Mansureh Ziaiee, Heydar Sadeghi, Mohammad Taghi Karimi, Masoud Rafiaei
{"title":"Assessment of Mandibular Kinematic Variables using a Motion Analysis System and a Regular Mobile Phone.","authors":"Mansureh Ziaiee, Heydar Sadeghi, Mohammad Taghi Karimi, Masoud Rafiaei","doi":"10.31661/jbpe.v0i0.2210-1555","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The development of a standard motion capture system is needed since the measurement of temporomandibular disorders is time-consuming and costly using laboratory tools.</p><p><strong>Objective: </strong>The current study aimed to investigate the mandibular kinematic variables using a regular mobile phone and the motion analysis system.</p><p><strong>Material and methods: </strong>In this quasi-laboratory and comparative study, ten healthy individuals participated, and three mobile cameras, nine red markers, and Kinovea software were also used to investigate the mandibular kinematic variables. Nine light reflective markers were used to check the sensitivity, accuracy, and reliability of the proposed system. The motion was analyzed using seven motion analysis infrared cameras and Qualisys Track Manager (QTM) software. Two other raters analyzed the kinematic variables obtained from the mobile to measure intra- and inter-rater reliability.</p><p><strong>Results: </strong>Pearson's correlation coefficient was obtained at 0.98, 0.75, 0.98, and 0.96, showing a high correlation. The accuracy and reliability validation tests showed an average error and an accuracy of 0.156 mm and 95%, respectively, with a mobile phone. The Intra Class Correlation coefficient showed a high internal correlation in the mentioned confidence interval (0.98 and 0.81, and 0.96 and 0.97). The intraclass correlation coefficient method also showed 97% inter-raster reliability.</p><p><strong>Conclusion: </strong>Mobile phones as a new system can evaluate the kinematic variables of mandibular disorders with appropriate accuracy and reliability.</p>","PeriodicalId":38035,"journal":{"name":"Journal of Biomedical Physics and Engineering","volume":"15 1","pages":"67-76"},"PeriodicalIF":0.0000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11833152/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Biomedical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31661/jbpe.v0i0.2210-1555","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The development of a standard motion capture system is needed since the measurement of temporomandibular disorders is time-consuming and costly using laboratory tools.
Objective: The current study aimed to investigate the mandibular kinematic variables using a regular mobile phone and the motion analysis system.
Material and methods: In this quasi-laboratory and comparative study, ten healthy individuals participated, and three mobile cameras, nine red markers, and Kinovea software were also used to investigate the mandibular kinematic variables. Nine light reflective markers were used to check the sensitivity, accuracy, and reliability of the proposed system. The motion was analyzed using seven motion analysis infrared cameras and Qualisys Track Manager (QTM) software. Two other raters analyzed the kinematic variables obtained from the mobile to measure intra- and inter-rater reliability.
Results: Pearson's correlation coefficient was obtained at 0.98, 0.75, 0.98, and 0.96, showing a high correlation. The accuracy and reliability validation tests showed an average error and an accuracy of 0.156 mm and 95%, respectively, with a mobile phone. The Intra Class Correlation coefficient showed a high internal correlation in the mentioned confidence interval (0.98 and 0.81, and 0.96 and 0.97). The intraclass correlation coefficient method also showed 97% inter-raster reliability.
Conclusion: Mobile phones as a new system can evaluate the kinematic variables of mandibular disorders with appropriate accuracy and reliability.
期刊介绍:
The Journal of Biomedical Physics and Engineering (JBPE) is a bimonthly peer-reviewed English-language journal that publishes high-quality basic sciences and clinical research (experimental or theoretical) broadly concerned with the relationship of physics to medicine and engineering.