Comparison of the diagnostic accuracy of VSBONE BSI versions for detecting bone metastases in breast and prostate carcinoma patients using conventional and CZT detector gamma cameras

IF 2.5 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Akiko Oya, Michihiro Nakayama, Takahiro Uno, Haruka Ueda, Ippei Yoshida, Atsushi Fujiya, Yurina Nomura, Atsutaka Okizaki
{"title":"Comparison of the diagnostic accuracy of VSBONE BSI versions for detecting bone metastases in breast and prostate carcinoma patients using conventional and CZT detector gamma cameras","authors":"Akiko Oya,&nbsp;Michihiro Nakayama,&nbsp;Takahiro Uno,&nbsp;Haruka Ueda,&nbsp;Ippei Yoshida,&nbsp;Atsushi Fujiya,&nbsp;Yurina Nomura,&nbsp;Atsutaka Okizaki","doi":"10.1007/s12149-025-02020-z","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>Bone scintigraphy is widely employed for detecting bone metastases, with the bone scan index (BSI) gaining traction as a quantitative tool in this domain. VSBONE BSI, an automated image analysis software, identifies abnormal hyperaccumulation areas in bone scintigraphy and computes BSI scores. The software, originally developed using data from conventional gamma cameras (C-Camera), has undergone two upgrades. This study hypothesized that the upgrades enhance the diagnostic accuracy for bone metastases and assessed the software’s applicability to images obtained using a cadmium–zinc–telluride detector gamma camera (CZT-Camera). The aim was to compare the diagnostic accuracy of VSBONE BSI across software versions using both conventional and CZT detectors and to evaluate its utility.</p><h3>Methods</h3><p>A total of 287 patients with breast or prostate carcinoma who underwent whole-body bone scintigraphy were included. VSBONE BSI automatically analyzed and calculated the BSI. The analysis results were compared with the presence or absence of metastases for each software version by using detector type of camera. The diagnostic agreement was evaluated.</p><h3>Results</h3><p>Receiver operating characteristic analysis showed an area under the curve (AUC) exceeding 0.7 across all groups, indicating good diagnostic performance. AUC values significantly increased with version upgrades for all patients and for breast carcinoma patients. In metastasis-negative cases, BSI values decreased with each software version upgrade, with the reduction being more pronounced in breast carcinoma patients scanned with the CZT-Camera.</p><h3>Conclusions</h3><p>Using the VSBONE BSI, version 2 or 3 had a higher rate of diagnostic concordance with the clinical prognosis than version 1. In metastasis-negative patients, newer software versions yielded lower BSI values, especially for breast carcinoma patients scanned using the CZT-Camera, highlighting the improved diagnostic accuracy of the updated software.</p></div>","PeriodicalId":8007,"journal":{"name":"Annals of Nuclear Medicine","volume":"39 5","pages":"466 - 475"},"PeriodicalIF":2.5000,"publicationDate":"2025-02-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Nuclear Medicine","FirstCategoryId":"3","ListUrlMain":"https://link.springer.com/article/10.1007/s12149-025-02020-z","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

Bone scintigraphy is widely employed for detecting bone metastases, with the bone scan index (BSI) gaining traction as a quantitative tool in this domain. VSBONE BSI, an automated image analysis software, identifies abnormal hyperaccumulation areas in bone scintigraphy and computes BSI scores. The software, originally developed using data from conventional gamma cameras (C-Camera), has undergone two upgrades. This study hypothesized that the upgrades enhance the diagnostic accuracy for bone metastases and assessed the software’s applicability to images obtained using a cadmium–zinc–telluride detector gamma camera (CZT-Camera). The aim was to compare the diagnostic accuracy of VSBONE BSI across software versions using both conventional and CZT detectors and to evaluate its utility.

Methods

A total of 287 patients with breast or prostate carcinoma who underwent whole-body bone scintigraphy were included. VSBONE BSI automatically analyzed and calculated the BSI. The analysis results were compared with the presence or absence of metastases for each software version by using detector type of camera. The diagnostic agreement was evaluated.

Results

Receiver operating characteristic analysis showed an area under the curve (AUC) exceeding 0.7 across all groups, indicating good diagnostic performance. AUC values significantly increased with version upgrades for all patients and for breast carcinoma patients. In metastasis-negative cases, BSI values decreased with each software version upgrade, with the reduction being more pronounced in breast carcinoma patients scanned with the CZT-Camera.

Conclusions

Using the VSBONE BSI, version 2 or 3 had a higher rate of diagnostic concordance with the clinical prognosis than version 1. In metastasis-negative patients, newer software versions yielded lower BSI values, especially for breast carcinoma patients scanned using the CZT-Camera, highlighting the improved diagnostic accuracy of the updated software.

Abstract Image

VSBONE BSI版本对常规和CZT伽玛仪检测乳腺癌和前列腺癌骨转移的诊断准确性比较
目的:骨显像被广泛用于检测骨转移,骨扫描指数(Bone scan index, BSI)作为一种定量工具在该领域受到关注。VSBONE BSI是一种自动图像分析软件,可识别骨闪烁成像中异常的超积累区域并计算BSI分数。该软件最初是利用传统伽马照相机(C-Camera)的数据开发的,经过了两次升级。本研究假设升级提高了骨转移的诊断准确性,并评估了该软件对使用碲化镉锌探测器伽马相机(CZT-Camera)获得的图像的适用性。目的是比较使用传统和CZT检测器的VSBONE BSI在不同软件版本的诊断准确性,并评估其实用性。方法:选取287例行全身骨显像检查的乳腺癌或前列腺癌患者。VSBONE BSI自动分析和计算BSI。采用检测器式相机对各软件版本的转移情况进行比较分析。评估诊断一致性。结果:受试者工作特征分析显示,所有组的曲线下面积(AUC)均超过0.7,表明诊断效果良好。所有患者和乳腺癌患者的AUC值随着版本升级而显著增加。在转移阴性的病例中,BSI值随着软件版本的每次升级而下降,在使用CZT-Camera扫描的乳腺癌患者中下降更为明显。结论:使用VSBONE BSI,版本2和版本3的诊断与临床预后的符合率高于版本1。在转移阴性患者中,更新的软件版本产生较低的BSI值,特别是对于使用CZT-Camera扫描的乳腺癌患者,突出了更新软件提高的诊断准确性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of Nuclear Medicine
Annals of Nuclear Medicine 医学-核医学
CiteScore
4.90
自引率
7.70%
发文量
111
审稿时长
4-8 weeks
期刊介绍: Annals of Nuclear Medicine is an official journal of the Japanese Society of Nuclear Medicine. It develops the appropriate application of radioactive substances and stable nuclides in the field of medicine. The journal promotes the exchange of ideas and information and research in nuclear medicine and includes the medical application of radionuclides and related subjects. It presents original articles, short communications, reviews and letters to the editor.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信