Thomas Mandel Clausen, Ryan Nakamura, Andie Conching, Joo Won Choi, Yi Jonathan Zhang, Ferdinand Hui, Samuel Tsappidi
{"title":"Flow diversion in the treatment of intracranial aneurysms using the novel FRED X device: An early experience from a single high-volume center.","authors":"Thomas Mandel Clausen, Ryan Nakamura, Andie Conching, Joo Won Choi, Yi Jonathan Zhang, Ferdinand Hui, Samuel Tsappidi","doi":"10.1177/15910199251319059","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The Flow Re-Direction Endoluminal Device X (FRED X) offers several benefits over other flow-diverter devices including an antithrombotic coating, optimized in-vessel stability, and increased flexibility for easier device placement. We present a to-date experience of the safety and utility of the FRED X device in the repair of posterior and anterior circulation aneurysms.</p><p><strong>Methods: </strong>A retrospective review was conducted on all endovascular procedures that utilized the FRED X device at our center from May 2022 to November 2023.</p><p><strong>Results: </strong>77 patients (72.7% women, mean age 58.9), underwent a total of 85 procedures using the FRED X device. Indications included treatment of incidentally discovered aneurysms, acute dissections, aneurysm rupture, repair of residual filling following prior intervention, and use of FRED X for recanalization of non-aneurysmal extracranial stroke. 31.3% of the aneurysms were in the posterior circulation, 68.7% were in the anterior circulation. 9.4% of patients presented with SAH due to acute aneurysm rupture. Patients treated with FRED X were separated into OFF-Label (40.0%) or ON-label (60.0%) indications. Occlusion rate at 6-month follow-up were 72.2% in the OFF-label group, 66.7% in the ON-label group, and 68.4% overall. Rate of major periprocedural complications was 1.2% and the cumulative rate of postprocedural complication at follow-up was 5.3%.</p><p><strong>Conclusion: </strong>This study shows that FRED X treatment of intracranial aneurysms is safe in both OFF-label and ON-label indications. Continued follow-up of our patient population will further establish the safety, efficacy, and long-term stability of this device.</p>","PeriodicalId":14380,"journal":{"name":"Interventional Neuroradiology","volume":" ","pages":"15910199251319059"},"PeriodicalIF":1.7000,"publicationDate":"2025-02-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11833799/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Interventional Neuroradiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/15910199251319059","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The Flow Re-Direction Endoluminal Device X (FRED X) offers several benefits over other flow-diverter devices including an antithrombotic coating, optimized in-vessel stability, and increased flexibility for easier device placement. We present a to-date experience of the safety and utility of the FRED X device in the repair of posterior and anterior circulation aneurysms.
Methods: A retrospective review was conducted on all endovascular procedures that utilized the FRED X device at our center from May 2022 to November 2023.
Results: 77 patients (72.7% women, mean age 58.9), underwent a total of 85 procedures using the FRED X device. Indications included treatment of incidentally discovered aneurysms, acute dissections, aneurysm rupture, repair of residual filling following prior intervention, and use of FRED X for recanalization of non-aneurysmal extracranial stroke. 31.3% of the aneurysms were in the posterior circulation, 68.7% were in the anterior circulation. 9.4% of patients presented with SAH due to acute aneurysm rupture. Patients treated with FRED X were separated into OFF-Label (40.0%) or ON-label (60.0%) indications. Occlusion rate at 6-month follow-up were 72.2% in the OFF-label group, 66.7% in the ON-label group, and 68.4% overall. Rate of major periprocedural complications was 1.2% and the cumulative rate of postprocedural complication at follow-up was 5.3%.
Conclusion: This study shows that FRED X treatment of intracranial aneurysms is safe in both OFF-label and ON-label indications. Continued follow-up of our patient population will further establish the safety, efficacy, and long-term stability of this device.
期刊介绍:
Interventional Neuroradiology (INR) is a peer-reviewed clinical practice journal documenting the current state of interventional neuroradiology worldwide. INR publishes original clinical observations, descriptions of new techniques or procedures, case reports, and articles on the ethical and social aspects of related health care. Original research published in INR is related to the practice of interventional neuroradiology...