{"title":"Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase thoracoabdominal CT.","authors":"Daniel Popp, Martin Siedlecki, Lena Friedrich, Mark Haerting, Christian Scheurig-Muenkler, Florian Schwarz, Thomas Kroencke, Stefanie Bette, Josua A Decker","doi":"10.1007/s00330-025-11409-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To compare image quality and iodine attenuation intra-individually in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium (CM) volume.</p><p><strong>Materials and methods: </strong>A prospectively acquired patient cohort between 04/2021 and 11/2023 was retrospectively screened if patients had the following combination of portal venous phase thoracoabdominal CT scans: (a) PCD-CT with 120 mL CM volume (PCD-CT<sub>120 mL</sub>), (b) PCD-CT with 100 mL CM volume (PCD-CT<sub>100 mL</sub>), and (c) prior energy-integrating detector CT (EID-CT) with 120 mL CM volume. On PCD-CT, virtual monoenergetic image (VMI) reconstructions at 70 keV were applied for both groups as well as additional VMI at 60 keV for PCD‑CT<sub>100 mL</sub>. Quantitative analyses including signal-to-noise (SNR) and contrast-to-noise ratios (CNR) and qualitative analyses were performed using a mixed linear effects model.</p><p><strong>Results: </strong>The final study cohort comprised 49 patients (mean age 67 [31-86] years, 12 female). Comparison to EID-CT was available in 33 patients. In standard 70 keV VMI reconstructions, PCD-CT<sub>100 mL</sub> was non-inferior to PCD-CT<sub>120 mL</sub> as well as to EID-CT<sub>120 mL</sub> for CNR in abdominal organs (all p > 0.050). The mixed linear effects model revealed significant differences between contrast volume groups for both contrast enhancement and image quality ratings. PCD-CT<sub>100 mL/70 keV</sub> demonstrated the smallest deviation from optimal contrast enhancement (-0.306, p < 0.001).</p><p><strong>Conclusion: </strong>In portal venous phase thoracoabdominal PCD-CT, a nearly 17% reduction in CM was achievable while maintaining subjective and objective image quality compared to prior higher CM volume PCD-CT scans within the same patients and still surpassing image quality of previous exams on an EID-CT system.</p><p><strong>Key points: </strong>Question How do image quality and iodine attenuation intra-individually compare in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium volume. Findings PCD-CT scans exhibit superior quantitative and qualitative image quality compared to energy-integrating detector-CT acquisitions and are not negatively affected by contrast volume reductions up to 17%. Clinical relevance This study provides further evidence that PCD-CT enables a considerable reduction in iodine dose for portal venous phase acquisition, benefiting both patients and healthcare system costs.</p>","PeriodicalId":12076,"journal":{"name":"European Radiology","volume":" ","pages":"4635-4648"},"PeriodicalIF":4.7000,"publicationDate":"2025-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00330-025-11409-3","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/2/12 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: To compare image quality and iodine attenuation intra-individually in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium (CM) volume.
Materials and methods: A prospectively acquired patient cohort between 04/2021 and 11/2023 was retrospectively screened if patients had the following combination of portal venous phase thoracoabdominal CT scans: (a) PCD-CT with 120 mL CM volume (PCD-CT120 mL), (b) PCD-CT with 100 mL CM volume (PCD-CT100 mL), and (c) prior energy-integrating detector CT (EID-CT) with 120 mL CM volume. On PCD-CT, virtual monoenergetic image (VMI) reconstructions at 70 keV were applied for both groups as well as additional VMI at 60 keV for PCD‑CT100 mL. Quantitative analyses including signal-to-noise (SNR) and contrast-to-noise ratios (CNR) and qualitative analyses were performed using a mixed linear effects model.
Results: The final study cohort comprised 49 patients (mean age 67 [31-86] years, 12 female). Comparison to EID-CT was available in 33 patients. In standard 70 keV VMI reconstructions, PCD-CT100 mL was non-inferior to PCD-CT120 mL as well as to EID-CT120 mL for CNR in abdominal organs (all p > 0.050). The mixed linear effects model revealed significant differences between contrast volume groups for both contrast enhancement and image quality ratings. PCD-CT100 mL/70 keV demonstrated the smallest deviation from optimal contrast enhancement (-0.306, p < 0.001).
Conclusion: In portal venous phase thoracoabdominal PCD-CT, a nearly 17% reduction in CM was achievable while maintaining subjective and objective image quality compared to prior higher CM volume PCD-CT scans within the same patients and still surpassing image quality of previous exams on an EID-CT system.
Key points: Question How do image quality and iodine attenuation intra-individually compare in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium volume. Findings PCD-CT scans exhibit superior quantitative and qualitative image quality compared to energy-integrating detector-CT acquisitions and are not negatively affected by contrast volume reductions up to 17%. Clinical relevance This study provides further evidence that PCD-CT enables a considerable reduction in iodine dose for portal venous phase acquisition, benefiting both patients and healthcare system costs.
期刊介绍:
European Radiology (ER) continuously updates scientific knowledge in radiology by publication of strong original articles and state-of-the-art reviews written by leading radiologists. A well balanced combination of review articles, original papers, short communications from European radiological congresses and information on society matters makes ER an indispensable source for current information in this field.