 17 post-LASIK, 3 post-ICL). Bland-Altman analysis showed good agreement between Pentacam and VX650 for anterior segment parameters. The VX650 measured marginally thicker corneal thickness (mean difference of central corneal thickness (CCT) of -4.0 ± 6.3 µm). Statistically significant differences in apical corneal thickness were noted for post-refractive surgery patients. K1 and K2 values showed significant variations in keratoconus but were not clinically significant. Anterior chamber depth was similar between devices, though volume differences were observed post-surgery. ICC indicated good to excellent repeatability (above 0.75) for most parameters.

Conclusion: Remidio Visionix VX650 showed good agreement with Pentacam for key anterior segment measurements, confirming its reliability for corneal diagnostics. Additionally, its multifunctional capabilities offer potential for streamlining workflows, especially in clinics seeking efficient, comprehensive eye assessments.

Abstract Image

Abstract Image

Abstract Image

基于 Scheimpflug 的前段成像系统在正常角膜和异常角膜中的前瞻性比较:Oculus Pentacam 与 Remidio Visionix VX650。
目的:角膜前段影像技术的进步显著提高了角膜诊断的准确性,有助于眼部疾病的治疗。该研究评估了两种Scheimpflug装置pentacam和Remidio Visionix 650 (VX650)在评估有或无角膜异常患者的角膜和前房参数方面的一致性。材料与方法:本前瞻性研究在一家三级眼科医院进行,测量了角膜测厚值(角膜中央厚度、最薄点)、前角膜表面参数(角膜测厚、曲率、散光)、后角膜表面参数(仰角、曲率)和前房参数(深度、体积、角度)。Bland-Altman分析评估了设备之间的一致性,类内相关系数(ICC)评估了VX650的可重复性。结果:202例患者(平均年龄25±4.33岁)中,正常141眼,异常61眼(圆锥角膜41眼,lasik术后17眼,icl术后3眼)。Bland-Altman分析显示Pentacam和VX650在前段参数上有很好的一致性。VX650测量的角膜厚度略厚(角膜中央厚度(CCT)的平均差值为-4.0±6.3µm)。屈光手术后患者的角膜根尖厚度差异有统计学意义。K1和K2值在圆锥角膜中有显著差异,但无临床意义。不同装置的前房深度相似,但术后观察到体积差异。ICC显示大多数参数的重复性良好至优异(0.75以上)。结论:Remidio Visionix VX650与Pentacam在关键前段测量上具有良好的一致性,证实了其在角膜诊断中的可靠性。此外,它的多功能功能提供了简化工作流程的潜力,特别是在寻求高效,全面的眼科评估的诊所。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.50
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信