V-PRO Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests

IF 2.6 4区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY
Anwar Borai, Wedyan Alsharif, Amirah Alhindi, Maha Alqahtani, Mohieldin Elsayid, Haitham Khalil, Salwa Al Marwani, Abobaker Yagoot, Janet Magjacot, Maha Al Meteiri, Rawan Alyamani, Hind Abdulhakim, Majid Al-Thaqafy
{"title":"V-PRO Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests","authors":"Anwar Borai,&nbsp;Wedyan Alsharif,&nbsp;Amirah Alhindi,&nbsp;Maha Alqahtani,&nbsp;Mohieldin Elsayid,&nbsp;Haitham Khalil,&nbsp;Salwa Al Marwani,&nbsp;Abobaker Yagoot,&nbsp;Janet Magjacot,&nbsp;Maha Al Meteiri,&nbsp;Rawan Alyamani,&nbsp;Hind Abdulhakim,&nbsp;Majid Al-Thaqafy","doi":"10.1002/jcla.70007","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>In accredited laboratories, each component of diagnostic products—such as laboratory instruments, reagents, and blood collection tubes must be validated before integration into routine patient testing. BD Vacutainers are commonly used in clinical laboratories compared to other blood collection tubes, while V-PRO tubes have recently been introduced to the market without prior laboratory validation. This study compares V-PRO tubes to BD Vacutainers to assess the validity of using V-PRO tubes for blood testing.</p>\n </section>\n \n <section>\n \n <h3> Materials and Methods</h3>\n \n <p>Blood samples were collected simultaneously into two different brands of tubes (V-PRO and BD) from 60 subjects. A standardized procedure was employed for sample collection, and analysis. A total of 28 chemistry tests and 20 immunoassays were analyzed using Abbott instruments, while high-performance liquid chromatography was used for testing glycated hemoglobin. The biases of V-PRO compared to BD were evaluated against current desirable quality specifications for bias derived from biological variation. For technical validation, a designated survey was distributed to various institutes using both tube types in their laboratories.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The V-PRO tube exhibited biases exceeding the desirable limits for CO<sub>2</sub> (3.2%), magnesium (2.0%), thyroid-stimulating hormone (11.7%), and estradiol (−8.5%). Survey results indicated a higher percentage of major pre-analytical, analytical, and post-analytical errors when using the V-PRO tube compared to the BD Vacutainer.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Laboratories currently using BD vacutainers should exercise caution if they intend to perform chemistry and immunoassay tests with V-PRO tubes. The technical validation outcomes for V-PRO were not acceptable due to significant faults identified in comparison to BD Vacutainer.</p>\n </section>\n </div>","PeriodicalId":15509,"journal":{"name":"Journal of Clinical Laboratory Analysis","volume":"39 6","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2025-03-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/jcla.70007","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Laboratory Analysis","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/jcla.70007","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background

In accredited laboratories, each component of diagnostic products—such as laboratory instruments, reagents, and blood collection tubes must be validated before integration into routine patient testing. BD Vacutainers are commonly used in clinical laboratories compared to other blood collection tubes, while V-PRO tubes have recently been introduced to the market without prior laboratory validation. This study compares V-PRO tubes to BD Vacutainers to assess the validity of using V-PRO tubes for blood testing.

Materials and Methods

Blood samples were collected simultaneously into two different brands of tubes (V-PRO and BD) from 60 subjects. A standardized procedure was employed for sample collection, and analysis. A total of 28 chemistry tests and 20 immunoassays were analyzed using Abbott instruments, while high-performance liquid chromatography was used for testing glycated hemoglobin. The biases of V-PRO compared to BD were evaluated against current desirable quality specifications for bias derived from biological variation. For technical validation, a designated survey was distributed to various institutes using both tube types in their laboratories.

Results

The V-PRO tube exhibited biases exceeding the desirable limits for CO2 (3.2%), magnesium (2.0%), thyroid-stimulating hormone (11.7%), and estradiol (−8.5%). Survey results indicated a higher percentage of major pre-analytical, analytical, and post-analytical errors when using the V-PRO tube compared to the BD Vacutainer.

Conclusions

Laboratories currently using BD vacutainers should exercise caution if they intend to perform chemistry and immunoassay tests with V-PRO tubes. The technical validation outcomes for V-PRO were not acceptable due to significant faults identified in comparison to BD Vacutainer.

Abstract Image

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Clinical Laboratory Analysis
Journal of Clinical Laboratory Analysis 医学-医学实验技术
CiteScore
5.60
自引率
7.40%
发文量
584
审稿时长
6-12 weeks
期刊介绍: Journal of Clinical Laboratory Analysis publishes original articles on newly developing modes of technology and laboratory assays, with emphasis on their application in current and future clinical laboratory testing. This includes reports from the following fields: immunochemistry and toxicology, hematology and hematopathology, immunopathology, molecular diagnostics, microbiology, genetic testing, immunohematology, and clinical chemistry.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信