High-Resolution EEG Amplifiers Are Feasible for Electrocochleography Without Time Restriction.

IF 2.1 Q1 AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY
Florian Josef Schertenleib, Sabine Hochmuth, Jana Annina Müller, Pascale Sandmann, Andreas Radeloff
{"title":"High-Resolution EEG Amplifiers Are Feasible for Electrocochleography Without Time Restriction.","authors":"Florian Josef Schertenleib, Sabine Hochmuth, Jana Annina Müller, Pascale Sandmann, Andreas Radeloff","doi":"10.3390/audiolres15010008","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>The gold standard for electrocochleography (ECochG) is using dedicated recording devices for auditory evoked potentials. However, these have a very limited time window for recording. The aim of this study is to evaluate EEG amplifiers for ECochG, in particular for recording cochlear microphonics (CMs) without time restriction.</p><p><strong>Methods: </strong>Three high-resolution EEG amplifiers and different types of electrodes were analyzed and compared with a clinical system for recording auditory evoked potentials. For this, CMs were recorded after stimulation with various stimuli in a dummy and in human subjects. In the latter, recordings were made from the tympanic membrane and, during otosurgical procedures, from the promontory. Our evaluation focused on comparing signal amplifiers and electrode types, considering the signal-to-noise ratio, recording characteristics, and measurement reliability.</p><p><strong>Results: </strong>Using a dummy model, we observed significant differences among devices, electrode types, and stimulus frequencies. These findings were subsequently confirmed in human participant measurements. Nevertheless, EEG amplifiers proved to be feasible for ECochG recordings and offered a recording fidelity comparable to proprietary clinical methods. Importantly, with EEG amplifiers, we were able to record cochlear potentials in response to speech stimuli, revealing a strong correlation (r = 0.78) between recorded signals and the input stimulus.</p><p><strong>Conclusions: </strong>Our findings indicate that high resolution EEG amplifiers are suitable for recording cochlear potentials, in particular, CMs. This allows for evaluating cochlear signals in response to extended stimuli, in particular, speech stimuli.</p>","PeriodicalId":44133,"journal":{"name":"Audiology Research","volume":"15 1","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2025-01-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11851963/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Audiology Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/audiolres15010008","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: The gold standard for electrocochleography (ECochG) is using dedicated recording devices for auditory evoked potentials. However, these have a very limited time window for recording. The aim of this study is to evaluate EEG amplifiers for ECochG, in particular for recording cochlear microphonics (CMs) without time restriction.

Methods: Three high-resolution EEG amplifiers and different types of electrodes were analyzed and compared with a clinical system for recording auditory evoked potentials. For this, CMs were recorded after stimulation with various stimuli in a dummy and in human subjects. In the latter, recordings were made from the tympanic membrane and, during otosurgical procedures, from the promontory. Our evaluation focused on comparing signal amplifiers and electrode types, considering the signal-to-noise ratio, recording characteristics, and measurement reliability.

Results: Using a dummy model, we observed significant differences among devices, electrode types, and stimulus frequencies. These findings were subsequently confirmed in human participant measurements. Nevertheless, EEG amplifiers proved to be feasible for ECochG recordings and offered a recording fidelity comparable to proprietary clinical methods. Importantly, with EEG amplifiers, we were able to record cochlear potentials in response to speech stimuli, revealing a strong correlation (r = 0.78) between recorded signals and the input stimulus.

Conclusions: Our findings indicate that high resolution EEG amplifiers are suitable for recording cochlear potentials, in particular, CMs. This allows for evaluating cochlear signals in response to extended stimuli, in particular, speech stimuli.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Audiology Research
Audiology Research AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY-
CiteScore
2.30
自引率
23.50%
发文量
56
审稿时长
11 weeks
期刊介绍: The mission of Audiology Research is to publish contemporary, ethical, clinically relevant scientific researches related to the basic science and clinical aspects of the auditory and vestibular system and diseases of the ear that can be used by clinicians, scientists and specialists to improve understanding and treatment of patients with audiological and neurotological disorders.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信