{"title":"Imaging and spectral performance of a wide-gap CdTe double-sided strip detector","authors":"Shunsaku Nagasawa , Takahiro Minami , Shin Watanabe , Tadayuki Takahashi","doi":"10.1016/j.nima.2025.170362","DOIUrl":null,"url":null,"abstract":"<div><div>The fourth flight of the Focusing Optics X-ray Solar Imager sounding rocket experiment (FOXSI-4) aimed to achieve the first imaging spectroscopic observations of mid-to-large class (<span><math><mo>≥</mo></math></span> GOES C5 class) solar flares, in contrast to the previous three flights that targeted relatively quiet regions of the Sun. To meet the emerging requirements for hard X-ray focal plane detectors for providing simultaneous diagnostics of spectrally (<span><math><mrow><mo><</mo><mn>1</mn></mrow></math></span> keV FWHM) and spatially (<span><math><mrow><mo><</mo><mn>50</mn><mspace></mspace><mi>μ</mi><mi>m</mi></mrow></math></span>) separated coronal and chromospheric emissions from solar flares, we developed a new strip-configuration detector called the wide-gap CdTe semiconductor double-sided strip detector (CdTe-DSD).</div><div>The wide-gap CdTe-DSD employs a unique design principle to enhance position resolution. This enhancement is realized by expanding the gaps between electrodes to induce charge-sharing across adjacent strip electrodes and using this sharing energy information for position reconstruction to a level finer than the strip-pitch. However, the detector response becomes complex and requires consideration of various factors, such as the charge loss due to wider gaps and the dependence on the depth of photon interaction. Thus, we developed an energy reconstruction method that fully leverages the energy information between adjacent strips and from both the cathode and anode sides, achieving an energy resolution of 0.75 keV (FWHM) at 14 keV. Furthermore, we conducted an X-ray scanning experiment using a synchrotron beam at Spring-8 to evaluate the detector response with a fine scale of 10 <span><math><mrow><mi>μ</mi><mi>m</mi></mrow></math></span>. Based on these results, we established a sub-strip position reconstruction method, demonstrating that X-rays interacting at the center of the gap can be determined with an accuracy of 20 <span><math><mrow><mi>μ</mi><mi>m</mi></mrow></math></span>, and even those at the strip center can be determined with an accuracy of 50 <span><math><mrow><mi>μ</mi><mi>m</mi></mrow></math></span>.</div></div>","PeriodicalId":19359,"journal":{"name":"Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment","volume":"1075 ","pages":"Article 170362"},"PeriodicalIF":1.5000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment","FirstCategoryId":"101","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0168900225001639","RegionNum":3,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0
Abstract
The fourth flight of the Focusing Optics X-ray Solar Imager sounding rocket experiment (FOXSI-4) aimed to achieve the first imaging spectroscopic observations of mid-to-large class ( GOES C5 class) solar flares, in contrast to the previous three flights that targeted relatively quiet regions of the Sun. To meet the emerging requirements for hard X-ray focal plane detectors for providing simultaneous diagnostics of spectrally ( keV FWHM) and spatially () separated coronal and chromospheric emissions from solar flares, we developed a new strip-configuration detector called the wide-gap CdTe semiconductor double-sided strip detector (CdTe-DSD).
The wide-gap CdTe-DSD employs a unique design principle to enhance position resolution. This enhancement is realized by expanding the gaps between electrodes to induce charge-sharing across adjacent strip electrodes and using this sharing energy information for position reconstruction to a level finer than the strip-pitch. However, the detector response becomes complex and requires consideration of various factors, such as the charge loss due to wider gaps and the dependence on the depth of photon interaction. Thus, we developed an energy reconstruction method that fully leverages the energy information between adjacent strips and from both the cathode and anode sides, achieving an energy resolution of 0.75 keV (FWHM) at 14 keV. Furthermore, we conducted an X-ray scanning experiment using a synchrotron beam at Spring-8 to evaluate the detector response with a fine scale of 10 . Based on these results, we established a sub-strip position reconstruction method, demonstrating that X-rays interacting at the center of the gap can be determined with an accuracy of 20 , and even those at the strip center can be determined with an accuracy of 50 .
期刊介绍:
Section A of Nuclear Instruments and Methods in Physics Research publishes papers on design, manufacturing and performance of scientific instruments with an emphasis on large scale facilities. This includes the development of particle accelerators, ion sources, beam transport systems and target arrangements as well as the use of secondary phenomena such as synchrotron radiation and free electron lasers. It also includes all types of instrumentation for the detection and spectrometry of radiations from high energy processes and nuclear decays, as well as instrumentation for experiments at nuclear reactors. Specialized electronics for nuclear and other types of spectrometry as well as computerization of measurements and control systems in this area also find their place in the A section.
Theoretical as well as experimental papers are accepted.