{"title":"Effect of suture anchor type, eyelet configuration, and loading condition on suture failure: An in vitro study.","authors":"Obinna P Fidelis, Pierre-Yves Mulon, David E Anderson, Dustin L Crouch","doi":"10.1111/vsu.14231","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To determine the effect of suture anchor type, eyelet configuration (raised or embedded), and loading condition (without or with cyclic loading) on failure load and failure mode of sutures in anchors with embedded and raised eyelets.</p><p><strong>Study design: </strong>Controlled laboratory study.</p><p><strong>Sample population: </strong>A total of 4 unique clinical anchor+suture combinations (n = 40 total); eyebolt screw reference samples (n = 30 total).</p><p><strong>Methods: </strong>A total of 10 samples of each anchor+suture combination were either (a) loaded to failure (n = 5) or (b) loaded from 0 N to 60 N for 1000 cycles then loaded to failure (n = 5). As a \"best-case\" reference, an eyebolt screw with a smooth, round cross-section was tested under the same two loading conditions and the same sutures (n = 10 per suture) used with the clinical anchors. The percent difference in the maximum load at failure ( <math> <semantics><mrow><mo>%</mo> <mo>∆</mo> <msub><mi>F</mi> <mi>max</mi></msub> </mrow> <annotation>$$ \\%\\Delta {F}_{\\mathrm{max}} $$</annotation></semantics> </math> ) between each anchor and suture-matched eyebolt screw was used to quantify the effect of each anchor on suture strength.</p><p><strong>Results: </strong>There was no effect of loading condition (i.e., without or with cyclic loading) on F<sub>max</sub>. The Anika anchor had a significantly lower <math> <semantics><mrow><mo>%</mo> <mo>∆</mo> <msub><mi>F</mi> <mi>max</mi></msub> </mrow> <annotation>$$ \\%\\Delta {F}_{\\mathrm{max}} $$</annotation></semantics> </math> than the Arthrex anchor (p = .015), IMEX anchor (p = .004) and Jorvet anchor (p < .001). A total of 14 out of 20 samples of anchors with raised eyelets failed at the suture mid-section, whereas all 20 samples of anchors with embedded eyelets failed at the suture knot.</p><p><strong>Conclusion: </strong>Anchors with raised eyelets had more failures at the suture mid-section indicative of wear and/or cutting of the suture against the eyelet.</p><p><strong>Clinical significance: </strong>Eyelet geometry and surface finish of metallic screw-type anchors have effects on suture strength.</p>","PeriodicalId":23667,"journal":{"name":"Veterinary Surgery","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2025-02-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary Surgery","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.1111/vsu.14231","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To determine the effect of suture anchor type, eyelet configuration (raised or embedded), and loading condition (without or with cyclic loading) on failure load and failure mode of sutures in anchors with embedded and raised eyelets.
Study design: Controlled laboratory study.
Sample population: A total of 4 unique clinical anchor+suture combinations (n = 40 total); eyebolt screw reference samples (n = 30 total).
Methods: A total of 10 samples of each anchor+suture combination were either (a) loaded to failure (n = 5) or (b) loaded from 0 N to 60 N for 1000 cycles then loaded to failure (n = 5). As a "best-case" reference, an eyebolt screw with a smooth, round cross-section was tested under the same two loading conditions and the same sutures (n = 10 per suture) used with the clinical anchors. The percent difference in the maximum load at failure ( ) between each anchor and suture-matched eyebolt screw was used to quantify the effect of each anchor on suture strength.
Results: There was no effect of loading condition (i.e., without or with cyclic loading) on Fmax. The Anika anchor had a significantly lower than the Arthrex anchor (p = .015), IMEX anchor (p = .004) and Jorvet anchor (p < .001). A total of 14 out of 20 samples of anchors with raised eyelets failed at the suture mid-section, whereas all 20 samples of anchors with embedded eyelets failed at the suture knot.
Conclusion: Anchors with raised eyelets had more failures at the suture mid-section indicative of wear and/or cutting of the suture against the eyelet.
Clinical significance: Eyelet geometry and surface finish of metallic screw-type anchors have effects on suture strength.
期刊介绍:
Veterinary Surgery, the official publication of the American College of Veterinary Surgeons and European College of Veterinary Surgeons, is a source of up-to-date coverage of surgical and anesthetic management of animals, addressing significant problems in veterinary surgery with relevant case histories and observations.
It contains original, peer-reviewed articles that cover developments in veterinary surgery, and presents the most current review of the field, with timely articles on surgical techniques, diagnostic aims, care of infections, and advances in knowledge of metabolism as it affects the surgical patient. The journal places new developments in perspective, encompassing new concepts and peer commentary to help better understand and evaluate the surgical patient.