A Computational Study on the Neck-Stem Rectangular Tapered Connection: Effects of Angular Mismatch, Assembly, and Cyclic Loading

IF 2.2 4区 医学 Q3 ENGINEERING, BIOMEDICAL
R. Cromi, L. Ciriello, F. Berti, L. La Barbera, T. Villa, G. Pennati
{"title":"A Computational Study on the Neck-Stem Rectangular Tapered Connection: Effects of Angular Mismatch, Assembly, and Cyclic Loading","authors":"R. Cromi,&nbsp;L. Ciriello,&nbsp;F. Berti,&nbsp;L. La Barbera,&nbsp;T. Villa,&nbsp;G. Pennati","doi":"10.1002/cnm.3909","DOIUrl":null,"url":null,"abstract":"<p>The bi-modular hip prosthesis is characterized by two tapered connections: a circular cross-section at the head–neck interface and a rectangular cross-section at the neck-stem interface. Even if the latter guarantees customization, it concerns a high rate of early failure. The connection resistance is relatable to machining (tolerances cause angular mismatch), implantation (hammering force or manual), and usage (Body Mass Index [BMI]). Due to the lack of literature about the neck-stem coupling, this work aims to investigate how the geometry of the rectangular taper connection and the external loads affect the fatigue strength of a bi-modular hip prosthesis through a 3D Finite Element Model (FEM). Nine combinations of neck-stem coupling are obtained considering the tolerances' limits on frontal and lateral angles as <span></span><math>\n <semantics>\n <mrow>\n <mn>4</mn>\n <mo>°</mo>\n <msup>\n <mspace></mspace>\n <mrow>\n <mtable>\n <mtr>\n <mtd>\n <mo>+</mo>\n <msup>\n <mn>6</mn>\n <mo>′</mo>\n </msup>\n </mtd>\n </mtr>\n <mtr>\n <mtd>\n <msup>\n <mn>0</mn>\n <mo>′</mo>\n </msup>\n </mtd>\n </mtr>\n </mtable>\n </mrow>\n </msup>\n </mrow>\n <annotation>$$ {4}^{{}^{\\circ}\\begin{array}{c}+{6}^{\\prime}\\\\ {}{0}^{\\prime}\\end{array}} $$</annotation>\n </semantics></math>. The CoCr neck and the Ti6Al4V stem, studied in their halved, are constrained and loaded inspired by the standard ISO 7206: the stem is distally encastered simulating the embedding and tilted by 10° concerning the sagittal plane, while the force is applied vertically. First, the influence of the assembly is investigated using <span></span><math>\n <semantics>\n <mrow>\n <mn>0.3</mn>\n <mspace></mspace>\n <mi>kN</mi>\n </mrow>\n <annotation>$$ 0.3\\ \\mathrm{kN} $$</annotation>\n </semantics></math>, <span></span><math>\n <semantics>\n <mrow>\n <mn>2</mn>\n <mspace></mspace>\n <mi>kN</mi>\n </mrow>\n <annotation>$$ 2\\ \\mathrm{kN} $$</annotation>\n </semantics></math>, and <span></span><math>\n <semantics>\n <mrow>\n <mn>4</mn>\n <mspace></mspace>\n <mi>kN</mi>\n </mrow>\n <annotation>$$ 4\\ \\mathrm{kN} $$</annotation>\n </semantics></math>; then, a cyclical vertical force varying from <span></span><math>\n <semantics>\n <mrow>\n <mn>2.67</mn>\n <mspace></mspace>\n <mi>kN</mi>\n </mrow>\n <annotation>$$ 2.67\\ \\mathrm{kN} $$</annotation>\n </semantics></math> to <span></span><math>\n <semantics>\n <mrow>\n <mn>5.34</mn>\n <mspace></mspace>\n <mi>kN</mi>\n </mrow>\n <annotation>$$ 5.34\\ \\mathrm{kN} $$</annotation>\n </semantics></math> is imposed. Finally, one combination is analyzed in its integrity to evaluate the effect of the out-of-plane load. The study's findings concern: (i) a positive angular mismatch, which is responsible for proximal contacts, improves fatigue life, reducing <i>Sines</i> stress up to 33%; (ii) the higher the assembly force the higher the neck stability and the lower the extension of the overstressed lateral area; (iii) the implant fatigue resistance is directly proportional to the patient's BMI; and (iv) the out-of-plane external load causes a 40% increment in the fatigue failure risk.</p>","PeriodicalId":50349,"journal":{"name":"International Journal for Numerical Methods in Biomedical Engineering","volume":"41 2","pages":""},"PeriodicalIF":2.2000,"publicationDate":"2025-02-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/cnm.3909","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal for Numerical Methods in Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cnm.3909","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

The bi-modular hip prosthesis is characterized by two tapered connections: a circular cross-section at the head–neck interface and a rectangular cross-section at the neck-stem interface. Even if the latter guarantees customization, it concerns a high rate of early failure. The connection resistance is relatable to machining (tolerances cause angular mismatch), implantation (hammering force or manual), and usage (Body Mass Index [BMI]). Due to the lack of literature about the neck-stem coupling, this work aims to investigate how the geometry of the rectangular taper connection and the external loads affect the fatigue strength of a bi-modular hip prosthesis through a 3D Finite Element Model (FEM). Nine combinations of neck-stem coupling are obtained considering the tolerances' limits on frontal and lateral angles as 4 ° + 6 0 $$ {4}^{{}^{\circ}\begin{array}{c}+{6}^{\prime}\\ {}{0}^{\prime}\end{array}} $$ . The CoCr neck and the Ti6Al4V stem, studied in their halved, are constrained and loaded inspired by the standard ISO 7206: the stem is distally encastered simulating the embedding and tilted by 10° concerning the sagittal plane, while the force is applied vertically. First, the influence of the assembly is investigated using 0.3 kN $$ 0.3\ \mathrm{kN} $$ , 2 kN $$ 2\ \mathrm{kN} $$ , and 4 kN $$ 4\ \mathrm{kN} $$ ; then, a cyclical vertical force varying from 2.67 kN $$ 2.67\ \mathrm{kN} $$ to 5.34 kN $$ 5.34\ \mathrm{kN} $$ is imposed. Finally, one combination is analyzed in its integrity to evaluate the effect of the out-of-plane load. The study's findings concern: (i) a positive angular mismatch, which is responsible for proximal contacts, improves fatigue life, reducing Sines stress up to 33%; (ii) the higher the assembly force the higher the neck stability and the lower the extension of the overstressed lateral area; (iii) the implant fatigue resistance is directly proportional to the patient's BMI; and (iv) the out-of-plane external load causes a 40% increment in the fatigue failure risk.

Abstract Image

求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal for Numerical Methods in Biomedical Engineering
International Journal for Numerical Methods in Biomedical Engineering ENGINEERING, BIOMEDICAL-MATHEMATICAL & COMPUTATIONAL BIOLOGY
CiteScore
4.50
自引率
9.50%
发文量
103
审稿时长
3 months
期刊介绍: All differential equation based models for biomedical applications and their novel solutions (using either established numerical methods such as finite difference, finite element and finite volume methods or new numerical methods) are within the scope of this journal. Manuscripts with experimental and analytical themes are also welcome if a component of the paper deals with numerical methods. Special cases that may not involve differential equations such as image processing, meshing and artificial intelligence are within the scope. Any research that is broadly linked to the wellbeing of the human body, either directly or indirectly, is also within the scope of this journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信