{"title":"Performance of the cobas 5800 System for Hepatitis B virus DNA and Hepatitis C virus RNA quantification","authors":"Erwan Vo-Quang , Delphine Rosse , Valérie Ortonne , Olivia Garrigou , Patrick Ingiliz , Vincent Leroy , Jean-Michel Pawlotsky , Stéphane Chevaliez","doi":"10.1016/j.diagmicrobio.2025.116753","DOIUrl":null,"url":null,"abstract":"<div><div>Hepatitis B and C infections are an underdiagnosed global health problem. Measurement of HBV DNA or HCV RNA levels using nucleic acid-based molecular diagnostic assays has been established as the standard of care for assessing diagnosis, guiding the treatment decision, and evaluating responses to antiviral therapy.</div><div>In the present study, we examined the performance of the cobas 5800 System for HBV DNA and HCV RNA quantification in a large series of patients chronically infected.</div><div>Specificity of the cobas HBV and HCV Tests on the 5800 System was high (99.1 % and 100 %, respectively). Linearity using the AcroMetrix panels was excellent. Repeatability and intermediate precision coefficients of variation were within 5 %. Of the 334 clinical specimens tested in parallel on the cobas 5800 and cobas 4800 Systems for HBV and the m2000 RealTime or Alinity m Systems for HCV, only 12 (3.6 %) yielded discrepant results that were at or near the limit of quantification of the cobas 5800 assays. The correlation between viral load results was extremely high, and only weak bias were observed across the entire range of concentrations tested without clinical impact in patients who are eligible for antiviral therapy.</div><div>This comparison study demonstrated equivalent performance of the new cobas 5800 System compared with other molecular platforms widely used in clinical practice for HBV DNA and HCV RNA quantification. The cobas 5800 System can be confidently used in clinical practice. A few clinical specimens with low viral loads may be missed. Further studies are warranted to confirm or refute this finding.</div></div>","PeriodicalId":11329,"journal":{"name":"Diagnostic microbiology and infectious disease","volume":"112 2","pages":"Article 116753"},"PeriodicalIF":2.1000,"publicationDate":"2025-02-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diagnostic microbiology and infectious disease","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0732889325000768","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
引用次数: 0
Abstract
Hepatitis B and C infections are an underdiagnosed global health problem. Measurement of HBV DNA or HCV RNA levels using nucleic acid-based molecular diagnostic assays has been established as the standard of care for assessing diagnosis, guiding the treatment decision, and evaluating responses to antiviral therapy.
In the present study, we examined the performance of the cobas 5800 System for HBV DNA and HCV RNA quantification in a large series of patients chronically infected.
Specificity of the cobas HBV and HCV Tests on the 5800 System was high (99.1 % and 100 %, respectively). Linearity using the AcroMetrix panels was excellent. Repeatability and intermediate precision coefficients of variation were within 5 %. Of the 334 clinical specimens tested in parallel on the cobas 5800 and cobas 4800 Systems for HBV and the m2000 RealTime or Alinity m Systems for HCV, only 12 (3.6 %) yielded discrepant results that were at or near the limit of quantification of the cobas 5800 assays. The correlation between viral load results was extremely high, and only weak bias were observed across the entire range of concentrations tested without clinical impact in patients who are eligible for antiviral therapy.
This comparison study demonstrated equivalent performance of the new cobas 5800 System compared with other molecular platforms widely used in clinical practice for HBV DNA and HCV RNA quantification. The cobas 5800 System can be confidently used in clinical practice. A few clinical specimens with low viral loads may be missed. Further studies are warranted to confirm or refute this finding.
期刊介绍:
Diagnostic Microbiology and Infectious Disease keeps you informed of the latest developments in clinical microbiology and the diagnosis and treatment of infectious diseases. Packed with rigorously peer-reviewed articles and studies in bacteriology, immunology, immunoserology, infectious diseases, mycology, parasitology, and virology, the journal examines new procedures, unusual cases, controversial issues, and important new literature. Diagnostic Microbiology and Infectious Disease distinguished independent editorial board, consisting of experts from many medical specialties, ensures you extensive and authoritative coverage.