Baptiste Dubot, Maxime Deniau, Donatien Ramiandrisoa, Jean-Gabriel Minonzio
{"title":"Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population.","authors":"Baptiste Dubot, Maxime Deniau, Donatien Ramiandrisoa, Jean-Gabriel Minonzio","doi":"10.3791/66985","DOIUrl":null,"url":null,"abstract":"<p><p>Fragility fractures are still a worldwide health burden in the context of population aging. In particular, the global number of hip fractures is expected to double between 2020 and 2050. Therefore, it is essential to detect patients at risk of fragility fracture at the population scale. The current golden standard is dual X-ray absorptiometry (DXA), providing the areal bone mineral density (aBMD). Ultrasonic devices, usually more portable and cheaper than X-ray devices, represent interesting DXA alternatives as screening tools. However, operator dependency is usually recognized as their main drawback. In this study, the measurement protocol of the bi-directional axial transmission (BDAT) ultrasonic device is presented in detail. The dedicated ultrasonic probe is placed at the one-third distal radius of the non-dominant forearm using conventional coupling gel. The guided interface provides in quasi-real time (about 2 Hz) four parameters of interest: velocities of the first arriving signal (vFAS) and the A0 mode (vA0), cortical thickness (Ct.Th) and porosity (Ct.Po), as well as four quality parameters. The operator moves the probe slowly at the measurement site, carefully observing the feedback provided by the interface until finding a stable position and starting a series of 10 acquisitions. When at least four consistent series are obtained, the measurement ends, and an automatic report is generated. The measurement usually takes about 5 min to complete. To determine the robustness of this protocol, a reproducibility study was conducted among 3 operators (one expert and two novices) and 14 healthy participants (6 women, 8 men, 21-53 years old). The intraclass correlation coefficients (ICC) were found good for vA0 (0.76), Ct.Po (0.80) or excellent for Ct.Th (0.87) and vFAS (0.91). The standard deviations were found to be less than 10% of the total ranges in clinical practice.</p>","PeriodicalId":48787,"journal":{"name":"Jove-Journal of Visualized Experiments","volume":" 215","pages":""},"PeriodicalIF":1.2000,"publicationDate":"2025-01-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jove-Journal of Visualized Experiments","FirstCategoryId":"103","ListUrlMain":"https://doi.org/10.3791/66985","RegionNum":4,"RegionCategory":"综合性期刊","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MULTIDISCIPLINARY SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
Fragility fractures are still a worldwide health burden in the context of population aging. In particular, the global number of hip fractures is expected to double between 2020 and 2050. Therefore, it is essential to detect patients at risk of fragility fracture at the population scale. The current golden standard is dual X-ray absorptiometry (DXA), providing the areal bone mineral density (aBMD). Ultrasonic devices, usually more portable and cheaper than X-ray devices, represent interesting DXA alternatives as screening tools. However, operator dependency is usually recognized as their main drawback. In this study, the measurement protocol of the bi-directional axial transmission (BDAT) ultrasonic device is presented in detail. The dedicated ultrasonic probe is placed at the one-third distal radius of the non-dominant forearm using conventional coupling gel. The guided interface provides in quasi-real time (about 2 Hz) four parameters of interest: velocities of the first arriving signal (vFAS) and the A0 mode (vA0), cortical thickness (Ct.Th) and porosity (Ct.Po), as well as four quality parameters. The operator moves the probe slowly at the measurement site, carefully observing the feedback provided by the interface until finding a stable position and starting a series of 10 acquisitions. When at least four consistent series are obtained, the measurement ends, and an automatic report is generated. The measurement usually takes about 5 min to complete. To determine the robustness of this protocol, a reproducibility study was conducted among 3 operators (one expert and two novices) and 14 healthy participants (6 women, 8 men, 21-53 years old). The intraclass correlation coefficients (ICC) were found good for vA0 (0.76), Ct.Po (0.80) or excellent for Ct.Th (0.87) and vFAS (0.91). The standard deviations were found to be less than 10% of the total ranges in clinical practice.
期刊介绍:
JoVE, the Journal of Visualized Experiments, is the world''s first peer reviewed scientific video journal. Established in 2006, JoVE is devoted to publishing scientific research in a visual format to help researchers overcome two of the biggest challenges facing the scientific research community today; poor reproducibility and the time and labor intensive nature of learning new experimental techniques.