Effect of artificial aging on color change and mechanical properties of high consistency maxillofacial silicone elastomers.

IF 4.3 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Nuran Ozyemisci, Meral Kurt, Secil Karakoca Nemli
{"title":"Effect of artificial aging on color change and mechanical properties of high consistency maxillofacial silicone elastomers.","authors":"Nuran Ozyemisci, Meral Kurt, Secil Karakoca Nemli","doi":"10.1016/j.prosdent.2025.02.019","DOIUrl":null,"url":null,"abstract":"<p><strong>Statement of problem: </strong>Discoloration and the deterioration of mechanical properties lead to the replacement of maxillofacial prostheses. However, studies on the change in color and mechanical properties of high viscosity elastomers are lacking.</p><p><strong>Purpose: </strong>The purpose of this in vitro study was to evaluate the influence of artificial aging on the mechanical and optical properties of high-consistency silicone elastomer (HCR) and compare these properties with a widely used liquid silicone rubber (LSR).</p><p><strong>Material and methods: </strong>The specimens were fabricated from HCR (Derma-sil; Spectromatch Ltd.) and LSR (M511; Technovent Ltd.) silicone materials according to the manufacturers' instructions. A total of 325 specimens were used (50 for the tear test, 50 for the tensile and elongation test, 25 for the hardness test, 50 for absorption and solubility, and 50 for color and surface roughness measurements). Four subgroups of HCR with different Shore A hardness values (5, 20, 55, 70) were used. The specimens were subjected to tear (ASTM D624), tensile and elongation (ASTM D41215), hardness (ASTM D2240), absorption and solubility (ADA 12), surface roughness, and color change tests before and after the artificial aging process. Both water bath and ultraviolet (UV) chamber aging methods were used for the absorption and solubility tests. Contrast ratio and relative translucency parameters were also examined in addition to color change. Accelerated aging including UV light, humidity, and heat was carried out for 168 hours. The analyses were performed using 2-way factorial ANOVA and mixed ANOVA designs. Bonferroni multiple comparison test was used for significant results (α=.05).</p><p><strong>Results: </strong>The hardness values of all groups increased after aging except for the HCR subgroup with Shore A hardness 20 (P<.001). No significant change was found in surface roughness (P=.118), tear strength (P=.374), or elongation percentage (P=.053) after aging for all silicone groups. The only group with a significant change in tensile strength was HCR70 (P=.01). According to the water absorption results, absorption rates were significantly higher after artificial aging compared with bath aging in all silicone groups (P<.001) except LSR (P>.05). The aging method had significant effect on the water solubility of all groups (P<.001). Color change was significantly different among the silicone groups (P<.001).</p><p><strong>Conclusions: </strong>The HCR silicone elastomer tested in the study showed optical and mechanical properties compatible with those of a commonly used LSR silicone.</p>","PeriodicalId":16866,"journal":{"name":"Journal of Prosthetic Dentistry","volume":" ","pages":""},"PeriodicalIF":4.3000,"publicationDate":"2025-03-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthetic Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.prosdent.2025.02.019","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Statement of problem: Discoloration and the deterioration of mechanical properties lead to the replacement of maxillofacial prostheses. However, studies on the change in color and mechanical properties of high viscosity elastomers are lacking.

Purpose: The purpose of this in vitro study was to evaluate the influence of artificial aging on the mechanical and optical properties of high-consistency silicone elastomer (HCR) and compare these properties with a widely used liquid silicone rubber (LSR).

Material and methods: The specimens were fabricated from HCR (Derma-sil; Spectromatch Ltd.) and LSR (M511; Technovent Ltd.) silicone materials according to the manufacturers' instructions. A total of 325 specimens were used (50 for the tear test, 50 for the tensile and elongation test, 25 for the hardness test, 50 for absorption and solubility, and 50 for color and surface roughness measurements). Four subgroups of HCR with different Shore A hardness values (5, 20, 55, 70) were used. The specimens were subjected to tear (ASTM D624), tensile and elongation (ASTM D41215), hardness (ASTM D2240), absorption and solubility (ADA 12), surface roughness, and color change tests before and after the artificial aging process. Both water bath and ultraviolet (UV) chamber aging methods were used for the absorption and solubility tests. Contrast ratio and relative translucency parameters were also examined in addition to color change. Accelerated aging including UV light, humidity, and heat was carried out for 168 hours. The analyses were performed using 2-way factorial ANOVA and mixed ANOVA designs. Bonferroni multiple comparison test was used for significant results (α=.05).

Results: The hardness values of all groups increased after aging except for the HCR subgroup with Shore A hardness 20 (P<.001). No significant change was found in surface roughness (P=.118), tear strength (P=.374), or elongation percentage (P=.053) after aging for all silicone groups. The only group with a significant change in tensile strength was HCR70 (P=.01). According to the water absorption results, absorption rates were significantly higher after artificial aging compared with bath aging in all silicone groups (P<.001) except LSR (P>.05). The aging method had significant effect on the water solubility of all groups (P<.001). Color change was significantly different among the silicone groups (P<.001).

Conclusions: The HCR silicone elastomer tested in the study showed optical and mechanical properties compatible with those of a commonly used LSR silicone.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Prosthetic Dentistry
Journal of Prosthetic Dentistry 医学-牙科与口腔外科
CiteScore
7.00
自引率
13.00%
发文量
599
审稿时长
69 days
期刊介绍: The Journal of Prosthetic Dentistry is the leading professional journal devoted exclusively to prosthetic and restorative dentistry. The Journal is the official publication for 24 leading U.S. international prosthodontic organizations. The monthly publication features timely, original peer-reviewed articles on the newest techniques, dental materials, and research findings. The Journal serves prosthodontists and dentists in advanced practice, and features color photos that illustrate many step-by-step procedures. The Journal of Prosthetic Dentistry is included in Index Medicus and CINAHL.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信