Rapid Microfluidic Ion-Exchange Optode System for Point-of-Care Determination of Sodium Concentration in Serum.

IF 4.9 3区 工程技术 Q1 CHEMISTRY, ANALYTICAL
Kuan-Hsun Huang, Cheng-Xue Yu, Chia-Chun Lee, Chin-Chung Tseng, Lung-Ming Fu
{"title":"Rapid Microfluidic Ion-Exchange Optode System for Point-of-Care Determination of Sodium Concentration in Serum.","authors":"Kuan-Hsun Huang, Cheng-Xue Yu, Chia-Chun Lee, Chin-Chung Tseng, Lung-Ming Fu","doi":"10.3390/bios15020104","DOIUrl":null,"url":null,"abstract":"<p><p>A microfluidic system for detecting sodium ions (Na<sup>+</sup>) has been developed, incorporating a micro finger-pump chip and a micro-spectrometer platform to measure Na<sup>+</sup> concentration in human serum. A small volume (10 μL) of serum sample is introduced into the microchip and reacted with a preloaded reagent mixture through a two-step finger-pump actuation process. The resulting purple complex is directed into the detection area of the chip and analyzed using the micro-spectrometer at wavelengths of 555 and 666 nm. The Na<sup>+</sup> concentration is then inversely derived from the measured A<sub>555</sub>/A<sub>666</sub> absorbance ratio using self-written software installed on a Raspberry Pi. The entire detection process is completed in just 3 min, offering a significant advantage in meeting clinical needs compared to the traditional reporting turnaround time of several hours in medical institutions. The experimental results indicate a linear relationship between the measured absorbance ratio and Na<sup>+</sup> concentration within the range of 1-200 mM, with a correlation coefficient of R<sup>2</sup> = 0.9989. Additionally, the detection results from 60 serum samples collected from chronic kidney disease (CKD) patients showed a strong agreement with those obtained using the conventional indirect ion-selective electrode (ISE) method, achieving a correlation coefficient of R<sup>2</sup> = 0.9885 and an average recovery rate of 99.4%. In summary, the proposed system provides a practical, affordable, and rapid alternative to conventional Na<sup>+</sup> detection methods, making it highly promising for point-of-care (POC) testing applications.</p>","PeriodicalId":48608,"journal":{"name":"Biosensors-Basel","volume":"15 2","pages":""},"PeriodicalIF":4.9000,"publicationDate":"2025-02-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11853684/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biosensors-Basel","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.3390/bios15020104","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

A microfluidic system for detecting sodium ions (Na+) has been developed, incorporating a micro finger-pump chip and a micro-spectrometer platform to measure Na+ concentration in human serum. A small volume (10 μL) of serum sample is introduced into the microchip and reacted with a preloaded reagent mixture through a two-step finger-pump actuation process. The resulting purple complex is directed into the detection area of the chip and analyzed using the micro-spectrometer at wavelengths of 555 and 666 nm. The Na+ concentration is then inversely derived from the measured A555/A666 absorbance ratio using self-written software installed on a Raspberry Pi. The entire detection process is completed in just 3 min, offering a significant advantage in meeting clinical needs compared to the traditional reporting turnaround time of several hours in medical institutions. The experimental results indicate a linear relationship between the measured absorbance ratio and Na+ concentration within the range of 1-200 mM, with a correlation coefficient of R2 = 0.9989. Additionally, the detection results from 60 serum samples collected from chronic kidney disease (CKD) patients showed a strong agreement with those obtained using the conventional indirect ion-selective electrode (ISE) method, achieving a correlation coefficient of R2 = 0.9885 and an average recovery rate of 99.4%. In summary, the proposed system provides a practical, affordable, and rapid alternative to conventional Na+ detection methods, making it highly promising for point-of-care (POC) testing applications.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Biosensors-Basel
Biosensors-Basel Biochemistry, Genetics and Molecular Biology-Clinical Biochemistry
CiteScore
6.60
自引率
14.80%
发文量
983
审稿时长
11 weeks
期刊介绍: Biosensors (ISSN 2079-6374) provides an advanced forum for studies related to the science and technology of biosensors and biosensing. It publishes original research papers, comprehensive reviews and communications. Our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible. There is no restriction on the length of the papers. The full experimental details must be provided so that the results can be reproduced. Electronic files and software regarding the full details of the calculation or experimental procedure, if unable to be published in a normal way, can be deposited as supplementary electronic material.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信