Within- and between-day test-retest reliability of responses to rapid bilateral anterolateral magnetic phrenic nerve stimulation in healthy humans (ReStim).
Kyle G P J M Boyle, Andrea A Beglinger, Heinrich Häusler, Anna Stahel, Esther I Schwarz, Christina M Spengler
{"title":"Within- and between-day test-retest reliability of responses to rapid bilateral anterolateral magnetic phrenic nerve stimulation in healthy humans (ReStim).","authors":"Kyle G P J M Boyle, Andrea A Beglinger, Heinrich Häusler, Anna Stahel, Esther I Schwarz, Christina M Spengler","doi":"10.3389/fphys.2025.1481766","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Mechanical ventilation can lead to lung injury and diaphragmatic dysfunction. Rapid bilateral anterolateral magnetic phrenic nerve stimulation (rBAMPS) may attenuate both of the aforementioned issues by inducing diaphragm activation. However, in order for rBAMPS to become part of standard of care, the reliability of inspiratory responses to rBAMPS needs to be established.</p><p><strong>Methods: </strong>Eighteen healthy participants (9F) underwent five blocks of 1-s rBAMPS at 25 Hz starting at 20% of maximal stimulator output with 10% increments. Three blocks were completed on the same day to test within-day reliability, and two additional blocks were each completed on subsequent days to test between-day reliability. Mean transdiaphragmatic pressure (P<sub>di,mean</sub>), tidal volume (V<sub>T</sub>), discomfort, pain, and paresthesia were recorded for each rBAMPS. Relative and absolute reliability of both P<sub>di,mean</sub> and V<sub>T</sub> were quantified by calculating intraclass correlation coefficients (ICC) and standard error of measurements (SE<sub>M</sub>), respectively. An ordinal regression was used to determine changes of sensory ratings within and between days.</p><p><strong>Results: </strong>At all stimulator outputs, within-day P<sub>di,mean</sub> displayed \"good\" reliability (ICC range 0.78-0.89). Between days, P<sub>di,mean</sub> reliability was also \"good\" (ICC range 0.79-0.87) at stimulator outputs of 20%-50% of maximum, but \"moderate\" (ICC range 0.56-0.72) at stimulator outputs of 60%-100%. SE<sub>M</sub> for P<sub>di,mean</sub> within day ranged from 0.9 to 3.4 across tested stimulator outputs and increased on average by 1.4 ± 0.9 between days. The V<sub>T</sub> reliability was \"good\" to \"excellent\" within (ICC range 0.82-0.94) and between (ICC range 0.81-0.96) days at all stimulator outputs. SE<sub>M</sub> for V<sub>T</sub> within day ranged from 0.08 to 0.36 and from 0.11 to 0.30 between days and tended to be larger at stimulator outputs greater than 50% of maximum. Subsequent blocks within day were associated with decreased discomfort and pain (P ≤ 0.043), while subsequent days were associated with decreased discomfort and paresthesia (P < 0.001).</p><p><strong>Discussion: </strong>rBAMPS appears to induce reliable diaphragmatic contractions, while select sensory responses become blunted over repeated stimulations. However, as reliability is slightly lower between days compared to within day, stimulation parameters may need to be adjusted to achieve similar responses on different days.</p>","PeriodicalId":12477,"journal":{"name":"Frontiers in Physiology","volume":"16 ","pages":"1481766"},"PeriodicalIF":3.2000,"publicationDate":"2025-02-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11850319/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in Physiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3389/fphys.2025.1481766","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"PHYSIOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Mechanical ventilation can lead to lung injury and diaphragmatic dysfunction. Rapid bilateral anterolateral magnetic phrenic nerve stimulation (rBAMPS) may attenuate both of the aforementioned issues by inducing diaphragm activation. However, in order for rBAMPS to become part of standard of care, the reliability of inspiratory responses to rBAMPS needs to be established.
Methods: Eighteen healthy participants (9F) underwent five blocks of 1-s rBAMPS at 25 Hz starting at 20% of maximal stimulator output with 10% increments. Three blocks were completed on the same day to test within-day reliability, and two additional blocks were each completed on subsequent days to test between-day reliability. Mean transdiaphragmatic pressure (Pdi,mean), tidal volume (VT), discomfort, pain, and paresthesia were recorded for each rBAMPS. Relative and absolute reliability of both Pdi,mean and VT were quantified by calculating intraclass correlation coefficients (ICC) and standard error of measurements (SEM), respectively. An ordinal regression was used to determine changes of sensory ratings within and between days.
Results: At all stimulator outputs, within-day Pdi,mean displayed "good" reliability (ICC range 0.78-0.89). Between days, Pdi,mean reliability was also "good" (ICC range 0.79-0.87) at stimulator outputs of 20%-50% of maximum, but "moderate" (ICC range 0.56-0.72) at stimulator outputs of 60%-100%. SEM for Pdi,mean within day ranged from 0.9 to 3.4 across tested stimulator outputs and increased on average by 1.4 ± 0.9 between days. The VT reliability was "good" to "excellent" within (ICC range 0.82-0.94) and between (ICC range 0.81-0.96) days at all stimulator outputs. SEM for VT within day ranged from 0.08 to 0.36 and from 0.11 to 0.30 between days and tended to be larger at stimulator outputs greater than 50% of maximum. Subsequent blocks within day were associated with decreased discomfort and pain (P ≤ 0.043), while subsequent days were associated with decreased discomfort and paresthesia (P < 0.001).
Discussion: rBAMPS appears to induce reliable diaphragmatic contractions, while select sensory responses become blunted over repeated stimulations. However, as reliability is slightly lower between days compared to within day, stimulation parameters may need to be adjusted to achieve similar responses on different days.
期刊介绍:
Frontiers in Physiology is a leading journal in its field, publishing rigorously peer-reviewed research on the physiology of living systems, from the subcellular and molecular domains to the intact organism, and its interaction with the environment. Field Chief Editor George E. Billman at the Ohio State University Columbus is supported by an outstanding Editorial Board of international researchers. This multidisciplinary open-access journal is at the forefront of disseminating and communicating scientific knowledge and impactful discoveries to researchers, academics, clinicians and the public worldwide.