FACTORS AFFECTING ASPIRATION VOLUMES OF BACK-FLUSH NEEDLES AND INFUSION VOLUMES OF VITRECTOMY DEVICE.

IF 2.3 2区 医学 Q2 OPHTHALMOLOGY
Masaharu Mizuno, Takashi Koto, Kosuke Nakajima, Tomoka Ishida, Makoto Inoue
{"title":"FACTORS AFFECTING ASPIRATION VOLUMES OF BACK-FLUSH NEEDLES AND INFUSION VOLUMES OF VITRECTOMY DEVICE.","authors":"Masaharu Mizuno, Takashi Koto, Kosuke Nakajima, Tomoka Ishida, Makoto Inoue","doi":"10.1097/IAE.0000000000004326","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The stability of the eye is important during vitrectomy, which makes it important to balance the infusion rates (IRs) and the aspiration rates (ARs). The purpose of this study was to determine the effects of the IR from the vitrectomy device and the AR with back-flush needles.</p><p><strong>Methods: </strong>We compared the ARs of 25-gauge or 27-gauge back-flush needles manufactured by four companies. The AR of balanced salt solution (BSS), ethylene glycol mimicking subretinal fluid, and air was measured for each aspiration pressure. The IR of BSS and air was measured at different intraocular pressures.</p><p><strong>Results: </strong>The AR increased in proportion to the increase in the aspiration pressure, and it was positively correlated with the cross-sectional area of the soft tips and negatively with the total length of the soft tips. The IR of BSS and air increased in proportion to the increase in the intraocular pressures settings. The IR of BSS exceeded the AR of BSS and ethylene glycol for all parameters, but the IR of air could be less than the AR at some settings.</p><p><strong>Conclusion: </strong>The results suggest that the aspiration of air may cause a collapse of the eye under some settings.</p>","PeriodicalId":54486,"journal":{"name":"Retina-The Journal of Retinal and Vitreous Diseases","volume":"45 3","pages":"478-485"},"PeriodicalIF":2.3000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Retina-The Journal of Retinal and Vitreous Diseases","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/IAE.0000000000004326","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The stability of the eye is important during vitrectomy, which makes it important to balance the infusion rates (IRs) and the aspiration rates (ARs). The purpose of this study was to determine the effects of the IR from the vitrectomy device and the AR with back-flush needles.

Methods: We compared the ARs of 25-gauge or 27-gauge back-flush needles manufactured by four companies. The AR of balanced salt solution (BSS), ethylene glycol mimicking subretinal fluid, and air was measured for each aspiration pressure. The IR of BSS and air was measured at different intraocular pressures.

Results: The AR increased in proportion to the increase in the aspiration pressure, and it was positively correlated with the cross-sectional area of the soft tips and negatively with the total length of the soft tips. The IR of BSS and air increased in proportion to the increase in the intraocular pressures settings. The IR of BSS exceeded the AR of BSS and ethylene glycol for all parameters, but the IR of air could be less than the AR at some settings.

Conclusion: The results suggest that the aspiration of air may cause a collapse of the eye under some settings.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.70
自引率
9.10%
发文量
554
审稿时长
3-6 weeks
期刊介绍: ​RETINA® focuses exclusively on the growing specialty of vitreoretinal disorders. The Journal provides current information on diagnostic and therapeutic techniques. Its highly specialized and informative, peer-reviewed articles are easily applicable to clinical practice. In addition to regular reports from clinical and basic science investigators, RETINA® publishes special features including periodic review articles on pertinent topics, special articles dealing with surgical and other therapeutic techniques, and abstract cards. Issues are abundantly illustrated in vivid full color. Published 12 times per year, RETINA® is truly a “must have” publication for anyone connected to this field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信