Diurnal variations in corneal biomechanics in healthy young adults.

IF 1.7 4区 医学 Q3 OPHTHALMOLOGY
DaTian Zhu, LuMeng Wang, ZhanXin Qu, Pu Wang, YongYi Yuan, Lan Yang, XinYu Yao, XiaoBo Zheng, HengLi Lian, Chong Wang, JunJie Wang, YuFeng Ye, Ahmed Elsheikh, FangJun Bao, JunHua Li
{"title":"Diurnal variations in corneal biomechanics in healthy young adults.","authors":"DaTian Zhu, LuMeng Wang, ZhanXin Qu, Pu Wang, YongYi Yuan, Lan Yang, XinYu Yao, XiaoBo Zheng, HengLi Lian, Chong Wang, JunJie Wang, YuFeng Ye, Ahmed Elsheikh, FangJun Bao, JunHua Li","doi":"10.1186/s12886-025-03913-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To assess the diurnal changes in corneal biomechanics in healthy young adults.</p><p><strong>Methods: </strong>At Wenzhou Medical University Eye Hospital in China, a prospective case series. Each healthy subject had six Corvis ST examinations, from 10:00 on a first day to 6:00 on the following day, at 4-hour intervals. The data collected included the biomechanically corrected IOP (bIOP), the central corneal thickness (CCT) and six dynamic corneal response (DCR) parameters, namely deformation amplitude (DA), the ratio between DA values at the apex and 2 mm from the apex (DAR2), integrated inverse radius (IIR), the stiffness parameter at first applanation (SP-A1), Corvis biomechanical index (CBI) and the updated stress-strain index (SSIv2). A multilevel model was used to assess the changes in DCR parameters of patients over time during the examination period.</p><p><strong>Results: </strong>Sixty-one subjects were included and intra-operator reproducibility was good for most DCR parameters at the 6 time points. DA, DAR2, IIR and CBI showed small fluctuations during the 24-hour cycle with mean values of 0.1, 0.16, 0.34, and 0.04 units, respectively, while SP-A1 showed higher variations averaging 20.57 units and SSIv2 had the least fluctuation within 0.10 units during 24-hour cycle. After correcting for the effects of CCT and bIOP, the diurnal fluctuations in IIR, CBI and SSIv2 were not statistically significant, unlike the fluctuations in DA, DAR2 and SP-A1, which remained significant.</p><p><strong>Conclusions: </strong>Compared with the DA, DAR2, IIR and CBI, SP-A1 was more affected while SSIv2 was the less affected by the diurnal variations and associated corneal edema.</p>","PeriodicalId":9058,"journal":{"name":"BMC Ophthalmology","volume":"25 1","pages":"90"},"PeriodicalIF":1.7000,"publicationDate":"2025-02-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11863582/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"BMC Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s12886-025-03913-3","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To assess the diurnal changes in corneal biomechanics in healthy young adults.

Methods: At Wenzhou Medical University Eye Hospital in China, a prospective case series. Each healthy subject had six Corvis ST examinations, from 10:00 on a first day to 6:00 on the following day, at 4-hour intervals. The data collected included the biomechanically corrected IOP (bIOP), the central corneal thickness (CCT) and six dynamic corneal response (DCR) parameters, namely deformation amplitude (DA), the ratio between DA values at the apex and 2 mm from the apex (DAR2), integrated inverse radius (IIR), the stiffness parameter at first applanation (SP-A1), Corvis biomechanical index (CBI) and the updated stress-strain index (SSIv2). A multilevel model was used to assess the changes in DCR parameters of patients over time during the examination period.

Results: Sixty-one subjects were included and intra-operator reproducibility was good for most DCR parameters at the 6 time points. DA, DAR2, IIR and CBI showed small fluctuations during the 24-hour cycle with mean values of 0.1, 0.16, 0.34, and 0.04 units, respectively, while SP-A1 showed higher variations averaging 20.57 units and SSIv2 had the least fluctuation within 0.10 units during 24-hour cycle. After correcting for the effects of CCT and bIOP, the diurnal fluctuations in IIR, CBI and SSIv2 were not statistically significant, unlike the fluctuations in DA, DAR2 and SP-A1, which remained significant.

Conclusions: Compared with the DA, DAR2, IIR and CBI, SP-A1 was more affected while SSIv2 was the less affected by the diurnal variations and associated corneal edema.

目的:评估健康年轻人角膜生物力学的昼夜变化:在中国温州医科大学附属眼科医院进行前瞻性病例系列研究。每个健康受试者从第一天的 10:00 到第二天的 6:00,每隔 4 小时进行六次 Corvis ST 检查。收集的数据包括生物力学校正后的眼压(bIOP)、角膜中央厚度(CCT)和六个动态角膜反应(DCR)参数,即形变幅度(DA)、角膜顶点和距角膜顶点2毫米处的DA值之比(DAR2)、综合反比半径(IIR)、首次角膜屈光检查时的硬度参数(SP-A1)、Corvis生物力学指数(CBI)和更新的应力应变指数(SSIv2)。采用多层次模型评估患者在检查期间DCR参数随时间的变化:结果:共纳入 61 名受试者,在 6 个时间点上,大多数 DCR 参数的术者内重现性良好。DA、DAR2、IIR和CBI在24小时周期内波动较小,平均值分别为0.1、0.16、0.34和0.04个单位,而SP-A1波动较大,平均为20.57个单位,SSIv2在24小时周期内波动最小,为0.10个单位。在校正了 CCT 和 bIOP 的影响后,IIR、CBI 和 SSIv2 的昼夜波动在统计学上并不显著,而 DA、DAR2 和 SP-A1 的波动仍然显著:与 DA、DAR2、IIR 和 CBI 相比,SP-A1 受昼夜变化和相关角膜水肿的影响更大,而 SSIv2 受昼夜变化和相关角膜水肿的影响较小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
BMC Ophthalmology
BMC Ophthalmology OPHTHALMOLOGY-
CiteScore
3.40
自引率
5.00%
发文量
441
审稿时长
6-12 weeks
期刊介绍: BMC Ophthalmology is an open access, peer-reviewed journal that considers articles on all aspects of the prevention, diagnosis and management of eye disorders, as well as related molecular genetics, pathophysiology, and epidemiology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信