Image quality assessment of the CARE kV for pediatric CT angiography: A feasibility study of over 50% effective dose reduction

IF 1.7 4区 综合性期刊 Q2 MULTIDISCIPLINARY SCIENCES
Chunmei Fan , Runtong Lu , Dilisaer Naisula , Fei Bi , Xin Liu , Shui Wang , Jihang Sun
{"title":"Image quality assessment of the CARE kV for pediatric CT angiography: A feasibility study of over 50% effective dose reduction","authors":"Chunmei Fan ,&nbsp;Runtong Lu ,&nbsp;Dilisaer Naisula ,&nbsp;Fei Bi ,&nbsp;Xin Liu ,&nbsp;Shui Wang ,&nbsp;Jihang Sun","doi":"10.1016/j.jrras.2025.101386","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To evaluate whether the application of CARE kV automatic tube voltage modulation technology can reduce the radiation dose required for CT angiography (CTA) in children rapidly, conveniently, and effectively while maintaining image quality.</div></div><div><h3>Methods</h3><div>57 pediatric abdominal CTA were enrolled in the study and control groups. The tube voltage was automatically modulated using CARE kV. Dose-saving optimization indices used for the study and control groups were 12 (preference for a lower tube voltage) and 3 (preference for a higher tube voltage), respectively. Tube voltage and effective dose (ED) were recorded for the two groups, and a multiple linear regression analysis was used to scrutinize factors influencing voltage selection and radiation dose. Two reviewers subjectively evaluated the image quality using a 5-point Likert scale (3 points: qualified; 5 points: best), and the contrast-to-noise ratio (CNR) was calculated.</div></div><div><h3>Results</h3><div>The most commonly selected tube voltage was 70 kV (36/57, 63.16%) and 100 kV (30/57, 52.63%) in the study and control groups, respectively (<em>p</em> &lt; 0.001). The ED of the study group was 1.58 ± 1.05 mSv, which was 57.07% lower than that in the control group (<em>p</em> &lt; 0.001). Both voltage selection and ED exhibited a negative correlation with the DI value. The small artery display ability of the two groups was 4.05 ± 0.64 and 4.11 ± 0.65 (<em>p</em> = 0.31). No statistically significant difference was found for the aortic CNR between the two groups (<em>p</em> = 0.15).</div></div><div><h3>Conclusion</h3><div>CARE kV is a rapid, convenient, and effective way to reduce the radiation dose in CTA by &gt;50% in children.</div></div>","PeriodicalId":16920,"journal":{"name":"Journal of Radiation Research and Applied Sciences","volume":"18 2","pages":"Article 101386"},"PeriodicalIF":1.7000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiation Research and Applied Sciences","FirstCategoryId":"103","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1687850725000986","RegionNum":4,"RegionCategory":"综合性期刊","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MULTIDISCIPLINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

To evaluate whether the application of CARE kV automatic tube voltage modulation technology can reduce the radiation dose required for CT angiography (CTA) in children rapidly, conveniently, and effectively while maintaining image quality.

Methods

57 pediatric abdominal CTA were enrolled in the study and control groups. The tube voltage was automatically modulated using CARE kV. Dose-saving optimization indices used for the study and control groups were 12 (preference for a lower tube voltage) and 3 (preference for a higher tube voltage), respectively. Tube voltage and effective dose (ED) were recorded for the two groups, and a multiple linear regression analysis was used to scrutinize factors influencing voltage selection and radiation dose. Two reviewers subjectively evaluated the image quality using a 5-point Likert scale (3 points: qualified; 5 points: best), and the contrast-to-noise ratio (CNR) was calculated.

Results

The most commonly selected tube voltage was 70 kV (36/57, 63.16%) and 100 kV (30/57, 52.63%) in the study and control groups, respectively (p < 0.001). The ED of the study group was 1.58 ± 1.05 mSv, which was 57.07% lower than that in the control group (p < 0.001). Both voltage selection and ED exhibited a negative correlation with the DI value. The small artery display ability of the two groups was 4.05 ± 0.64 and 4.11 ± 0.65 (p = 0.31). No statistically significant difference was found for the aortic CNR between the two groups (p = 0.15).

Conclusion

CARE kV is a rapid, convenient, and effective way to reduce the radiation dose in CTA by >50% in children.
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
5.90%
发文量
130
审稿时长
16 weeks
期刊介绍: Journal of Radiation Research and Applied Sciences provides a high quality medium for the publication of substantial, original and scientific and technological papers on the development and applications of nuclear, radiation and isotopes in biology, medicine, drugs, biochemistry, microbiology, agriculture, entomology, food technology, chemistry, physics, solid states, engineering, environmental and applied sciences.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信